Title: A MICRO OPTICAL FORCE SENSOR FOR FORCE MEASUREMENT
1A MICRO OPTICAL FORCE SENSOR FOR FORCE
MEASUREMENT
J. Peirs, J. Clijnen, D. Reynaerts, H. Van
Brussel, P. Herijgers, B. Corteville, S.
Boone Katholieke Universities Leuven, Dept. of
Mechanical Engineering, Belgium
- Vivek Kumbala (K00217072)
2Overview
- Introduction
- Force Range and Sensor Design
- Principle of Measurement
- Flexible Structure
- Calibration of the Sensor
- Conclusion
- Summary
3Force Range
- A test set-up was built to measure the forces
occurring during suturing (stitching). The figure
shows a needle driver equipped with a 2-component
force sensor based on strain gauges. - The 4 strain gauges are glued on the instrument
shaft with 90 interval. Two opposing strain
gauges form a half bridge measuring the X or Y
component of the force at the tip. - The Z component is assumed to be of the same
order of magnitude. The strain gauges are
connected to AC bridge amplifiers and sampled at
250 Hz.
4Needle driver with 2-component force sensor based
on strain gauges (enlarged).
5Sensor Design
-
- Specifications
- The sensor will be mounted at the tip of a 5 mm
diameter - instrument driver with two bending degrees of
freedom. - This instrument driver has an internal channel of
2 mm - diameter through which surgical instruments
can be inserted. - The sensor should have the same internal and
external - Diameters.
6- The sensor should be as short as possible because
it will be - mounted in front of the local degrees of
freedom - Three components (FX, FY, FZ) should be measured
with a - range of 2.5 N and a resolution of 0.01 N
(0.2 of range). - The radial forces act on the instrument tip, at a
distance of - 15 mm from the sensor front. The sensor should
be - biocompatible, sterilisable and robust.
7Principle of Measurement
- The working principle of the sensor Three
optical - fibres measure the deformation of the
flexible structure - through the intensity of the reflected
light. -
- The basic layout of the sensor consists of
consists of two - parts connected by a flexible connection.
- The upper part is connected to the tool
- The lower part is connected to the instrument
shaft.
8Basic Layout of the Sensor
9- Three optical fibres, arranged at 120 intervals
in the lower part, measure the relative
displacement between upper and lower part through
the intensity of the reflected signal. - The fibres are placed axially because bending
them inside the sensor over 90 would violate the
minimum bending radius. - Three sensing fibres are used here each of which
have two possible configurations.
10- The first configuration uses separate emitter and
receiver fibres. - Here coupling losses do not occur, but high
losses occur at the mirror. - As the receiver fibre is located next to the
emitter fibre, it can pick up only a small
portion of the reflected light, which is far less
than 25 . - An additional drawback is that this configuration
requires twice as much fibres to be integrated in
the sensor. Therefore, the second configuration
is chosen.
11- In the second configuration the light is
reflected back into the same fibre. - An optocoupler has to be used to couple the
emittor (LED) - and receiver (photodiode) to the same
measurement fibre. - Maximum sensitivity is obtained when a 50/50
optocoupler - is used. This means that 50 of the emitted
signal is - coupled into the measurement fibre while 50
goes to the - reference photodiode.
12- Similarly, the reflected signal is split equally
into the emitter - and receiver fibres.
- This means that maximally 25 of the originally
emitted - light is sent back to the receiver (when the
sensor reflects all - incoming light).
- The sensing fibre can measure the perpendicular
distance - from the surface or the lateral distance from
an edge of this - surface, depending on the sensor design.
13Flexible Structure
- The design tries to decouple the deformations
caused by axial and radial forces by using
parallelograms. - An axial force causes the thin horizontal beams
to bend, while the thick vertical beams deform
negligibly. - For a radial force the vertical beams deform
while the horizontal beams are mainly stressed
longitudinally, a direction in which they have a
high stiffness. - The flexible structure consists of 4 identical
parallelograms - placed in an axisymmetric arrangement.
14- The sensor is made of a strong titanium alloy
(Ti6Al4V) because this material has a good
corrosion resistance, superior biocompatibility,
low Youngs modulus and high strength (both
maximizing sensor displacement), high fatigue
resistance, and good shock resistance.
15Flexible structure of the sensor (scale in
millimeters).
Dimensions of the flexible structure.
16- The circular holes are starting holes for the
wire-EDM (Electro- Discharge Machine) process
used to machine the slits. - The EDM wire has a diameter of 30 µm.
- The square features located just above the upper
holes are end stops protecting the sensor from
axial overload. - The vertical gap is 50 µm corresponding to an
axial load of 2.5 N - The end stop in radial direction is the core the
gap between the core and the flexible tube is 85
µm and closes at a radial force of 1.7N
17Calibration
- To calibrate the sensor, a short rod is fixed to
its tip to act as a lever for applying torque. -
- The length of the lever, 15 mm, corresponds to
the distance from the instrument tip to the
sensor front. - The compliance matrix A between the applied
forces Fi and the displacements di measured by
the fibres is defined as follows
18Reflected signal as a function of perpendicular
and lateral distance between fibre and Si surface
edge.
19- The design and calibration values for the
compliance matrix are as shown (in µm/N). - These values can be converted to mV/N by
multiplying them with the sensitivity of the
optical system, 10.3 mV/µm - The design sensitivity matrix is not symmetric
due to the 120 spacing between the fibres and
the chosen coordinate frame. - The calibrated compliance matrix shows large
differences with the design values because of
production errors. But, the resolution of the
optical measurement system is 0.3 µm,
corresponding to about 0.04 N.
20Conclusion
- A 5 mm diameter tri-axial force sensor has been
developed - for minimally invasive robotic surgery.
- To define the required force range and
resolution, a needle driver has been equipped
with strain gauges. The vivo-tests with different
types of needles and tissue showed that the
required force range and resolution are
respectively 2.5 N and 0.01 N. - The new sensor is based on a flexible titanium
structure of which the deformations are measured
through reflective measurements with 3 optical
fibres. It has a range of 2.5 N in axial
direction and 1.7 N in radial direction
21References
- http//www.mech.kuleuven.be/micro/pub/medic/Pape
r_MME2003_MIS_ sensor.pdf -
- J. Rosen et al, Surgeon-Tool Force/Torque
Signatures - Evaluation of Surgical Skills in
Minimally Invasive Surgery, Proc. of medicine
meets virtual reality, SanFrancisco, CA - I. Brouwer et al, Measuring in vivo animal soft
tissue properties for haptic modeling in surgical
simulation. Medicine meets virtual reality - J. Peirs, H. Van Brussel, D. Reynaerts, G. De
Gersem, A Flexible Distal Tip With Two Degrees of
Freedom for Enhanced Dexterity in Endoscopic
Robot Surgery, Proc. Micromechanics Europe
Workshop, Sinaia, Romania,
22Summary
- The design of a prototype gastro intestinal
intervention system based on an inch-worm type of
mobile robot was discussed in which the modular
actuator for realizing the bending motion was
presented and many technologies were discussed.
- It has been, hence, concluded that the device
can be used as a kind of vehicle for inspection
in the intestinal system. - A 5mm diameter optical force sensor for
minimally invasive robotic surgery has been
developed, based on the specifications from
in-vivo tests. And it has been found that it has
a range of 2.5 N in axial direction and 1.7 N in
radial direction.
23THANKS!
QUESTIONS?