Title: Myocardial contrast echocardiography in routine clinical practice
1Myocardial contrast echocardiography in routine
clinical practice
2WHY ?
3Clinical and economic outcomes assessment with
MCE, Leslee et al, Heart 1999.
4- Principles of Contrast echocardiography
- Blood appears black on conventional two
dimensional echocardiography, not because blood
produces no echo, but because the ultrasound
scattered by red blood cells at conventional
imaging frequencies is very weakseveral thousand
times weaker than myocardiumand so lies below
the displayed dynamic range
5- Contrast ultrasound results from scattering of
incident ultrasound at a gas/liquid interface,
increasing the strength of returning signal. - However, the bubble/ultrasound interaction is
complex. - Understanding this interaction is key to
performing, understanding, and interpreting a
contrast echo study.
6- Physics in ultrasound
- Unlike solid tissue, gas bubbles have acoustic
properties that vary with the strength of the
insonating signal. When insonated ----gt gas
bubbles pulsate-----gt with compression at the
peak of the ultrasound wave and expansion at the
nadir. -
- An ultrasound beam with a frequency of 3 MHz,
will result in bubble oscillation three million
times per second AND hence are several million
times more effective at scattering sound than red
blood cells. With this movement, sound is
generated and, combined with that of thousands of
other bubbles, results in the scattered echo from
the contrast agent. - Characterising this echo, from that of tissue,
form the basis of contrast echocardiographic
studies.
7- It is a remarkable coincidence that gas bubbles
of a size required to cross the pulmonary
capillary vascular bed (15 mm) resonate in a
frequency range of 1.57 MHz, precisely that used
in diagnostic ultrasound.
8- At low poweroutput (PO) settings, there is mostly
a linear response (fundamental enhancement) with
some generation of harmonic frequencies. - As the PO is increased, the bubbles generate more
nonlinear resonance and thus generate greater
harmonic frequencies. - At a high power setting, fracture and destruction
of the microbubble occur, allowing the air or gas
inside to be released.
(lt100 kPa)
(100 kPa1 Mpa)
9- Contrast imaging requires ultrasound machine
settings to be optimised. Generally, this
requires variation in the system power output,
indicated on clinical systems as the mechanical
index (MI) (most important parameter) The MI is a
unit-less number that serves as an indicator of
the nonthermal bioeffects. - This is an estimate of the peak negative pressure
within insonated tissue defined as the peak
negative pressure divided by the square root of
the ultrasound frequency. - Display of the MI was made mandatory in the USA
as a safety measure, to enable an estimate of the
tissue effects of ultrasound exposure to be made.
As this also reflects the mechanical effect of
ultrasound on a contrast bubble, this has proved
useful in developing machine settings for
contrast ultrasound.
10- Standard clinical echocardiography imaging
utilises an MI of around 1.0, but a lower
setting, around 0.5, is usually optimal for left
ventricular opacification. - To achieve myocardial perfusion imaging the
extremes of power output are utilised - high MI (gt 1.2) is used to achieve bubble
destruction in power Doppler imaging, and - ultra low MI (lt 0.1) required to induce linear
oscillation of microbubbles required for real
time myocardial perfusion imaging.
11(No Transcript)
12Current generation of contrast agents comprises
small, stabilized, gas-filled microbubbles that
can pass through the smallest capillaries.
- Current microbubbles can be visualized within the
left heart chambers after an intravenous
injection for two reasons - (1) they are smaller than red blood cells
and so can pass through the pulmonary
capillaries, and - (2) they are stable, relative to time and
pressure (they can persist long enough to travel
through the pulmonary capillary bed, and they can
withstand left-sided pressures).
The larger the microbubbles, the better the
contrast effect
13- The properties of the Air- filled microbubble are
- - a strong reflector of ultrasound and highly
soluble - But has low persistence and lacks stability
because air diffuses out. This shrinking
microbubble becomes less and less reflective. - WHEREAS
- Gases are of high molecular weight are-
- not very soluble hence stable with longer
persistence. - Persistence prolongs contrast effect so that
assessment of left ventricular borders and wall
motion can be made.
14(No Transcript)
15- The ultrasonic characteristics depends on -
- a) size of the bubble
- b) composition of the shell and the gas in the
shell. - The outer shell of the microbubbles is composed
of many different substances, including - albumin, polymers, palmitic acid,or
phospholipids. - This composition determines its elasticity, its
behavior in an ultrasonic field, and the methods
for metabolism and elimination. - In general, the stiffer the shell, the more
easily it will crack or break with ultrasonic
energy. Conversely, the more elastic the shell,
the greater its ability to be compressed or
resonated and to produce a nonlinear backscatter
signature.
