Title: Anesthesia Machine
1Anesthesia Machine
-
- Presented by Gil Soto C.R.N.A
2Danger
3Lecture Outline
- The Machine
- Gas Supply Systems Hospital
pipeline Cylinder - High Pressure System (exposed to cylinder
pressure) - Intermediate Pressure System (exposed to pipeline
press) - Low Pressure System (distal to flowmeter needle
valve) - Circle System CO2 Absorber System
Unidirectional Valves - Ventilator
- Scavenger System
4Anesthesia Machine Checkout
- General
- Anesthesia Apparatus Checkout Recommendations,
1993 (Taken from the FDA) - This checkout, or a reasonable equivalent, should
be conducted before administration of anesthesia.
These recommendations are only valid for an
anesthesia system that conforms to current and
relevant standards and includes an ascending
bellows ventilator and at least the following
monitors capnograph, pulse oximeter, oxygen
analyzer, respiratory volume monitor (spirometer)
and breathing system pressure monitor with high
and low pressure alarms. This is a guideline
which users are encouraged to modify to
accommodate differences in equipment design and
variations in local clinical practice. Such local
modifications should have appropriate peer
review. Users should refer to the operator's
manual for the manufacturer's specific procedures
and precautions, especially the manufacturer's
low pressure leak test (step 5). - If an anesthesia provider uses the same machine
in successive cases, these steps need not be
repeated or may be abbreviated after the initial
checkout.
5Anesthesia Machine Checkout
- Steps 1-3
- Emergency Ventilation Equipment 1. Verify
Backup Ventilation Equipment is Available
Functioning - High Pressure System 2. Check Oxygen Cylinder
Supply a. Open 02 cylinder and verify at least
half full (about 1000 psi). b. Close cylinder.
3. Check Central Pipeline Supplies a. Check
that hoses are connected and pipeline gauges read
about 50 psi.
6Anesthesia Machine Checkout
- Steps 4-7
- Low Pressure Systems 4. Check Initial Status
of Low Pressure System a. Close flow control
valves and turn vaporizers off. b. Check fill
level and tighten vaporizers' filler caps. 5.
Perform Leak Check of Machine Low Pressure System
a. Verify that the machine master switch and
flow control valves are OFF. b. Attach
"Suction Bulb" to common Fresh gas outlet. c.
Squeeze bulb repeatedly until fully collapsed.
d. Verify bulb stays fully collapsed for at
least 10 seconds. e. Open one vaporizer at a
time and repeat 'c' and 'd' as above. f.
Remove suction bulb, and reconnect fresh gas
hose. 6. Turn On Machine Master Switch and all
other necessary electrical equipment. 7. Test
Flowmeters a. Adjust flow of all gases through
their full range, checking for smooth operation
of floats and undamaged flowtubes. b. Attempt
to create a hypoxic 02/N20 mixture and verify
correct changes in flow and/or alarm.
7Anesthesia Machine Checkout
- Scavenging System 8. Adjust and Check
Scavenging System a. Ensure proper connections
between the scavenging system and both APL
(pop-off) valve and ventilator relief valve.
b. Adjust waste gas vacuum (if possible).
c. Fully open APL valve and occlude Y-piece.
d. With minimum 02 flow, allow scavenger
reservoir bag to collapse completely and verify
that absorber pressure gauge reads about zero.
e. With the 02 flush activated allow the
scavenger reservoir bag to distend fully, and
then verify that absorber pressure gauge reads
lt10 cm H20.
8Anesthesia Machine Checkout
- Breathing System 9. Calibrate 02 Monitor a.
Ensure monitor reads 21 in room air. b.
Verify low 02 alarm is enabled and functioning.
c. Reinstall sensor in circuit and flush
breathing system with 02. d. Verify that
monitor now reads greater than 90. 10. Check
Initial Status of Breathing System a. Set
selector switch to "Bag" mode. b. Check that
breathing circuit is complete, undamaged and
unobstructed. c. Verify that C02 absorbent is
adequate. d. Install breathing circuit
accessory equipment (e.g. humidifier, PEEP valve)
to be used during the case. 11. Perform Leak
Check of the Breathing System a. Set all gas
flows to zero (or minimum). b. Close APL
(pop-off) valve and occlude Y-piece. c.
