Title: Measurement Systems International, Inc.


1
Measurement SystemsInternational, Inc.
Integrated Load Sensor Training
2
Table of Contents
  • 3. Basic Load Pin Overview and Specs
  • Load Pin Applications
  • Load Pin Application Questionnaire
  • Load Pin/Sensor System Diagram
  • Load Pin Installation Guidelines
  • Load Link Specs and Applications
  • Load Link Installation Guidelines
  • Wire Rope Clamp-On Sensor Specs and Applications
  • Sensor Cable Interface Guidelines
  • MSI-3550 Cable Interface
  • MSI-3550/ Scoreboard Interface
  • MSI-3650 Cable Interface
  • 14. MSI-3650/ Scoreboard Interface
  • MSI-9000 Cable Interface
  • MSI-9000/ Scoreboard Interface
  • General Calibration Guidelines
  • Calibration Setup
  • MSI-3550 Calibration Instructions
  • MSI-3550 Cal Instructions (contd)
  • MSI-3650/3750 Cal Instructions
  • MSI-9000 Cal Instructions
  • MSI-9000/CVM Cal Instructions

3
Load pin training
  • Overview, Specs, Applications, System Diagram
    Installation Guidelines

4
Basic Load Pin Overview and Specs
  • Standard Specifications
  • Overload
  • 200 without zero shift
  • 500 without failure (minimum)
  • Bridge
  • Full bridge 350 Ohm (nominal)
  • Excitation
  • 12 V AC or DC (Maximum)
  • Output Signal
  • 1 - 2-mv/V (Nominal)
  • Operating Temp Range
  • -40 to 150 F
  • Material
  • -- 17-4 Stainless Steel
  • Accuracy
  • 1-5 of FS depending on System Parameters
  • Electrical Connection
  • 6-Pin receptacle or built-in cable
  • Designed and manufactured specific to each
    integrated solution, the MSI load pin is strictly
    industrial grade.
  • Each load pin design is precision machined from
    17-4 stainless steel for safety, strength and
    corrosion resistance. Each strain gage location
    is precisely calculated and correctly placed in
    the optimum concentrated stress area applicable
    to each installation.
  • MSI load pin designs apply only internally
    mounted strain gages for complete protection from
    the outside environment. To ensure the highest
    level of accuracy, performance and quality, each
    load pin is also internally balanced and
    temperature compensated.
  • The MSI load pin is a key system component toward
    providing the highest quality integrated overhead
    weighing solutions. Each pin is specifically
    designed to replace existing sheave or shackle
    pins and pins located in a Wire Rope dead-end.
  • To complete the integrated solution, MSI provides
    a comprehensive product line of instrumentation
    for output signal conditioning, load indicating
    and wireless transmission of data for integration
    with existing plant process control equipment.

5
Load Pin Applications
  • Applications
  • Overload Monitoring
  • Harsh Environment
  • Slack Line Monitoring
  • Caught Load Monitoring
  • Low or No Headroom Applications
  • 1 - 5 Accuracy weighing system
  • Industries
  • Manufacturing
  • Foundry/Metals
  • Shipping/Container
  • Energy
  • Crane/ Hoist

6
Applications (in Pictures)
Container Crane Load Monitoring System
Integrated Coil Weighing System
7
Applications (in Pictures)
Integrated Spreader Bar
Basic Integrated Sheave System
8
Applications (in Pictures)
Wire Tension Monitoring System
9
Load Pin Application Questionnaire
When requesting quote for load pin system, please
complete MSIs Load Pin Application
Questionnaire. If you require any assistance,
please contact your MSI Sales Representative.
Pin Dimensions
Connector Type
Crane/ Hoist Information
Cable Length
10
MSI- Load Pin Diagram
Load Pin
Outer Sections
Load Arrow
Center Section
Each MSI Load Pin has a minimum of two principal
loading areas referred to as Center and Outer
Sections. Each MSI Load Pin has an etched Load
Arrow to aid the installer to position load pin
in correct orientation. Rotate the load pin to
position the Load Arrow in the same direction as
the load is applied on the Center Section.
11
Load Pin/Integrated Sensor System Diagram
12
Idler Sheave Installation
Support
1/32 Clearance Both sides
Load
  • Reduce errant effects of friction and/or binding,
    by allowing a minimum
  • clearance of 1/32 on each side of the
    idler sheave.
  • Position the load pin so that its Load Arrow is
    in the same direction as load is applied to the
    Center Section.
  • 3. Firmly secure the load pin Anti-Rotational
    Keeper Plate with the idler sheave support plate..

