Title: Scintillation, Anger, Gamma Camera & Quality Control
1Scintillation, Anger, Gamma Camera Quality
Control
2The performance of a scintillation camera must be
assessed everyday it is used to assure the
acquisition of diagnostically reliable images.
3Scintillation, Anger, and Gamma cameras
- Cameras come in a variety of configurations,
single, double, and triple head models - Some allow for whole-body imaging, portable use
and tomographic imaging ( SPECT single photon
emission computed tomography) - Planar imaging is appropriate for many studies,
SPECT imaging allows for greater sensitivity and
resolution of imaging deep tissues.
4Gamma Camera
5Gamma Camera
6Almost anything can affect proper function and
performance of a scintillation camera.
- Changes or failure of individual system
components or subsystems
- Environmental conditions
- electrical power supply fluctuations
- physical shock
- temperature changes
- humidity
- dirt
- background radiation
7Crystals
- Crystals used in anger (gamma) cameras are
extremely sensitive to moisture and are sealed in
a aluminum housing. - They are also sensitive to temperature (rapid
changes in temp.) can produce fractures of the
crystal.
8Crystals
- Vary in size from 7 to 25
- Either circular or rectangular
- Crystals are ¼ to ½ thick with 3/8 being the
most common. Thicker the crystal the greater the
probability the incoming photon will interact,
deposit its energy, and be detected.
9Crystals
- The thicker the crystal the better the
sensitivity will behowever - The thicker the crystal the poorer the spatial
resolution will be..because of the complex
interaction b/w crystal, photomultiplier tube,
and light pipe that is used to optically couple
the two.
10Positioning logic
- Anger cameras have an array of PMT
(photomultiplier tubes) attached to the back of
the scintillation crystal - of tubes is determined by size shape of both
the crystal and individual PMT - Common for cameras to have either 37, 55, or 61
PMT
11Positioning logic
- When a scintillation event occurs each PMT
produces an output pulse - Amplitude of pulse from a PMT is directly
proportional to the amount of light
(scintillations) the photocathode has received. - PMTs closest to scintillation event produce
largest output.
12Positioning Logic
- Older PMTround cross section
- Current PMT-hexagonal cross section to cover more
of crystal area which allows for more efficient
detection of scintillation photons. - The more the number of PMTs, the better the
spatial resolution and linearity.
13Positioning Logic
- If only PMT with the largest pulse were used for
x,y positioning, the spatial resolution would be
equal to the cross-sectional size of each tube. - Combining PMT pulses allows better resolution
with x,y coordinates based on Centroid (center of
mass) approach.
14Positioning Logic
- Application of Centroid to Analog Cameras
- Analog network creates four signals (x, x-, y,
y-). - PMT coordinate directions are inversely
proportional to the square of the distance from
respective coordinate and is controlled in a
predetermined, fixed way by the resistor network.
15Positioning Logic
- The sum of these 4 signals z pulse
- The x coordinate of the interaction is
- x (x - x-) / z
- The y coordinate of the interaction is
- y (y - y-) / z
16Energy discrimination
- Pulse Height Analyzer is used in Anger (gamma)
cameras to select a centerline and window - Photopeak energy is used as the centerline and
size of window is determined by percentage of the
centerline energy ex10 of 140Kev to a range
126Kev 154Kev /- 10
17Energy discrimination
- Wider the window the greater the sensitivity.
- However if the window is to wide it will allow
compton scatter and resolution will decrease. - Window width ie 10-20 etc is selected based on
resolution and sensitivity requirements of study.
18Energy Discrimination
- If photon is scattered in the patient, direction
of flight will not intersect the site of origin,
only the site of the Compton interaction. - A large of photons striking the crystal have
been scattered in the patient. - Z pulse in PHA used to discriminate against
scattered photons.
19Energy Discrimination
- Window must be properly peaked or field of view
will not be uniform. - Windows off set on high side of peak result in
appearance of hot PM tubes on the flood. - Windows off set on low side of peak result in
appearance of cold PM tubes on the flood.
