Electromyography: Recording - PowerPoint PPT Presentation

1 / 22
About This Presentation
Title:

Electromyography: Recording

Description:

Electromyography: Recording D. Gordon E. Robertson, PhD, FCSB Biomechanics Laboratory, School of Human Kinetics, University of Ottawa, Ottawa, Canada – PowerPoint PPT presentation

Number of Views:63
Avg rating:3.0/5.0
Slides: 23
Provided by: DGERob
Category:

less

Transcript and Presenter's Notes

Title: Electromyography: Recording


1
ElectromyographyRecording
  • D. Gordon E. Robertson, PhD, FCSB
  • Biomechanics Laboratory,
  • School of Human Kinetics,
  • University of Ottawa, Ottawa, Canada

2
EMG Recording Topics
  • Surface or indwelling
  • Electrode placement
  • Type of amplifier
  • Common Mode Rejection Ratio (CMRR)
  • Dynamic range and Gain
  • Input impedance and skin resistance
  • Frequency response
  • Telemetry versus directly wired

3
Types of Electrodes
4
Surface Electrodes
  • lower frequency spectrum
  • (20 to 500 Hz)
  • relatively noninvasive, cabling does encumber
    subject, telemetry helps
  • skin preparation usually necessary
  • surface muscles only
  • global pickup (whole muscle)
  • inexpensive and easy to apply

5
Surface Electrodes
  • pre-gelled disposable electrodes are most common
    and inexpensive
  • MLS pre-amplified electrodes reduce movement
    artifact
  • Delsys Trigno includes 3D accelerometers

6
Indwelling Electrodes
  • fine wire or needle
  • localized pickup
  • difficult to insert
  • invasive, possible nerve injury
  • produces higher frequency
  • spectrum (10 to 2000 Hz)
  • can record deep muscles

7
Electrode Placement
  • electrode pairs in parallel with fibres
  • midway between motor point and myotendinous
    junction (or near belly of muscle)
  • approximately 2 cm apart, better if electrodes
    are fixed together to reduce relative movement

8
Surface Electrode Placement
9
Noise Reduction and Grounding
  • leads should be immobilized to skin
  • surgical webbing can help reduce movement
    artifacts
  • ground electrode placed over
  • electrically neutral area usually
  • bone
  • N.B. there should be only one ground electrode
    per person to prevent ground loops that could
    cause an electrical shock

10
Surface Electrode System(preamplifier type)
11
Type of Amplifier
  • because EMG signals are small (lt 5 mV) and
    external signals (radio, electrical cables,
    fluorescent lighting, television, etc.) are
    relatively large, EMG signals cannot be
    distinguished from background noise
  • background noise (hum) is a common mode signal
    (i.e., arrives at all electrodes simultaneously)
  • common mode signals can be removed by
    differential amplifiers
  • single-ended (SE) amplifiers may be used after
    differential preamplified electrodes

12
Common Mode Rejection Ratio (CMRR)
  • ability of a differential amplifier to perform
    accurate subtractions (attenuate common mode
    noise)
  • usually measured in decibels (y 20 log10 x)
  • EMG amplifiers should be gt80 dB (i.e., S/N of 10
    0001, the difference between two identical 1 mV
    sine waves would be 0.1 mV)
  • most modern EMG amplifiers are gt100 dB

13
Dynamic Range and Gain
  • dynamic range is the range of linear
    amplification of an electrical device
  • typical A/D computers use /10 V or /5 V
  • amplifiers usually have /10 V or more,
    oscilloscopes and multimeters /200 V or more
  • audio tape or minidisk recorders have /1.25 V
  • EMG signals must be amplified by usually 1000x or
    more but not too high to cause amplifier
    saturation (signal overload)
  • if too low, numerical resolution will comprised
    (too few significant digits)

14
Input Impedance
  • impedance is the combination of electrical
    resistance and capacitance
  • all devices must have a high input impedance to
    prevent loading of the input signal
  • if loading occurs the signal strength is reduced
  • typically amplifiers have a 1 MW (megohm) input
    resistance, EMG amplifiers need 10 MW or greater
  • 10 GW bioamplifiers need no skin preparation

15
Skin Impedance
  • dry skin provides insulation from static
    electricity, 9-V battery discharge, etc.
  • unprepared skin resistance can be 2 MW or greater
    except when wet or sweaty
  • if using electrodes with lt 1 GW input
    resistances, skin resistance should be reduced to
    lt 100 kW
  • Vinput Rinput / (Rinput Rskin) VEMG

16
Skin Impedance Example
  • Vinput Rinput / (Rinput Rskin) VEMG
  • If skin resistance is 2 MW (megohm) and input
    resistance is 10 MW then voltage at amplifier
    will be 10/(10 2) 0.833 83.3 of its true
    value.
  • By reducing skin resistance to 100 kW this can be
    improved to 99.
  • By also using a 100 MW resistance amplifier the
    signal will be 99.9.

17
Frequency Response
  • frequency responses of amplifier and recording
    systems must match frequency spectrum of the EMG
    signal
  • since raw surface EMGs have a frequency
    spectrum from 20 to 500 Hz, amplifiers and
    recording systems must have same frequency
    response or wider
  • since relative movements of electrodes cause low
    frequency artifacts, high-pass filtering is
    necessary (10 to 20 Hz cutoff)
  • since surface EMG signals only have frequencies
    as high as 500 Hz, low-pass filtering is
    desirable (500 to 1000 Hz cutoff)
  • therefore use a band-pass filter (e.g., 20 to
    500 Hz)

18
Frequency Response
  • Typical frequency spectrum of surface EMG

19
Typical Band Widths
EMG 20500 Hz 101000 Hz surface indwelling
ECG 0.0530 Hz 0.05100 Hz standard diagnostic
EEG 13 Hz 47 Hz 812 Hz 1230 Hz 30100 Hz delta waves theta waves alpha waves beta waves gamma waves
muscle forces or human movements DC10 Hz muscle moments joint trajectories
audio 208000 Hz 2015 000 Hz 2020 000 Hz voice tape CD
20
EMG Sampling Rate
  • since highest frequency in surface EMG signal is
    500 Hz, A/D (computer) sampling rates should be
    1000 Hz or greater (gt2 times maximum frequency)
  • raw EMGs cannot be correctly recorded by pen
    recorders since pen recorders are essentially 50
    Hz low-pass filters
  • mean or median frequencies of unfatigued muscles
    are around 70 to 80 Hz
  • notch filters should not be used to remove
    50/60 cycle (line frequency) interference because
    much of the EMG signal strength is in this range

21
Telemetry versus Direct Wire
  • telemetry has less encumbrance and permits
    greater movement volumes
  • radio telemetry can be affected by interference
    and external radio sources
  • radio telemetry may have limited range due to
    legislation (e.g., IC, FCC, CRTC)
  • cable telemetry (e.g., Bortec) can reduce
    interference from electrical sources
  • telemetry is usually more expensive than directly
    wired systems
  • telemetry has limited bandwidth (more channels
    reduce frequency bandwidths)

22
Telemetered EMG
  • Delsyss Trigno EMG and accelerometry telemetry
    system
Write a Comment
User Comments (0)
About PowerShow.com