Pressurize breathing system to about 30 cm H20
with 02 flush. d. Ensure that pressure
remains fixed for at least 10 seconds. e.
Open APL (Pop-off) valve and ensure that pressure
decreases.
9Anesthesia Machine Checkout
- Manual and Automatic Ventilation Systems 12.
Test Ventilation Systems and Unidirectional
Valves a. Place a second breathing bag on
Y-piece. b. Set appropriate ventilator
parameters for next patient. c. Switch to
automatic ventilation (Ventilator) mode. d.
Fill bellows and breathing bag with 02 flush and
then turn ventilator ON. e. Set 02 flow to
minimum, other gas flows to zero. f. Verify
that during inspiration bellows delivers
appropriate tidal volume and that during
expiration bellows fills completely. g. Set
fresh gas flow to about 5 L/min. h. Verify
that the ventilator bellows and simulated lungs
fill and empty appropriately without sustained
pressure at end expiration. i. Check for
proper action of unidirectional valves. j.
Exercise breathing circuit accessories to ensure
proper function. k. Turn ventilator OFF and
switch to manual ventilation (Bag/APL) mode.
l. Ventilate manually and assure inflation and
deflation of artificial lungs and appropriate
feel of system resistance and compliance. m.
Remove second breathing bag from Y-piece.
10Anesthesia Machine Checkout
- Monitors 13. Check, Calibrate and/or Set Alarm
Limits of all Monitors Capnometer, Pulse
Oximeter, Oxygen Analyzer, Respiratory Volume
Monitor (Spirometer), Pressure Monitor with High
and Low Airway Alarms - Final Position 14. Check Final Status of
Machine a. Vaporizers off b. AFL valve
open c. Selector switch to "Bag" d. All
flowmeters to zero e. Patient suction level
adequate f. Breathing system ready to use
11The Anesthesia Machine
- The anesthesia gas machine is a device which
delivers a precisely-known but variable gas
mixture, including anesthetizing and
life-sustaining gases.
12The MachineOhmeda N.A.Drager (Narkomed)
13 Anesthesia Machine Jackson Memorial Hospital
14Manufacturers Names
- North American Dräger (Telford, PA) is the
manufacturer of the Narkomed 2C, Narkomed 4,
Narkomed GS, Narkomed 6000, Narkomed Julian,
Narkomed MRI and Narkomed Mobile models. - Datex-Ohmeda (Madison WI) manufactures the AS/3
ADU, Aestiva, Modulus SE, Excel 210, and Excel
110
15Some Numbers to Remember
- The hospital pipeline is the primary gas source
at 50 psi (normal working pressure of most
machines). - Cylinders O2 is supplied at around 2000 psi
(regulated to approximately 45 psi after it
enters the machine). - Oxygen flush is a "straight shot" from supply to
delivery point, 35-75 L/min. - OSHA Fact Sheet (1991) on Waste Anesthetic Gases
(WAGs) occupational exposure should be limited to
an eight hour time-weighted average of not more
than 2 ppm halogenated agents (Halothane,
Enflurane, Isoflurane, Sevoflurane, Desflurane) - If Halogenated agent is used in combination with
nitrous oxide, then ONLY 0.5 ppm OF THE
HALOGENATED AGENT IS ALLOWED - No more than 25 ppm nitrous oxide can be used at
all times (with or without Halogenated Agent)
16Minimal Components
- O2 Pipeline N2O Pipeline
-
- O2 Flowmeter N2O Flowmeter
- Container with VAA
- Bag-valve-mask device
- Patient
17Straight-line model SPDD (Supply/Processing/Deliv
ery/Disposal)
18Oxygen has five "tasks
- It powers the ventilator driving gas
- O2 flush
- Activation of low pressure alarms
- Activation of fail-safe mechanisms (O2
pressure sensor shut-off ) - Proceeding through the flowmeter
19Other gases One task Only
- Transported via flowmeter breathing circuit to
- Anesthetize pt (N2O)
- Sustain Life (Air)
20(No Transcript)
21Basic Schematics
22Gas Supply SystemsHospital Pipeline
23DISS
- Pipeline inlets are connected with DISS (diameter
index safety system) non-interchangeable
connections. - The check valve, located down stream from the
pipeline inlet, prevents reverse flow of gases
(from machine to pipeline, or to atmosphere),
which allows use of the gas machine when pipeline
gas sources are unavailable.