20Image formation
- Image can be formed in two ways analog and
digital - Virtually all cameras (analog digital) form
images via digital acquisition - Virtually all cameras are interfaced with a
computer - Counts from the radioactivity that are collected
are stored in the computers memory
21Image formation
- Camera crystals are electronically divided into a
matrix consisting of many small areas called
pixels (picture elements) - Each pixel is assigned a separate storage
location on the computers memory. - At the beginning each storage locations are set
at zero.
22Image formation
- Once a count (radioactivity) is detected on the
crystal the storage location responds by
assigning it a 1 - At the conclusion of an acquisition, each storage
location contains a number of counts
(radioactivity) that have been registered at that
location on the crystal.
23Image formation
- Matrix size determines amount of pixels used.
- If a 64 x 64 matrix is used than 4096 pixels are
used in acquisitions - Matrix size influences spatial resolution of an
image. The larger the matrix size the better the
spatial resolution
24Image formation
- Acquisitions can be acquired in either frame mode
(most common) or list mode. - Framing rate is the amount of each image is
acquired - Frame mode all counts are collected into a
storage matrix in a given time period. Matrix
size and framing rate are preselected.
25Image formation
- One frame of data is acquired by either number of
counts acquired or a fixed length o time - Dynamic frame mode is the framing rate in
sec/frame for a determined amount of time ex 20
sec per frame for 30mins. - Matrix and framing rate cannot be changed after
acquisition.
26Image formation
- Two factors are considered when selecting dynamic
framing study - Purpose of study and counting rate.
- Quantitative study usually requires a shorter
framing rate to provide more data ex renal scan.
- Qualitative study requires more counts per frame
to provide better resoultion
27Image formation
- Higher the counting rate, the shorter the framing
rate. Ex 0.2-0.5sec/frame more frames per time
than 2-5sec/frame - Most common matrix sizes are 64x64, 128x128,
256x256, larger that matrix better the spatial
resolution.
28Image formation
- Zoom mode may be used when a small organ occupies
only a portion of the field of view. - Zoom mode increases resolution prior to
acquisition, postacquisition zoom makes image
larger yet decreases resoultion.
29Image formation
- Information density number of counts per square
centimeter, directly affects resolution - Must obtain adequate counts in a frame resolve
activities b/w adjacent pixels contrast. - Increasing counts decreases signal-to-noise ratio
and improves quality of image.
30The most useful tests to help determine the
proper function of a camera are reflected in the
parameters of field uniformity, spatial
resolution, linearity and sensitivity.
- These are all measured at the time of
installation - confirm specifications
- provide standard for all subsequent performance
evaluations/test - Initial measurements are part of acceptance
testing upon receipt of a new camera as well as
after a camera has been serviced.
31UNIFORMITY
- The most basic measurement of camera performance
is a flood-field uniformity - - this is the ability of the camera to
depict a uniform distribution of activity as
uniform (full-field activity should cause a
scintillation event across the entire camera
head, the source should cover the entire camera
head) - It is assessed by flooding the field of the
camera with a uniform field of radiation and then
assessing the uniformity of the image that it
produces. - Flood fields are examined daily for homogeneity.
32Uniformity
33Inhomogeneous flood fields can result from
- Inadequate mixing of Technetium flood phantom
- Computer correction turned off
- Loss of coupling between crystal and PMTs
- Bad PMT voltage adjustment
- Camera pulse height analyzer off-peak for the
radionuclide
34- Most of the nonuniformity in a camera detector
occurs as a result of spatial distortion (the
mispositioning of events) - To correct distortion, reference images are
acquired and digital correction maps are
generated and stored. - Each map contains values that represent x,y
coordinate shifts. - Microprocessor circuitry repositions each count
in real time during acquisition of the study
using these shifts. - With many new cameras it is best to create
correction maps with the same radionuclide that
is used for patient imaging. - In some cameras several correction maps are
maintained for all the possible radionuclides
that might be used and it is up to the
technologist to select the appropriate one.
