ftir - PowerPoint PPT Presentation

About This Presentation
Title:

ftir

Description:

ftir – PowerPoint PPT presentation

Number of Views:3073
Slides: 28
Provided by: shivaram49
Tags: ftir

less

Transcript and Presenter's Notes

Title: ftir


1
FOURIER TRANSFORM OF INFRARED SPECTROSCOPHY
  • Shivaram
  • B.PHARMACY
  • Shiva.pharmacist_at_gmail.com

2
DEFINITION OF INFRARED SPECTROSCOPHY
  • The absorption of light, as it passes through a
    medium, varies linearly with the distance the
    light travels and with concentration of the
    absorbing medium. Where a is the absorbance, the
    Greek lower-case letter epsilon is a
    characteristic constant for each material at a
    given wavelength (known as the extinction
    coefficient or absorption coefficient), c is
    concentration, and l is the length of the light
    path, the absorption of light may be expressed by
    the simple equation a epsilon times c times l.

3
INFRARED SPECTROSCOPHY
  • Infrared spectroscopy is the measurement of the
    wavelength and intensity of the absorption of
    mid-infrared light by a sample. Mid-infrared is
    energetic enough to excite molecular vibrations
    to higher energy levels.
  • The wavelength of infrared absorption bands is
    characteristic of specific types of chemical
    bonds, and infrared spectroscopy finds its
    greatest utility for identification of organic
    and organometallic molecules. The high
    selectivity of the method makes the estimation of
    an analyte in a complex matrix possible

4
EXAMPLE OF IR
5
THEORY OF INFRARED ABSORPTION SPECTROSCOPHY
  • For a molecule to absorb IR, the vibrations or
    rotations within a molecule must cause a net
    change in the dipole moment of the molecule. The
    alternating electrical field of the radiation
    (remember that electromagnetic radiation consists
    of an oscillating electrical field and an
    oscillating magnetic field, perpendicular to each
    other) interacts with fluctuations in the dipole
    moment of the molecule.
  • If the frequency of the radiation matches the
    vibrational frequency of the molecule then
    radiation will be absorbed, causing a change in
    the amplitude of molecular vibration

6
MOLECULAR ROTATIONS
  • Rotational transitions are of little use to the
    spectroscopist. Rotational levels are quantized,
    and absorption of IR by gases yields line
    spectra.
  • However, in liquids or solids, these lines
    broaden into a continuum due to molecular
    collisions and other interactions

7
INSTRUMENT OF IR
8
  • INTRODUCTION TO FOURIER TRANSFORM OF INFRARED
    SPECTROSCOPHY

9
WHAT IS FTIR SPECTROMETER
  • A spectrometer is an optical instrument used to
    measure properties of light over a specific
    portion of the electromagnetic spectrum, 5
    microns to 20 microns.
  • FTIR (Fourier Transform InfraRed) spectrometer is
    a obtains an infrared spectra by first collecting
    an interferogram of a sample signal using an
    interferometer, then performs a Fourier Transform
    on the interferogram to obtain the spectrum.
  • An interferometer is an instrument that uses the
    technique of superimposing (interfering) two or
    more waves, to detect differences between them.
    The FTIR spectrometer uses a Michelson
    interferometer.

10
FOURIER TRANSFORMS
  • Fourier transform defines a relationship between
    a signal in time domain and its representation in
    frequency domain.
  • Being a transform, no information is created or
    lost in the process, so the original signal can
    be recovered from the Fourier transform and vice
    versa.
  • The Fourier transform of a signal is a continuous
    complex valued signal capable of representing
    real valued or complex valued continuous time
    signals