24PISS
PISS (pin-index safety system) prevents
misconnection of a cylinder to the wrong yoke.
Keep cylinders closed except when checking
machine, or while in use (if O2 from pipeline is
unavailable)
25(No Transcript)
26Gas Supply SystemsCylinder
Pin Index Safety System
O2 2,5
N2O 3,5
27High Pressure System(parts which receive gas at
cylinder pressure)
- hanger yoke (including filter and unidirectional
valve) - yoke block (with check valves)
- cylinder pressure gauge
- cylinder pressure regulators
28Bourdon Gauge
29Hanger Yoke Check Valve
- Hanger Yoke
- orients cylinders
- provides unidirectional flow
- ensures gas-tight seal.
- Check Valve
- minimize trans-filling
- allows change of cylinders during use
- minimize leaks to atmosphere if a yoke is empty.
30Check Valve
31(No Transcript)
32More on Cylinders
- The cylinder pressure regulator converts high,
variable cylinder pressure to a constant pressure
of approximately 45 psi downstream of the
regulator. - This is intentionally slightly less than pipeline
pressure, to prevent silent depletion of cylinder
contents if a cylinder is inadvertently left open
after checking its pressure. - Cylinder pressure gauge indicates pressure in the
higher-pressure cylinder only (if two are opened
simultaneously).
33E cylinder Characteristics
- Gas US (International) PSI
Capacity (L) PISS - O2 Green (white)
1900 660 2-5 - N2O Blue (blue) 745
1590 3-5 - Air Yellow (B W)
1900 625 1-5 - Well use 2000psi for O2 instead of
1900psi
34Intermediate Pressure System
Hospital Pipeline Outlets
Machine piping guts
Hospital Pipeline Inlets
Gauges-pipeline (intermediate press. )
35Intermediate Pressure System
- (receives gases at low, relatively constant
pressures (37-55 psi, pipeline pressure) - (For consistency well use 50 psi)
- pipeline inlets and pressure gauges
- ventilator power inlet
- Oxygen pressure-failure device (fail-safe) and
alarm - flowmeter valves
- oxygen second-stage regulator
- oxygen flush valve
36Oxygen pressure-failure device (fail-safe) and
alarm
- What happens if you lose oxygen pipeline
pressure? - The fail safe device ensures that "Whenever
oxygen pressure is reduced and until flow ceases,
the set oxygen concentration shall not decrease
at the common gas outlet" (from ASTM F1161). - The loss of oxygen pressure results in alarms,
audible and visible, at 30 psi pipeline pressure.
- Fail-safe systems don't prevent hypoxic mixtures.
37Fail-safe systems don't prevent hypoxic mixtures
- as long as there is pressure in the O2 line,
nothing in the fail safe system prevents you from
turning on a gas mixture of 100 nitrous oxide
(however, this should be prevented by the hypoxic
guard system) - or 100 helium (which wouldnt be prevented by
the hypoxic guard). - Datex-Ohmeda terms their fail safe a "pressure
sensor shut off valve"- at 20 psi oxygen, the
flow of all other gases are shut off. Dräger's,
"oxygen failure protection device" (OFPD)
threshold is proportional, unlike Ohmeda's which
is off-or-on.
38Fail-safe systems don't prevent hypoxic mixtures
(Cont)
- Ohmeda uses a second-stage O2 pressure regulator
(ensures constant oxygen flowmeter input until
supply pressure is less than 12-16 psi). The
oxygen ratio monitor controller (ORM newer or
ORMC, both by Dräger) shuts off nitrous oxide
when oxygen pressure is less than 10 psi
39 Pipeline Trouble
- Pipeline sources are not trouble free
contamination (particles, bacteria, viral,
moisture), inadequate pressure, excessive
pressures, and accidental crossover (switch
between oxygen and some other gas such as nitrous
oxide or nitrogen) are all reported.
