1Entering the North American Market
- The Regulatory Landscape Device Regulation in
Canada - Nancy Ruth
- Director, Medical Devices
- AdvaMed September 21, 2008
2Canadian Medical Devices Regulations (CMDR)
- Authority from Food and Drug Act
- Health Canada - regulator
- Mandate protect Health of citizens
- Safe, effective therapeutics - medical devices
- Pre-market and post-market activities
- US - CFR
- Food Drug and Cosmetic Act
- FDA
- Protection safe/effective
- Pre-market and post-market activities
3Pre-market Activities
- Clinical studies Investigational Testing
Authorization (ITA) US IDE - Medical Device Licences US 510(k), PMA
- Quality system ISO 13485 certified 3rd party
recognized registrars US QS Reg not certified
FDA inspection - Medical Device Establishment Licence US
Establishment registration/device listing
4Post-Market Activities
- Medical Device Establishment Licence annual
update/renewal US FDA Establishment
registration/device listing annual update/renewal - Inspection Program for Medical Devices
- Health Canada inspectors
- Regulatory compliance not quality systems
- Importers, distributors, manufacturers US FDA
inspection cycle -
- Investigations noncompliance, safety US FDA
investigations - Post-market data collection, trend analysis
- Mandatory problem reports, recalls, global
information sharing - US Medical Device Reports, recalls, global
information
5Canadian Medical Devices Regulations Risk based
classification system
- Class 1 (least risk)
- Class 2
- Class 3
- Class 4 (most risk)
6Canadian Medical Devices Regulations Licences
- Class 1
- No product licence needed for manufacturer
- No ISO 13485 certificate
- Medical Device Establishment Licence (MDEL)
product class, medical specialty - Manufacturer can be exempt if import/distribute
through MDEL holder - Retailer, healthcare institution, healthcare
professional exempt - 2010 annually, attestations
- fee increases expected in 2009
7Canadian Medical Devices Regulations Licences
- Class 2, 3, 4
- Medical device licence held by manufacturer
- Apply to Health Canada
- Class 2 minimal safety/effectiveness fee 200
- Class 3, 4 premarket review document for
safety/effectiveness fees vary per submitted
information -approx 2000 - 10,000 - Amendment for significant changes
- No substantial equivalence process stand alone
submissions - Private label licence provision
- Exact product, intended use as original licence
attestation - Authorization by original licence holder
cross-reference - No ISO 13485 for private labeler
- fee increases expected in 2009
8Canadian Medical Devices Regulations Target
timelines- medical device licence review
- Class 2 15 days administrative
- Class 3 15 day screen, 60 day review
administrative - Class 4 15 day screen, 75 day review
administrative - Questions screening, review (additional
information) new cycles - Reality some delays notably IVDD backlog
9Canadian Medical Devices Regulations Quality
system
- Class 2, 3, 4 licence applications
- Quality system certificate submitted
- CMDCAS (Canadian Medical Device Conformity
Assessment System) - Recognized registrar third party
- Audit, issue certificate
- 3 year cycle certification, 2 x annual
surveillance, recertification - Manufacturer contracts registrar, pays fees
- Global picture
10Canadian Medical Devices Regulations
Investigational Testing Authorization
- Clinical study- submission to Health Canada
safety, likelihood of achieving study purpose - Class 2 fewer requirements
- Protocol, description, instructions for use,
informed consent ethics approval/investigator
agreement not submitted - Class 3, 4 - higher risk, more information
submitted - Pre-clinical, bench testing, risk analysis
- Manufacturing information
- ISO 13485 not required but quality/safety
assurance needed - Ethics board approval/investigator agreement
submitted - 30 day review target authorization letter
needed - No fee
11Canadian Medical Device Regulations Differences
- US to Canada
- ISO 13485 certified quality system CMDCAS
recognized registrar - No substantial equivalence process more than
510(k) dossier for Class III, IV - Review fees less costly than PMA
- Increasing safety documentation scrutiny
- FDA guidances may be used by Health Canada
- Classification correlation
- device/drug/NHP (natural health product)
designation correlation Canada vs US vs ROW - Manufacturer definition
- Labeling, trade mark issues
12Canadian Medical Devices Regulationsmanufacturer
definition
- manufacturer means a person who sells a medical
device under their own name, or under a
trade-mark, design, trade name or other name or
mark owned or controlled by the person, and who
is responsible for designing, manufacturing,
assembling, processing, labelling, packaging,
refurbishing or modifying the device, or for
assigning to it a purpose, whether those tasks
are performed by that person or on their behalf.
(fabricant) - Label, licence, QS certificate
- Trade-mark
13US to Canada
- Already cleared/approved in US?
- QS Regulations compliance
- Based on ISO 134851996 nothing contradictory
- CMDCAS gap add quality manual, CMDR compliance
references - Procedures- recall, mandatory problem reporting,
classification, labeling - Recognized registrar SCC, Health Canada
- Audit certificate
- CMDR compliant labeling
- Licence application Class 2,3,4
- Class 1 manufacturer MDEL (can be exempt
import/distribute MDEL holder) - Importer MDEL
14EU to Canada
- US companies already selling to EU product
already CE marked for EU? - Canada and EU are similar QS, classification
- ISO 134852003 standard
- Additional specific requirements
- CMDCAS adds CMDR (Canadian Medical Devices
Regulations) references, recall, mandatory
problem reporting, classification, labeling, etc - Registrar recognized by SCC Health Canada
- Certificate SCC logo, CMDCAS statement
- CMDR compliant labeling
15Canadian Medical Devices Regulations Importation
to Canada
- Medical Device Establishment Licence entity who
imports for sale - Money or not samples
- Exceptions importer of devices for
Investigational Testing Authorization, Special
Access/Custom Made, retailer, healthcare
professional/institution - Attestations
- Documented procedures
- Complaints, distribution records, mandatory
problem reports, recall - Shipping/handling, corrective action,
installation, servicing as applicable - Record keeping, classification, compliant labels
- Domestic distributors
- No ISO 13485 certificate for MDEL
16Comparison US, Canada, EU