A MICRO OPTICAL FORCE SENSOR FOR FORCE MEASUREMENT - PowerPoint PPT Presentation

1 / 23
About This Presentation
Title:

A MICRO OPTICAL FORCE SENSOR FOR FORCE MEASUREMENT

Description:

The 4 strain gauges are glued on the instrument shaft with 90 interval. ... diameter through which surgical instruments can be inserted. ... – PowerPoint PPT presentation

Number of Views:240
Avg rating:3.0/5.0
Slides: 24
Provided by: tam97
Category:

less

Transcript and Presenter's Notes

Title: A MICRO OPTICAL FORCE SENSOR FOR FORCE MEASUREMENT


1
A MICRO OPTICAL FORCE SENSOR FOR FORCE
MEASUREMENT
J. Peirs, J. Clijnen, D. Reynaerts, H. Van
Brussel, P. Herijgers, B. Corteville, S.
Boone Katholieke Universities Leuven, Dept. of
Mechanical Engineering, Belgium
- Vivek Kumbala (K00217072)
2
Overview
  • Introduction
  • Force Range and Sensor Design
  • Principle of Measurement
  • Flexible Structure
  • Calibration of the Sensor
  • Conclusion
  • Summary

3
Force Range
  • A test set-up was built to measure the forces
    occurring during suturing (stitching). The figure
    shows a needle driver equipped with a 2-component
    force sensor based on strain gauges.
  • The 4 strain gauges are glued on the instrument
    shaft with 90 interval. Two opposing strain
    gauges form a half bridge measuring the X or Y
    component of the force at the tip.
  • The Z component is assumed to be of the same
    order of magnitude. The strain gauges are
    connected to AC bridge amplifiers and sampled at
    250 Hz.

4
Needle driver with 2-component force sensor based
on strain gauges (enlarged).
5
Sensor Design
  • Specifications
  • The sensor will be mounted at the tip of a 5 mm
    diameter
  • instrument driver with two bending degrees of
    freedom.
  • This instrument driver has an internal channel of
    2 mm
  • diameter through which surgical instruments
    can be inserted.
  • The sensor should have the same internal and
    external
  • Diameters.

6
  • The sensor should be as short as possible because
    it will be
  • mounted in front of the local degrees of
    freedom
  • Three components (FX, FY, FZ) should be measured
    with a
  • range of 2.5 N and a resolution of 0.01 N
    (0.2 of range).
  • The radial forces act on the instrument tip, at a
    distance of
  • 15 mm from the sensor front. The sensor should
    be
  • biocompatible, sterilisable and robust.

7
Principle of Measurement
  • The working principle of the sensor Three
    optical
  • fibres measure the deformation of the
    flexible structure
  • through the intensity of the reflected
    light.
  • The basic layout of the sensor consists of
    consists of two
  • parts connected by a flexible connection.
  • The upper part is connected to the tool
  • The lower part is connected to the instrument
    shaft.

8
Basic Layout of the Sensor
9
  • Three optical fibres, arranged at 120 intervals
    in the lower part, measure the relative
    displacement between upper and lower part through
    the intensity of the reflected signal.
  • The fibres are placed axially because bending
    them inside the sensor over 90 would violate the
    minimum bending radius.
  • Three sensing fibres are used here each of which
    have two possible configurations.

10
  • The first configuration uses separate emitter and
    receiver fibres.
  • Here coupling losses do not occur, but high
    losses occur at the mirror.
  • As the receiver fibre is located next to the
    emitter fibre, it can pick up only a small
    portion of the reflected light, which is far less
    than 25 .
  • An additional drawback is that this configuration
    requires twice as much fibres to be integrated in
    the sensor. Therefore, the second configuration
    is chosen.

11
  • In the second configuration the light is
    reflected back into the same fibre.
  • An optocoupler has to be used to couple the
    emittor (LED)
  • and receiver (photodiode) to the same
    measurement fibre.
  • Maximum sensitivity is obtained when a 50/50
    optocoupler
  • is used. This means that 50 of the emitted
    signal is
  • coupled into the measurement fibre while 50
    goes to the
  • reference photodiode.

12
  • Similarly, the reflected signal is split equally
    into the emitter
  • and receiver fibres.
  • This means that maximally 25 of the originally
    emitted
  • light is sent back to the receiver (when the
    sensor reflects all
  • incoming light).
  • The sensing fibre can measure the perpendicular
    distance
  • from the surface or the lateral distance from
    an edge of this
  • surface, depending on the sensor design.