13
Dead End Installation
  • Reduce errant effects of friction and/or binding,
    by allowing a minimum
  • clearance of 1/32 between sides of clevis
    support and lug.
  • Position the load pin so that its Load Arrow is
    in the same direction as load is applied to the
    Center Section.
  • 3. Firmly secure the load pin Anti-Rotational
    Keeper Plate with the support clevis.

14
Load Link training
  • Overview, Specs, Applications Installation
    Guidelines

15
Load Link Specification Overview
  • Load Link Overview
  • This load link style load cell is designed for
    tension applications such as Wire Rope Dead Ends.
    Other applications include winching, crane
    hoist scaling, and load monitoring systems. 
  • Openings at both ends of the load link allow for
    easy connection to shackles, wire rope, or other
    rigging.  
  • Load Links are available in capacities ranging
    from 250 pounds to over 500,000 lbs.  Custom size
    load links with specified safety factors are
    available upon request. 
  • Standard Specifications
  • Accuracy 0.1 of rated capacity and 0.5 on
    capacities over 50 tons
  • Operating Temperature -4 to 140 F (-20 to
    60 C)
  • Enclosure Corrosion resistant and
    environmentally sealed
  • Overload 200 safe 500 ultimate
  • Calibration Digital

16
Load Link Applications
  • Installed on crane wire rope dead-ends.
  • 1 to 5 accurate weighing systems
  • Load Monitoring Applications
  • Caught Load
  • Slack Line
  • Overload
  • Stage Theater Rigging
  • Industries
  • Energy
  • Crane/Hoist
  • Stage Theater
  • Manufacturing
  • Metals

17
Load Link
Ensuring that its performance is unrestricted
install MSI Load Link at wire rope dead-end.
Route sensor cable so that it doesnt bind or
induce excessive bending forces on its
connector.
18
Load Link Applications
Dead-End Installation
Caught Load Monitoring Installation
19
Clamp on device training
  • Overview, Specs, Applications Installation
    Guidelines

20
Clamp On Device Overview
  • WRCS are low cost line tension sensors that fit a
    variety of cable sizes. This design does not
    require the user to cut any cables and
    create dead ends for load monitoring devices. 
    Instead, it simply clamps onto the middle of the
    wire rope and measures cable deflection to
    determine how much load is being applied.  
  • Benefits
  • Used in tension monitoring installations where a
    dead-end is not available.
  • Can attach to the middle of  any cable run.
  • Can be moved easily to other wire rope as
    required.
  • Designed to handle any capacity.
  • Not precision accurate and classified as a safety
    device rather than a weighing device.
  • Features
  • Accurate strain gauge measurement
  • Designed for all deadline tension, electronic
    sensing applications
  • Simple clamp-on design requires no deadline
    anchor
  • Environmentally sealed load cell operates in the
    harshest environments
  • Easy to use. Installs on wire rope from 3/4" to
    2" diameter
  • Custom engineering available 

21
Clamp On Device
Mounting Nuts
Clamp down plate
Install the wire rope Clamp-on Sensor at a
location that will not inhibit lifting height.
Note Once installed, the sensor shall remain
static while load is moving. To install position
wire rope in parallel with sensor, install clamp
down plate and tighten mounting nuts until wire
rope is firmly against the sensor center section.
22
Load Link training
  • Sensor Cable Interface for MSI-3550, MSI-9000,
    and Scoreboards

23
  • Sensor Interface to System Electronics
  • The Load Sensor and its cable assembly have
    either 4 or 6 conductors plus shield to interface
    to the System electronics. Two of these
    conductors are dedicated to Excitation and two
    for Signal. For long cable runs (over 100 ft.),
    six conductors are generally provided to detect
    voltage loss. With 6 conductor cables, the last
    two conductors are for Sense.
  • Excitation
  • - Excitation
  • Sense
  • - Sense
  • Signal
  • Sense
  • Refer to Sensor and Cable Assembly documentation
    for color designations.