16- The major technical difficulties in using
contrast agents for left ventricular
opacification are - the need to resonate but not burst microbubbles
and - to maintain adequate bubble concentration within
the cavity. - injection of a sufficient concentration of
microbubbles and - the proper ultrasonographic equipment settings
to optimize image quality. - Inadequate bubble concentration, causes less
opacification of the chamber, which was a
difficulty with first-generation agents because
of excessive bubble destruction when insonified
with ultrasound. - Left ventricular opacification requires a long
duration of the contrast effect for the
assessment of left ventricular borders in
suboptimal patients both in rest and under
stress. - Contrast agents that are easily destroyed will be
useful in myocardial perfusion studies.
17- CONTRAST AGENTS FOR ULTRASOUND
- Initially contrast echocardiography utilised free
air in solution but these large, unstable bubbles
were not capable of crossing the pulmonary
vascular capillary bed, allowing right heart
contrast effects only. - The first agents capable of left heart contrast
after intravenous injection (first generation
agents) were air bubbles stabilised by
encapsulation (Albunex) or by adherence to
microparticles (Levovist). - Replacing air with a low solubility fluorocarbon
gas stabilises bubbles still further (second
generation agentsfor example, Optison, SonoVue),
greatly increasing the duration of the contrast
effect. - Third generation agentsnot yet commercially
availablewill use polymer shell and low
solubility gas and should producemuch more
reproducible acoustic properties.
18- PERFORMING A CONTRAST STUDY
- Bubbles are prone to destruction by physical
pressure, hence requires meticulous attention to
the preparation and administration. - The agent should be prepared immediately before
injection and vents used when withdrawing the
agent into the syringe. Bubbles tend to float
towards the surface and the contrast vial or
syringe should be gently agitated each time fresh
contrast administration is required. - Injection through a small lumen catheter
increases bubble destructiona 20 G or greater
cannula should be used. - Very small volumes of contrast are needed using
second generation agents (lt 1 ml) and a flush is
required by using a three way tap,with contrast
injected along the direct path to minimise bubble
destruction, and saline flush injected into the
right angle bend. - For myocardial perfusion work, infusion produces
more reproducible results with the potential for
quantification6 but brings its own problems.
Bubbles in agents currently available are buoyant
and will tend to rise to the surface of the
syringe. - Purpose designed infusion pumps which agitate
contrast continuously are in development but not
yet widely available.
19(No Transcript)
20(No Transcript)
21- A PFO is diagnosed if more than three
microbubbles pass from right to left atrium
within three cardiac cycles of right atrial
opacification. -
- Crude quantification is possible with
- a small shunt defined as 310 bubbles,
- a medium shunt 1020, and
- a large shunt gt 20 bubbles.
- An initial study should be performed during
normal respiration, when the normal reversal of
atrial pressure gradient in early systole may be
sufficient to allow shunting if a large defect is
present.
22(No Transcript)
23(No Transcript)
24LEFT VENTRICULAR OPACIFICATION
- Assessment of left ventricular (LV) systolic
function is the most common indication for
echocardiography. Accurate assessment and
quantification is dependent on visualising the
entire endocardium in cross section. - Tissue harmonic imaging has greatly improved
image quality and reduced the number of
non-diagnostic studies. - Contrast opacification of the left ventricular
cavity enhances endocardial border definition and
has been shown to increase diagnostic accuracy in
suboptimal studies at rest and during stress. - This is the major clinical use of left heart
contrast echocardiography at present
25- Contrast echo is useful in the assessment of LV
function in patients ventilated in intensive
care, reducing the time required to obtain
diagnostic information and obviating the need for
trans-oesophageal echocardiography. - While apical imaging is greatly enhanced,
para-sternal views may deteriorate, at least
initially,with contrast in the right ventricle
attenuating visualisation of the left. For this
reason, apical views should routinely be obtained
first in any contrast study. - LV opacification is also used to delineate LV
anatomy, particularly apically, confirming
pseudo-aneurysm, apical hypertrophy or
ventricular non- compaction and demonstrating
filling defects, typically apical thrombus.