11
SAMPLE ANALYSIS PROCESS
  • 1. The Source Infrared energy is emitted from a
    glowing black-body source. This beam passes
    through an aperture which controls the amount of
    energy presented to the sample (and, ultimately,
    to the detector).
  •  
  • 2. The Interferometer The beam enters the
    interferometer where the spectral encoding
    takes place. The resulting interferogram signal
    then exits the interferometer.
  •  
  • 3. The Sample The beam enters the sample
    compartment where it is transmitted through or
    reflected off of the surface of the sample,
    depending on the type of analysis being
    accomplished. This is where specific frequencies
    of energy, which are uniquely characteristic of
    the sample, are absorbed.
  •  
  • 4. The Detector The beam finally passes to the
    detector for final measurement. The detectors
    used are specially designed to measure the
    special interferogram signal.
  •  
  • 5. The Computer The measured signal is digitized
    and sent to the computer where the Fourier
    transformation takes place. The final infrared
    spectrum is then presented to the user for
    interpretation and any further manipulation.
  •  

12
(No Transcript)
13
FTIR THEORY
  • The spectrometer described here is a modified
    Bomem MB-100 FTIR.
  • The heart of the FTIR is a Michelson
    interferometer .
  • The mirror moves at a fixed rate. Its position is
    determined accurately by counting the
    interference fringes of a collocated Helium-Neon
    laser.
  • The Michelson interferometer splits a beam of
    radiation into two paths having different
    lengths, and then recombines them.
  • A detector measures the intensity variations of
    the exit beam as a function of path difference.
  • A monochromatic source would show a simple sine
    wave of intensity at the detector due to
    constructive and destructive interference as the
    path length changes.

14
  • In the general case, a superposition of
    wavelengths enter spectrometer, and the detector
    indicates the sum of the sine waves added
    together.
  • shows some idealized light sources, and the
    interferograms that they would theoretically
    produce.
  • The difference in path length for the radiation
    is known as the retardation d (OM OF d) .
  • When the retardation is zero, the detector sees
    a maximum because all wavenumbers of radiation
    add constructively.
  • When the retardation is l/2, the detector sees a
    minimum for the wavelength l. An interferogram is
    the sum of all of the wavenumber intensities

15
FTIR BASICS
16
SCHEMATIC OF MICHELSON INTERFEROMETER
17
SAMPLE INTERFEROGRAMS AND THEORETICAL SOURCE
INTENSITY
18
FTIR INSTUMENTATION
  • In a conventional IR spectrophotomer, a sample IR
    beam is directed through the sample chamber and
    measured against a reference beam at each
    wavelength of the spectrum. The entire spectral
    region must be scanned slowly to produce good
    quality spectrum. In 5.32, we will be using a
    Nicolet FTIR Spectrophotometer (Nicolet was
    heavily involved in the design of the Hubble
    telescope!). IR spectroscopy has been
    dramatically improved by the development of the
    Fourier Transform method in much the same way as
    NMR has been revolutionized by this method.

19
  • The heart of an FTIR Spectrophotometer is a
    Michelson Interferometer built around the sample
    chamber. Radiation from an IR source is directed
    through the sample cell to a beam splitter. Half
    of the radiation is reflected from a fixed mirror
    while the other half is reflected from a mirror
    which moved continuously over a distance of about
    2.5 micrometers. When the two beams are
    recombined at the detector, an interference
    pattern is produced. A single scan of the entire
    distance takes about 2 seconds and is stored in
    the computer. In order that several scans may be
    added, they must coincide exactly. Obviously,
    this would be impossible
  • considering the thermal fluctuations and
    vibrations in the laboratory. In order to solve
    this problem, a helium-neon laser is
    simultaneously directed through the Michelson
    Interferometer and the interference pattern of
    the laser is used as a frequency reference. The
    performance of an FTIR is dramatically superior
    to that of conventional instruments. Generally,
    only a small amount of sample will produce an
    excellent spectrum in a fraction of the time.

20
PREPARATION OF SAMPLE
  • Due to the sensitivity of the FTIR instrument,
    the most convenient and satisfactory method
    involves simple evaporation of a solution of the
    sample (chloroform, ether, dichloromethane or
    even a CDCl3 NMR sample may be used) onto a KBr
    salt plate and acquisition of the spectrum from
    the thin film remaining. This method provides
    excellent spectra with flat baseline unless the
    thin film is too powdery in which case
  • excessive scattering of the light leads to an
    irregular baseline. The sample may alternatively
    be prepared as a nujol mull (mull accessories
    agate mortar and pestle, nujol and NaCl discs may
    be obtained from LS).