40What if you lose oxygen pipeline pressure?
- Open the emergency oxygen cylinder fully (not
just the three or four turns used for checking) - Disconnect the pipeline connection at the wall
- Why? Something is wrong with the oxygen pipeline.
- What if the supply problem evolves into a
non-oxygen gas in the oxygen pipeline? If so, it
will flow to the patient (pipeline pressure 50
psi) rather than your oxygen cylinder source
(down-regulated to 45 psi). - If you are lucky, the oxygen alarm will sound to
warn you of the change (you do set your alarms,
don't you?). - If for some reason the oxygen analyzer does not
warn of the crossover, the pulse oximeter will-
but only after the oxygen has been washed out by
ventilation from the patient's functional
residual capacity and vessel-rich group.
41Reinforcement!!!!
- Disconnect the pipeline connection at the wall if
oxygen pipeline pressure is lost. It's also
easier to remember one strategy which works for
any problem with the pipeline, rather than to
remember that sometimes you must, and sometimes
it is optional, to disconnect. And use that
oxygen analyzer always! - Ventilate by hand rather than with the mechanical
ventilator (which uses cylinder oxygen for the
driving gas if the pipeline is unavailable)
42HOW LONG BEFORE O2 TANK IS EXHAUSTED???
- -The time to exhaustion is calculated by
dividing the remaining O2 volume in the cylinder
by the rate of consumption of O2. - -Remaining volume in liters (L) in an E-cylinder
is calculated by dividing the cylinder pressure
in psig by 2000 psig and multiplying by 660 L.
43EXAMPLE
- If cylinder gauge reads 1,000 psig, this
represents (1000/2000) X 660 330 L left in that
tank. The rate of consumption of O2 during
mechanical ventilation is the sum of the O2 flow
meter setting and the patients minute
ventilation (VT in L x RR in breaths/min). - If FGF is 0.5 L/min O2 and 1.0 L/min N2O and VT
is 0.7 L and RR is 10 bpm, then the minute
ventilation is - 7 L/min (0.7L x 10 bpm)
- The total O2 consumption is 7.5 L/min. The
expected time to exhaustion is thus approximately
330 L divided by 7.5 L/min 44 min (ignoring the
gas sampled by the gas analyzer and leaks)
44The Low-pressure system(distal to flowmeter
needle valve)
- flowmeter tubes
- vaporizers
- check valves (if present)
- common gas outlet
45Flowmeters
-Thorpe tube is an older term for flowmeters.
-Components needle valve, indicator float,
knobs, valve stops. -Flow increases when the
knob is turned counterclockwise (same as
vaporizers). -At low flows, the annular-shaped
orifice around the float is (relatively) tubular
so (according to Poiseuille's Law) flow is
governed by viscosity. (laminar flow) -At high
flows (indicated on the wider top part of the
float tube), the annular opening is more like an
orifice, and density governs flows. (turbulent
flow)
46Low Pressure System
- Distal to Flowmeter Needdle Valve
- Flow Meters- measures and indicates the
rate of gas flowing through it. Variable
orifice/Thorpe tube-constant press. flow meters.
Rate of
flow r/t 1) pressure drop across the
constriction 2) size of annular
opening
3)
Physical properties of the gas
(viscosity and
density) Indicator,
float or bobbin- 1) rotometers
2) non-rotating floats
3) ball floats Sequence of
flowmeters tubes is very important to
decrease chance of hypoxic mixture.,
Gas flow is from left to right, O2 on
right side. Any leak in flowmeters will
vent other gas out or - entrain air
before O2 is added to gas mixture decreasing
chance that O2
will be lost or diluted.
FLOW
47More on Flowmeters
- Needle valve can be damaged if it is closed with
force - Flowtube (Thorpe tube) is tapered (narrower at
bottom) and gas-specific - If gas has 2 tubes, they are connected in series
with a single control valve
48Did anyone say Flowmeters??
- Care of flowmeters includes ensuring that
- floats spin freely
- qualified service personnel regularly maintain
gas machines - an O2 analyzer used always (of course, the
readings are erroneous during use of nasal
cannula) - one never adjusts a flowmeter without looking at
it - one includes flowmeters in visual monitoring
sweeps - one turns flowmeters off before opening
cylinders, connecting pipelines, or turning
machine "on".