26- MYOCARDIAL PERFUSION IMAGING
- Myocardial contrast echocardiography (MCE) is
the imaging of a contrast agent within the
myocardial capillary vascular bed is
reproducible, real time, noninvasive assessment
of myocardial perfusion during rest and stress,
at the bedside and in the cardiac catheterisation
laboratory. - At present, MCE remains difficult and suboptimal
imaging prohibits routine use. - While LV cavity opacification with contrast can
make a difficult study diagnostic, only those
with high quality baseline B mode images are
suitable for MCE with current equipment. - Intravenous injection of contrast results in
very low concentration of bubbles in the
myocardium, with only 510 of cardiac output
entering the coronary circulation. As more than
90 of intramyocardial blood volume is within the
capillary compartment, contrast bubbles are
imaged at low velocity with slow replenishment
following bubble destruction.
27- Dissolution and destruction of microbubbles by
both intramural pressure and ultrasound exposure
further limits any contrast effect. Thus,
contrast specific imaging modalities must be
used. Broadly, there are two approaches used to
overcome the problems inherent in myocardial
contrast perfusion imaging intermittent imaging
and pulse inversion or power modulation imaging. - Flash imaging, utilising a pulse of echo at high
MI to destroy all microbubbles within the
myocardium, combined with low MI real time MCE
allows assessment of perfusion in real time .
This technique can be used to quantify rate of
bubble replenishment in the myocardium
28(No Transcript)
29(No Transcript)
30(No Transcript)
31LVO PROTOCOLS FOR A BOLUS INJECTION OFCONTRAST
32(No Transcript)
33(No Transcript)
34- New possibilities have emerged with
second-generation contrast agents containing high
molecular- weight gases such as fluorocarbons
(Optison, MBI/Mallinckrodt Sonazoid, Nycomed
Amersham, Amersham, Bucks, U.K. and EchoGen,
Sonus/Abbott, Seattle, WA, U.S.A.) or sulphur
hexafluoride (SonoVue, Bracco, Milan, Italy). - In a phase III multi centre trial that compared a
first generation (Albunex) with a second
generation (EchoGen) contrast agent demonstrated
the superiority of fluorocarbon-containing
microbubbles over sonicated human albumin for LV
cavity opacification, endocardial border
definition, duration of effect, salvage of
suboptimal echocardiograms, diagnostic confidence
and potential to influence patient management.
35- Reilly et al demonstrated that a second
generation contrast agent could significantly
improve the quality of echocardiograms in the
intensive care unit. They evaluated wall motion
analysis with standard fundamental
echocardiography, harmonic echocardiography and
contrast echocardiography. Wall motion was seen
with greater confidence after contrast
echocardiography. - Another study, Kornbluth et al. in mechanically
ventilated patients, showed the use of contrast
was superior to tissue harmonic imaging for
endocardial border delineation, wall motion
scoring and quantification of ejection fraction.
36- In Stress echocardiography a sizeable proportion
of patients can benefit from the additional use
of contrast media to improve endocardial border
detection. - At peak exercise image quality is often low
because of tachycardia. Only when endocardial
borders are well delineated and reproducible can
useful results be expected with stress
echocardiography. - The use of contrast during stress
echocardiography in patients with suboptimal
imaging may have a positive impact on cost it
can indeed enhance diagnostic confidence in
stress echocardiography, with improvement in
image quality in approximately 50 of patients. - In a group of poorly echogenic patients
undergoing dopamine echocardiography,Voci et
al.30 demonstrated that infusion of insonicated
albumin provided adequate LV opacification in 90
of patients, with significant reduction in
inter-observer variability in ejection fraction
measurements.
37-
- At present, the best solution for improving
endocardial border visualization is a combination
of intravenous contrast LV opacification and
harmonic imaging. - In a review of 200 patients referred for stress
echocardiography, a combination of native tissue
harmonic imaging and contrast application was
used. The combination resulted in significantly
better visualization of the endocardium as
compared with fundamental contrast imaging, and
an increased inter-observer agreement for border
detection, which increased from 83 in
fundamental mode without contrast to 95 with
contrast native tissue harmonic imaging.
38- Contrast stress echocardiography, despite the
added cost of the agent, can still provide a cost
saving of 15, and accuracy and outcome
assessment are similar to those of nuclear
techniques. - For conventional stress echocardiography, recent
reports have documented that its diagnostic
accuracy is similar to that of myocardial
perfusion evaluation using nuclear techniques. - Currently, if the patient does not have a good
acoustic window then stress echocardiography
should not be performed without contrast
enhancement. - The maximum yield of contrast enhancement has
been demonstrated when approximately two to six
segments are not satisfactorily delineated.