21
PREPARATION OF INSTRUMENT
  • If the instrument has just been turned on, then
    it is necessary to runa TEST ( F10 ) to be sure
    that all components are ON. If the instrument is
    not turned on or does not check out when the TEST
    is performed, then ask the instrument TA for
    help. In addition, it is important that N2 is
    flowing through the chamber so that most of the
    CO2 and
  • H2O are flushed from the chamber and from inside
    the instrument. F4 SCAN BACKGROUND is performed
    with a blank IR plate in the chamber. F8 then F4
    DISPLAY BACKGROUND will show the spectrum of CO2
    and H2O that remain in the chamber. If the
    background shows excessive CO2 and H2O, then be
    sure the N2 is flowing briskly, wait a minute or
    two and try again. Once a good background has
    been obtained, several students in succession can
    use the same background.

22
SCANNING OF SAMPLE
  • Place the sample plate in the FTIR and wait for
    N2 to purge out the air.
  • F5 SCAN SAMPLE. Wait until the scan and Fourier
    transform are completed.
  • F8 then F1 DISPLAY SPECTRUM will automatically
    subtract the stored background and display the
    spectrum.
  • F7 PRINT. Important Make sure that the printer
    is on-line before pressing F7.
  • Type PEAKPICK S 4000 600 to find the peaks in the
    spectrum. This data is printed by pressing F7 .
    If no one else is using the instrument next,
    please turn off the nitrogen purge

23
FTIR METHODS
  • EPA Method 318 - Extractive FTIR Method for
    Measurement of Emissions from Mineral Wool and
    Wool Fiberglass Industries
  •  
  • EPA Performance Specification 15 for Extractive
    FTIR CEMS in Stationary Sources
  •  
  • EPA Method 320 -Vapor Phase Organic and
    Inorganic Emissions by FTIR (extractive)
  •  
  • EPA Method 321 - Determination of HCl for
    Portland Cement Industries
  •  
  • EPA Protocol for Extractive FTIR for Analysis
    of Gas Emissions
  •  
  • NIOSH Method 3800 - Organic and Inorganic gases
    by Extractive FTIR Spectrometry
  •  

24
FTIR BENEFITS
  • Real-time measurement results.
  •  
  • Simultaneous analysis of multiple gaseous
    compounds.
  •  
  • Measures a wide variety of volatile compounds
    (Inorganic and Organic).
  •  
  • Sensitivity from very low parts per million to
    high percent levels.
  •  
  • Provides a precise measurement method which
    requires no rigorous external calibration.
  •  
  • Speed. Measurements take only seconds

25
ADVANTAGES OF FTIR
  • Some of the major advantages of FT-IR over the
    dispersive technique include
  •  
  • Speed Because all of the frequencies are
    measured simultaneously, most measurements by
    FT-IR are made in a matter of seconds rather than
    several minutes. This is sometimes referred to as
    the
  • Felgett Advantage.
  •  
  • Sensitivity Sensitivity is dramatically
    improved with FT-IR for many reasons. The
    detectors employed are much more sensitive, the
    optical throughput is much higher (referred to as
    the Jacquinot Advantage) which results in much
    lower noise levels, and the fast scans enable the
    coaddition of several scans in order to reduce
    the random measurement noise to any desired level
    (referred to as signal averaging).
  •  
  • Mechanical Simplicity The moving mirror in the
    interferometer is the only continuously
  • moving part in the instrument. Thus, there is
    very little possibility of mechanical breakdown

26
APPLICATIONS OF FTIR
  • Identification of inorganic compounds and organic
    compounds
  • Identification of components of an unknown
    mixture
  • Analysis of solids, liquids, and gasses
  • In remote sensing
  • In measurement and analysis of Atmospheric
    Spectra
  • - Solar irradiance at any point on earth
  • - Longwave/terrestrial radiation spectra
  • Can also be used on satellites to probe the space
  •  

27
  • thanQ
Write a Comment
User Comments (0)
About PowerShow.com