49Low Pressure System
- Safety Devices-Purpose is to decrease risk of
hypoxic mixture Mandatory Minimum O2 Flow-
factory preset minimum O2 flow that always flows
when machine is on. Minimum O2/N2O Ratio
13 Device or proportioning system Flow
valves linked mechanically or pneumatically so O2
cannot be set below 25. Alarm will signal if
O2/NO2 ratio falls below preset value - O2/NO2 Proportioning Device-Automatically
mixes O2 and NO2 to setting selected on dial
50(No Transcript)
51(No Transcript)
52(No Transcript)
53Hypoxic breathing is POSSIBLE
- hypoxic guard systems CAN permit hypoxic
breathing mixtures IF - wrong supply gas in oxygen pipeline or cylinder,
- defective pneumatic or mechanical components,
- leaks exist downstream of flow control valves, or
- if third inert gas (such as helium) is used.
54Low Pressure System
- Vaporizers- Classification A. Method of
regulating output concentration 1.
Concentration calibrated 2. Measured
flow B. Method of vaporization 1.
Flow over 2. Bubble Through
3. Injection C. Temperature compensation
1. Thermocompensation 2.
Supplied heat D. Specificity 1.
Agent specific 2. Multiple agent
E. Resistance 1. Plenum
2. Low resistance
55VAPORIZERS
- Vapor Pressure (VP) Molecules escape from a
volatile liquid to the vapor phase, creating a
saturated vapor pressure at equilibrium - VP is independent of Atmospheric Press
- VP increases with Temperature
- VP depends ONLY on the Physical Characteristics
of the Liquid on its Temperature
56CLASSIFICATION
- Variable bypass
- Fresh gas flow from the flowmeters enters the
inlet of any vaporizer which is on. The
concentration control dial setting splits this
stream into bypass gas (which does not enter the
vaporizing chamber), and carrier gas (also called
chamber flow, which flows over the liquid agent)
57CLASSIFICATION
- Flow over
- Carrier gas flows over the surface of the liquid
volatile agent in the vaporizing chamber (as
opposed to bubbling up through it (as in the
copper kettle and Vernitrol)
58CLASSIFICATION
- Temperature compensated
- Equipped with automatic devices that ensure
steady vaporizer output over a wide range of
ambient temperatures - Agent-specific
- Only calibrated for a single gas, usually with
keyed fillers that decrease the likelihood of
filling the vaporizer with the wrong agent - Out of circuit
- As opposed to (much) older models such as the
Ohio 8 (Boyle's bottle) which were inserted
within the circle system.
59Vaporizer Interlock Mechanism
Safety mechanism that allows ONLY one vaporizer
at a time to be opened
60Circle System
Circle System- CO2 absorber housing and absorber,
unidirectional valves, inspiratory and expiratory
ports, fresh gas inlet, APL valve, pressure
gauge, breathing tubes, Y-piece, reservoir bag,
bag/vent switch selector, respiratory gas monitor
sensor.
61Circle System
- CO2 Absorber System Housing (canister support),
Absorbent, baffles, side tube - Unidirectional Valves-aka Flutter valves, one way
valves, check valves, directional valves, dome
valves - Canister-Air space 50, void space 42, pore
space 8 - Soda Lime 4 Sodium Hydroxide, 1 potassium
hydroxide, 14-19 H2O, and calcium hydroxide to
make 100, Silica and kielselguhr for
hardness
Indicator for color change with exhaustion of
CO2 absorption capabilities CO2H2O?H2CO3 2
NaOH2H2CO3Ca(OH)2 ?CaCO3NaCO34H2O heat
released 13,700 cal./mole CO2 absorbed - Barium Hydroxide Lime 20 Barium hydroxide, 80
calcium hydroxide, and /-
potassium hydroxide, Indicator for color
change with exhaustion of CO2 absorption
capabilities Ba(OH)2 . 8H2OCO2?BaCO39H2O 9H
2O9CO2 ?9H2CO3 9H2CO39Ca(OH) 2 ?