24
MSI Model 3350 Digital Weight Indicator
Remove back cover and connect load sensor per
diagram below
25
MSI Model 3350 Digital Weight Indicator
When also installing a remote scoreboard, connect
its RS-232 cable to the input connector block
adjacent to the indicators Sensor input. The
MSI-3550 RS-232 TXD line connects to the
scoreboard RXD line and conversely MSI-3350s RDX
line connects to the scoreboard TXD. Also ensure
Ground is connected to (GND).
26
9000 CellScale Electronics Load Sensor
Connector Connections
27
9000 CellScale Electronics 9850 Digital
Indicator Scoreboards
If interfacing a Non-RF Remote Scoreboard, its
cable assembly will connect to either a Model
MSI-9000 CellScale or MSI-9850 Remote RF
Indicator. Please refer to appropriate User Guide
for conductor color-guide to determine RTX, DTX
and GND signals.   NOTE The TXD line connects
to the Scoreboard RXD line, the RDX line connect
to the TDX line in the Scoreboard and GND to GND
9000 User Manual Starting on Page 7
9850 User Manual Starting on Page 14
28
CALIBRATION CONSIDERATIONS
As the position of a hoist load block changes,
system accuracy is effected. This is principally
due to changes in weight caused by increasing or
decreasing load representing the wire rope feed.
As load is lowered or raised the weight of the
wire rope distributed to the application will
change and displayed as more or less load. To
minimize these effects, the system should be
calibrated with a load at a position equal to the
anticipated application weighing position. For
example, if application weight readings will be
typically recorded at a load height of 10 ft.,
the system calibration should be performed at the
same position. If the crane is required to weigh
at different heights, then the hoist load block
should be located in the center of its minimum
and maximum travel positions during calibration.
29
CALIBRATION SETUP
Use Calibration Setup to perform initial
calibration or to revise Factory settings.
There are two forms of Calibration 1) Initial
calibration 2) Standard Calibration MSI-3550
Calibration To properly calibrate a system,
test-weights of sufficient capacity are required.
While having test weights equaling the system
capacity will result in most accurate system
calibration, typically 50 of the rated capacity
is recommended. Note Except when entering
weight values, turning the System power off will
safely terminate Calibration Procedure.
30
3550 Digital Indicator SYSTEM INITIALIZE
Caution System Initialization is only required
when Indicator Circuit Boards are replaced or the
system capacity requires changing. Function Usual
ly used when boards are replaced, system
troubleshooting or sensor replacement. The
procedure clears the calibration settings, but
retains the factory feature settings. Action 1)
Turn off the scale by pressing the POWER key. 2)
Remove seal port on left side of indicators
front panel. Depress the switch button (via
hole with a probe) and simultaneously press the
POWER key. 3) The indicator will display the
user-prompt SureE? to confirm technician
wants to perform a complete system
initialization. 4) To proceed with system
initialization, press the TARE button. To abort
initialization, press the ZERO Button. Note
Aborting this procedure will not reset
system parameters.
31
ZERO functions as CLEAR/EXIT Use the ZERO Button
to exit the Setup menu without affecting system
settings or exiting Setup Mode. While entering
numeric values, you can use the ZERO Button to
clear an errantly entered digit. TARE functions
as ENTER For menu selection and choosing menu
items. USER functions as SELECT Used to scroll
through menus and if held down the Select will
repeat automatically. Use the POWER function
button to enter a decimal point. (i.e. Tare or
Set Point).
Calibration Instructions begin on page 24 of
MSI-3550 User Manual
32
3650/3750 Digital Indicator
Model MSI-3650 Calibration Instructions begin on
page 38 of its User Manual
Model MSI-3750 Calibration Instructions begin on
page 112 of its User Manual
33
Model 9000 CellScale Calibration
  • The MSI-9000 CellScale can be calibrated with
  • 1) A computer interface via MSI-9000s Com port 1
  • To calibrate with the MSI-9000, refer to page 74
    of its User Manual
  • 2) Remote RF interface with the MSI-3750CS,
    MSI-9850 or MSI-9750A indicators
  • To calibrate with the MSI-3750CS, refer to page
    77 of its User Manual
  • To calibrate with the MSI-9850, refer to page 92
    of its User Manual
  • To calibrate with the MSI-9750A, refer to page 72
    of its User Manual
  • 3) Computer interface using MSI CVM software
  • Refer to following slides

34
Model 9000 CellScale Calibration
Model MSI-9000 Calibration Instructions with
Computer with CVM Software
Step 1
After selecting Scale then CellScale then
Calibration The following menu will appear.
35
Step 2
Select the proper scale then press Start The
following menu will appear
Step 3
Select Units then press Continue the
following menu will appear
36
Step 4
Enter capacity of system then press Continue
the following menu will appear
Step 5
Select Resolution then press Continue the
following menu will appear
37
Stop 6
Enter calibration weight and then load scale with
weight. Press Continue and the following will
appear
If a multi-point calibration is desired repeat
step six.
38
When the last calibration weight/value has been
entered and accepted, press Save and the
following will appear
Step 8
Close the calibration window, unload the test
weights and the computer will state Calibration
is Complete
Write a Comment