35- Variation in the position of a pulse from
different areas of the camera within the pulse
height window may also produce nonuniformities. - This spatially dependent energy variation can
also be corrected by microprocessor circuitry. - The combination of energy variation and spatial
distortion is responsible for the loss of spatial
resolution and imperfect linearity and
uniformity. - In more advanced cameras all of these things are
accounted for through really long boring math
that no one wants to know. - Some systems may even account for
collimator-specific uniformity corrections.
36Uniformity
- Field uniformity may be done either as intrinsic
or extrinsic - Intrinsiccollimators off. Monitors the condition
of sodium iodide crystals and electronics. - Extrinsiccollimators on. Monitors the camera as
it is used clinically.
37Uniformity
- Uniformity images must be inspected daily for
nonuniformity and compared to previous flood
(uniformity) images.
38SPATIAL RESOLUTION
- Can be defined in terms of the amount by which a
system smears out the image of a very small point
source or a very thin line of radioactivity. - It can be thought of as the distance by which two
small point sources must be separated to be
distinguished as a separate in the reconstructed
image.
39- It is usually measured through the use of a
transmission phantom (bar phantom) - It consists of alternating patterns in lead to
produce closely spaced areas of differing
activity levels, which will allow for the
analysis of resolution. - The better the spatial resolution, the better the
ability to detect small abnormalities which will
present themselves as different radionuclide
concentrations in clinical images.
40 - Resolution patterns should be used without a
collimator to measure intrinsic performance. - This can be difficult or impossible along with
time consuming with some multihead cameras. - It can also be useful to assess the resolution
with a point or line source. - The spread of the point or line is an indication
of the degree of blurring, or the loss of the
resolution in the camera. - Resolution should be checked weekly.
41 - Four quadrant resolution studies are usually
required weekly. - To perform four quadrant studies you must rotate
the barphantom 90 degrees so that all quadrants
of the detector are tested.
42LINEARITY
- The ability to reproduce a linear activity source
as linear in the image. - A phantom with either an arrangement of bars or
holes is usually used. - The image produced should look exactly like the
phantom that was used. - For example straight lines should be reproduced
as straight lines and holes should be reproduced
as holes
43- Linearity should be checked weekly
- Linear defects in a flood field are usually the
result of a cracked crystal and less likely the
result of a collimator defect. - Linearity is checked along with spatial
resolution with the bar phantoms
44Florida Administrative Code 64E-5.613Quality
Control of Diagnostic Instrumentation
- Each licensee shall establish written quality
control procedures for all equipment used to
obtain images or information from radionuclide
studies. The procedures shall be recommended by
equipment manufacturers or be approved by the
department. The Licensee shall perform quality
control as specified in written procedures and
retain a copy of the quality control results for
3 years.
45NEMA standards
- National Electrical Manufactures Association
(NEMA) is a trade association for the
manufactures of electrical products. - Nuclear section on diagnostic imaging and therapy
systems division developed the NEMA standards
publication for performance measurements of
scintillation cameras
46NEMA standards
- Document defines standards by which scintillation
cameras may be measured. - Published standards is to define methods by which
scintillation cameras performance can be measured
- A comparison can be made between performance
claims by different manufactors.
47NEMA standards
- American Association of Physicists in Medicine
has published information on scintillation camera
quality control - Published information can provide additional
guidance in designing quality control procedures.
48NEMA standards
- American Society of Nuclear Cardiology has also
published guidelines for Instrumentation quality
assurance, performance, and quality control
procedures for transmission emission tomographic
systems (SPECT)
49Question The most basic measurement of camera
performance is A) Linearity B) Spatial
Resolution C) Flood-Field Uniformity D)
Sensitivity
50Question The most basic measurement of camera
performance is A) Linearity B) Spatial
Resolution C) Flood-Field Uniformity D)
Sensitivity
51- Question
- How often should spatial resolution be checked?
- Daily
- Weekly
- Quarterly
- Annually
52- Question
- How often should spatial resolution be checked?
- Daily
- Weekly
- Quarterly
- Annually
53Are you ready..
- To explode!!!!!!! Turn on the lights!!!!!
- ANY QUESTIONS