9CaCO318H2O 2KOHH2CO3 ? K2CO32H2O Ca(OH
)2K2CO3 ? CaCO32KOH Regeneration (color
change loss) with rest can occur. Appears new but
is exhausted Granule size 4-8 mesh- 4 mesh equals
strainer with 4 openings/inch
62Circle system CO2 Absorber System
Removing canister soda lime
canisters unlocked
canister locking lever
Removing both canisters soda lime
Replacing fresh soda lime
Exhausted soda lime
63Circle system Unidirectional Valves
- Unidirectional valves-aka flutter valves, one way
valves, check valves, directional valves, dome
valves. Found on Inspiratory and Expiratory
flow ports
Narkomed Machine
Ohmeda Machine
64Ventilator
- Ventilator Components Driving gas supply,
injector, controls, alarms, safety-release valve,
bellows assembly, exhaust valve, spill valve,
connection for ventilator hose
Bellows assembly
Ventilator controls
65Ventilator
- Driving gas supply or power gas supply-O2
pneumatically drives (compresses) ventilator
bellows - Injector or Venturi mechanism-Increases the flow
of driving gas by using the BERNOULLI Principle-
As a gas flow meets a restriction, its lateral
pressure drops. Any opening in the tube at this
constriction will entrain air (suck air in) - Controls-Adjusts Flow, Volume, Timing, and
Pressure of the driving gas that compresses the
bellows Pneumatic-Uses pressure changes to
initiate changes in respiratory cycle Fluidic or
fluid logic-Uses gas streams through channels in
solid material. Allow for compact
ventilator Electronic-Electronic control of
many addition ventilation parameters powered by
a driving gas on newer machines. Must have both
power and pnuematics. - Alarms-ASTM standards group alarms into three
levels High, Medium, Low Priority correlates
tooperator immediate action, prompt action,or
awareness. Loss of main power is the only
required alarm with a required duration of at
least 2 minutes - Safety relief valve-aka pressure limiting valve,
drving gas pressure relief valve. Vents driving
gas if factory pre-set pressure is reached (65-80
cm H2O) or adjustable set pressure is reached.
66Bernoullis Principle
- At constriction
- Flow is higher
- Pressure is lower
67Ventilator
- Bellows Assembly Housing- Usually made of
hard rigid clear plastic Bellows
ASCENDING-standing, upright. Compressed downward
during inspiration. ASCEND DURING EXPIRATION - Pressure is always positive. PEEP 2-4 cm
H2O. DESCENDING-hanging, inverted. Compressed
upward during inspiration. DESCEND DURING
EXPIRATION. Weight of bellows results in negative
airway pressure during exhalation until bellow
refilled. -
IMPORTANT difference between ascending and
descending is that when there is a major leak
or disconnect, the ascending bellows will
collapse (unless prevented by scavenging system).
When a disconnection occurs with a descending
bellows system, the ventilator will continue
its upward movement and downward movements,
drawing in room air and driving gas during its
descent and discharging it during the upward
movement. Gas flow during upward movement may
generate enough pressure such that the low
pressure alarm is not activated.
What type is shown?
Remember that the type is described by how the
bellows move during EXPIRATION
68Scavenger System
- Scavenger System consists of 1) gas collecting
assembly, 2) a transfer means, 3) the interface,
4) gas disposal tubing, 5) gas disposal assembly.
(some or all components may be combined). AS
TM standard fitting size for scavenger hoses 19
mm ( international standard 30mm) to prevent
incorrect connection to breathing hoses (22mm).
3
2
1
45
45
1
69REFERENCES
- NZ pg 247-252
- MM pg 35-49
- DD pg 3-74
- http//chico.med.yale.edu/machine/agmpart1.htmGen
eral20features - http//www.anest.ufl.edu/eduweb/vam/
70 Unknown
The essence of intelligence is skill in
extracting meaning from
everyday experience
71Anesthesia Machine
- What is the PISS numbers for O2, for N2O?
- What is the line pressure for O2 for N2O?
- What is the tank pressure for O2 for N2O?
72Anesthesia Machine
- The O2 pressure alarm goes off, you should
- It is not possible to deliver a hypoxic mixture.