39Comparison of myocardial contrastechocardiography
with single-photon emissioncomputed tomography
- MCE may be compared with single-photon emission
computed tomography (SPECT). - Results of the earliest studies that compared MCE
with SPECT were encouraging. However, larger
clinical trials have shown that MCE, when
compared with SPECT, has a good specificity but
relatively low sensitivity. The question of
training is important. - The study conducted by Marwick et al., which
involved centres for which MCE was not initially
a routine investigation, showed that performance
and interpretation of perfusion echocardiography
requires special expertise. - The recent introduction of new techniques such as
harmonic power Doppler resulted in good
concordance between MCE and SPECT. - A study conducted by Rocchi et al revealed
excellent sensitivity and specificity (82 and
95, respectively) in detecting viable myocardium
in segments supplied by infarct-related arteries,
but more studies are needed to establish the
value of MCE.
40No-reflow phenomenon
- Restoration of epicardial flow in the
infarct-related artery, as indicated by
Thrombolysis in Myocardial Infarction (TIMI)
grade 3 flow, does not necessarily imply normal
flow at the level of the microcirculation. The
no-reflow phenomenon was initially observed in
the 1970s by Kloner et al and was subsequently
confirmed clinically by MCE, which can define the
presence and extent of flow at the level of the
microcirculation. - In the 39 patients with acute myocardial
infarction (AMI) before and immediately after
successful coronary recanalization studied by Ito
et al. up to 23 of patients showed lack of
microvascular reperfusion (or no-reflow)
despite a fully patent infarct-related artery. - In another study of 86 patients who underwent
coronary revascularization and MCE with
intra-coronary application of contrast, reduced
myocardial contrast effect was demonstrated in
all patients with TIMI grade 2 flow. In 16 of
patients with TIMI grade 3 flow, reduced
myocardial reperfusion was seen. Those patients
had a reduced wall motion score and ejection
fraction at 28 days as compared with patients
with normal myocardial perfusion by MCE, and
their outcome was similar to that of patients
with TIMI grade 2 flow. - These findings demonstrate significant
correlation between myocardial perfusion and LV
function.
41- Porter and coworkers demonstrated that adequate
perfusion imaging may be achieved with
intravenous injection of perfluorocarbon-exposed
sonicated dextrose albumin. In one study,
conducted in 45 consecutive patients 24 16
days after an AMI, 29 of the patients with TIMI
grade 3 flow showed evidence of contrast defect.
The patients had a significant increase in both
end-systolic volume and wall motion score index
at follow-up (4 weeks after myocardial
infarction). - In another study, in which power Doppler harmonic
imaging was used for identification of
reperfusion after AMI, showed very good
sensitivity and specificity (82 and 95,
respectively) of this technique in depicting
perfusion status in segments supplied by infarct-
elated arteries.
42- Moreover, the accuracies of power Doppler
harmonic imaging MCE and SPECT were similar (90
and 92 on segmental basis, and 98 versus 98 on
coronary artery territory basis), which
apparently provides evidence for the superiority
of this technique in the assessment of myocardial
perfusion over the techniques that are based on
greyscale modalities. - There was a correlation between perfusion defects
and functional recovery after 6 weeks, yielding
prognostic information for the recovery of
ventricular function and allowing differentiation
between stunned and necrotic myocardium. - MCE has great potential in this context because
it offers the unique possibility of exploring the
integrity of the microcirculation in vivo, a
fundamental prerequisite for myocardial viability
43Acute coronary syndromes
- In the emergency room, detection of normal
perfusion may allow safe discharge and may avoid
the costs of unnecessary hospitalization. - In the case of detectable perfusion defect
without previous myocardial infarction, an acute
coronary syndrome may be diagnosed and, depending
on the extent of this defect, a conservative or
aggressive diagnostic approach may be applied. - In the situation of established AMI, Agati et
al. compared the infarct size and microvascular
perfusion 1 month after AMI in relation to the
applied therapy primary coronary angioplasty or
thrombolysis. That study showed that, after
successful recanalization of the infarct related
artery, primary angioplasty is more effective
than thrombolysis in preserving microvascular
flow and preventing extension of myocardial
damage. - MCE is a simple bedside method to monitor the
success of therapy, and may help in the
decision-making process (i.e. whether to proceed
to coronary angiography and rescue interventions
when thrombolytic therapy fails to achieve full
reperfusion).