13
Flexible Structure
  • The design tries to decouple the deformations
    caused by axial and radial forces by using
    parallelograms.
  • An axial force causes the thin horizontal beams
    to bend, while the thick vertical beams deform
    negligibly.
  • For a radial force the vertical beams deform
    while the horizontal beams are mainly stressed
    longitudinally, a direction in which they have a
    high stiffness.
  • The flexible structure consists of 4 identical
    parallelograms
  • placed in an axisymmetric arrangement.

14
  • The sensor is made of a strong titanium alloy
    (Ti6Al4V) because this material has a good
    corrosion resistance, superior biocompatibility,
    low Youngs modulus and high strength (both
    maximizing sensor displacement), high fatigue
    resistance, and good shock resistance.

15
Flexible structure of the sensor (scale in
millimeters).
Dimensions of the flexible structure.
16
  • The circular holes are starting holes for the
    wire-EDM (Electro- Discharge Machine) process
    used to machine the slits.
  • The EDM wire has a diameter of 30 µm.
  • The square features located just above the upper
    holes are end stops protecting the sensor from
    axial overload.
  • The vertical gap is 50 µm corresponding to an
    axial load of 2.5 N
  • The end stop in radial direction is the core the
    gap between the core and the flexible tube is 85
    µm and closes at a radial force of 1.7N

17
Calibration
  • To calibrate the sensor, a short rod is fixed to
    its tip to act as a lever for applying torque.
  • The length of the lever, 15 mm, corresponds to
    the distance from the instrument tip to the
    sensor front.
  • The compliance matrix A between the applied
    forces Fi and the displacements di measured by
    the fibres is defined as follows

18
Reflected signal as a function of perpendicular
and lateral distance between fibre and Si surface
edge.
19
  • The design and calibration values for the
    compliance matrix are as shown (in µm/N).
  • These values can be converted to mV/N by
    multiplying them with the sensitivity of the
    optical system, 10.3 mV/µm
  • The design sensitivity matrix is not symmetric
    due to the 120 spacing between the fibres and
    the chosen coordinate frame.
  • The calibrated compliance matrix shows large
    differences with the design values because of
    production errors. But, the resolution of the
    optical measurement system is 0.3 µm,
    corresponding to about 0.04 N.

20
Conclusion
  • A 5 mm diameter tri-axial force sensor has been
    developed
  • for minimally invasive robotic surgery.
  • To define the required force range and
    resolution, a needle driver has been equipped
    with strain gauges. The vivo-tests with different
    types of needles and tissue showed that the
    required force range and resolution are
    respectively 2.5 N and 0.01 N.
  • The new sensor is based on a flexible titanium
    structure of which the deformations are measured
    through reflective measurements with 3 optical
    fibres. It has a range of 2.5 N in axial
    direction and 1.7 N in radial direction

21
References
  • http//www.mech.kuleuven.be/micro/pub/medic/Pape
    r_MME2003_MIS_ sensor.pdf
  • J. Rosen et al, Surgeon-Tool Force/Torque
    Signatures - Evaluation of Surgical Skills in
    Minimally Invasive Surgery, Proc. of medicine
    meets virtual reality, SanFrancisco, CA
  • I. Brouwer et al, Measuring in vivo animal soft
    tissue properties for haptic modeling in surgical
    simulation. Medicine meets virtual reality
  • J. Peirs, H. Van Brussel, D. Reynaerts, G. De
    Gersem, A Flexible Distal Tip With Two Degrees of
    Freedom for Enhanced Dexterity in Endoscopic
    Robot Surgery, Proc. Micromechanics Europe
    Workshop, Sinaia, Romania,

22
Summary
  • The design of a prototype gastro intestinal
    intervention system based on an inch-worm type of
    mobile robot was discussed in which the modular
    actuator for realizing the bending motion was
    presented and many technologies were discussed.
  • It has been, hence, concluded that the device
    can be used as a kind of vehicle for inspection
    in the intestinal system.
  • A 5mm diameter optical force sensor for
    minimally invasive robotic surgery has been
    developed, based on the specifications from
    in-vivo tests. And it has been found that it has
    a range of 2.5 N in axial direction and 1.7 N in
    radial direction.

23
THANKS!
QUESTIONS?
Write a Comment
User Comments (0)
About PowerShow.com