44Chronic ischaemic heart disease
- Currently available methods for detection of
myocardial ischaemia have several limitations. - Stress ECG is positive in only 50 patients with
single vessel disease (stenosis gt70) and
provides limited information on the extent of
ischemia. Stress echocardiography is currently
the most important diagnostic method. The
sensitivity and specificity for CAD are between
80 and 90. However, flow must be reduced to 50
in at least 5 of the myocardium to detect new
wall motion abnormalities. Additionally, stress
echocardiography cannot be used when an unstable
process is suspected. - Myocardial scintigraphy is currently the only
well established method that directly addresses
myocardial perfusion, but this technique is not
as widely available as echocardiography it is
more expensive and has a lower spatial
resolution. - The potential value of MCE in the diagnosis,
treatment, follow-up and outcome prediction of
CAD is clear, especially with the fact that MCE
can identify very early stages in the ischaemic
cascade.
45- In a study conducted by Meza et al. MCE had a
good negative predictive value (gt80) for
estimating attenuated regional function 60 days
after surgery. - In a study that assessed the ability to detect
angiographically significant CAD with accelerated
intermittent imaging after intravenous contrast
administration during stress echocardiography, an
agreement between regional perfusion and
quantitative angiographic findings was found in
217 of the 270 regions analyzed (k 061, 80
agreement, gt50 stenosis). - The greatest incremental benefit of accelerated
intermittent imaging versus wall motion was
gained with dobutamine-induced stress. The
contrast studies depicted 90 of the regions
supplied by a vessel with more than 50 stenosis,
whereas wall motion analysis depicted only 32 (P
0001). - In a study by Porter et al. accelerated
intermittent imaging allowed real-time assessment
of myocardial blood flow and wall thickening
simultaneously, though this requires further
assessment and validation.
46Collateral blood flow assessment
- The extent of coronary collateral circulation can
determine the percentage of necrotic myocardium
during AMI - In the study conducted by Sabia et al. MCE could
define the percentage of the risk area provided
by collaterals, and there was only partial
correlation with angiographic degree of
collaterals. - In the study conducted by Mills et al.in animals,
those investigators demonstrated the possibility
of following the development of coronary
collateral circulation using serial MCE
examinations. This method may permit
investigation of the physiology of collateral
development in vivo and may allow the results of
therapeutic angiogenesis to be monitored.
47- Recent advances in MCE
- Transthoracic echocardiography following contrast
injection helps to detect LV thrombi . - With high-resolution Doppler echocardiography
coronary vasomotion and flow reserve from the
chest without the need for contrast agents can be
measured. - Development of tissue specific microbubbles,
which would allow drug delivery or clot
identification. - Finally, the development of bubbles with a very
narrow distribution may open the possibility to
measure intra-cavitary pressures in a
non-invasive manner.
48Assessment of Myocardial Postreperfusion
Viability by Intravenous Myocardial Contrast
Echocardiography Analysis of the Intensity and
Texture of Opacification, Koji Ohmori, MD,
Circulation April 17, 2001
- examined the relationship between the
opacification pattern produced by a new
intravenous dodecafluoropentane contrast agent
and histological evidence of necrosis or
viability after reperfusion of coronary
occlusion.
49(No Transcript)
50- This is the first application of echo texture
analysis to quantify altered opacification
pattern produced by intravenous MCE in infarcted
segments after reperfusion. - Thus, these data suggested that texture
characterization of the contrast opacification
pattern has the potential to complement
conventional intensity measurements of
intravenous MCE in determining myocardial
viability after reperfusion.
51Identification of Hibernating Myocardium With
Quantitative Intravenous Myocardial Contrast
Echocardiography Comparison With Dobutamine
Echocardiography and Thallium-201 Scintigraphy
Sarah Shimoni, Circulation February 4, 2003
52(No Transcript)
53Conclusion
- Implementation of contrast into clinical practice
will increase the diagnostic power of
echocardiography, providing complex structural
and functional information in a fast one stop
shop examination, without a large increase in
overhead costs. - However, much work needs to be done to
understand the behaviour of bubbles to improve
image quality and to implement new solutions into
echocardiographic equipment. - However, rigorous cost-effectiveness analyses
have not been done, particularly in perfusion
studies. In an era of poor resources, it is
impossible to overlook economic constraints. - On the basis of current evidence, it appears that
contrast echocardiography can improve health
outcomes at reasonable cost and may represent
good value for money, even if it is a
cost-increasing technology. - This aim will be achieved once echocardiography
is refined to the point that it can offer a
high-quality, conclusive result in almost every
examination, and provide anatomical, perfusion
and functional data all in one examination. This
can hopefully be done with the help of contrast
media.
54Thank you