Title: Chromatography Integration and Troubleshooting


1
Chromatography Integration and Trouble-shooting
  • Presented by Marlene Moore mmoore_at_advancedsys.c
    om
  • October 2007

2
Agenda
  • Problems
  • Tools
  • Solutions

3
Problems
  • Baseline Disturbances
  • Irregular Peak Shapes, Sizes
  • Retention Time Shifts
  • Loss of Separation or Resolution
  • Quantitation Difficulties
  • Rapid Column Deterioration
  • Ghost Peaks
  • Broad Solvent Front

4
Baseline Disturbances
spiking
noise
wander
drift
offset
5
Spiking
  • Particulate Matter Passing Through Detector
  • Loose Connections
  • Cables
  • Circuit boards

6
Noise
  • Contaminated Injector/column
  • Physical Defect Detector
  • Defective Detector Board
  • Carrier Gas Leak at Septum or Column
  • FID, NPD, FPD
  • Column in Flame
  • Incorrect Combustion Gases or Flow Rate/leak
  • ECD, TCD
  • Air Leak

7
Wander
  • Contaminated Carrier Gas
  • Isothermal Conditions
  • Contaminated Gas Chromatograph
  • Poor Control of Carrier Gas or Detector Gas Flows
  • Poor Thermal Control of Detector

8
Drift
  • Septum Bleed
  • Upward
  • GC or column contamination
  • Damaged stationary phase
  • Downward
  • Incomplete conditioning of column
  • Unequilibrated detector

9
Offset
  • Contamination
  • Injector, column, detector
  • Carrier gas - change/install impurity traps
  • Column Inserted Into Flame (FID, NPD, FPD)
  • Malfunctioning or Improperly Set Recording Device

10
(No Transcript)
11
Peak Shape
Peak Shape tells the story of your
chromatography
12
Irregular Peak Shapes
tailing
fronting
split
rounded
negative
reduced
13
No Peaks
  • Plugged Syringe
  • Broken Column
  • Injecting Sample Into Wrong Injector
  • Column Installed in Wrong Detector
  • Integrator Reading Wrong Signal or Not Connected
  • Detector Not on
  • Temperature
  • Oven not heated
  • Column temperature too low
  • Injection port temperature too low
  • Detector Gases Improperly Set or Not on
  • Very Low or No Carrier Gas Flow

14
Reduced Peaks
  • Plugged Syringe
  • Change in Injection Technique
  • Large Leak - Injector
  • Split Ratio Too High
  • Too Short of Purge Activation Time for Splitless
    Injections
  • Very High Septum Purge Flow
  • Too Low Injector Temperature
  • Column Temperature Too Low
  • Sample Components Absorbed by Column or Inlet
    Liner

15
Reduced Peaks
  • Initial Temperature Column Too High for Splitless
    or On-column Injections
  • High Background Signal -
  • Contamination
  • Excessive column bleed
  • Check auto zero function and setting
  • Improperly Operated Detectors
  • Impurities Carrier Gas or Detector Gas
  • Detector Compound Mismatch
  • Excessive Attenuated Integrator Signal
  • Sample Concentration or Integrity Problems

16
Tailing Peaks
  • Active Injector Liner or Column
  • Contaminated
  • Injector liner
  • Column
  • Dead Volume Caused by Poorly Installed Liner,
    Column or Union
  • Poorly Cut Column End
  • Polarity Mismatch - Stationary Phase, Solute or
    Solvent
  • Cold Spot in the Flow Path

17
Tailing Peaks
  • Solid Debris in Liner or Column
  • Poor Injection Technique - Too Slow
  • Too Low of a Split Ratio
  • Overloading on a PLOT Column
  • Decrease sample amount
  • Amines and carboxylic acids tail on most columns
  • Need ph modified stationary phase
  • Some Peaks Always Exhibit Tailing

18
Rounded or Flat Topped Peaks
  • Overloaded Detector
  • Exceeding Range of Integrator

19
Split Peaks
  • Detector Overloaded
  • Poor Injection Technique
  • Poorly Installed Column Into Injector
  • Column Temperature Fluctuations
  • Coelution of Two or More Compounds
  • Mixed Sample Solvent for Splitless or On-column
    Injections.

20
Negative Peaks
  • All Peaks Negative - Polarity Check Connectors
  • TCD - Select Peaks
  • Thermal conductivity of compound greater than
    carrier gas
  • ECD - After Positive Peak
  • Dirty or old ECD cell

21
Retention Time Shifts
  • Different Column Temperature
  • Different Carrier Gas Flow or Linear Velocity
  • Leak in Injector - Septum
  • Contaminated Column
  • Change in Sample Solvent

22
Loss of Separation/Resolution
  • Contaminated Column
  • Damaged Stationary Phase
  • Different Column Temperature, Carrier Flow Rate,
    Column
  • Large Changes in Sample Concentrations
  • Improper Injector Operation

23
Quantitation Difficulties
  • Injection Techniques
  • Split Discrimination
  • Different Purge Activation Time for Splitless
    Injection
  • Baseline Disturbances
  • Improper Integrator Settings
  • Inconsistent Detector Gas Flows or Temperatures
  • Column or Liner Activity

24
Rapid Column Deterioration
  • Exposure of Column to Air at Elevated
    Temperatures
  • Exceeding Upper Temperature Limit of Column
  • Chemical Damage
  • Contamination of Column With High Molecular
    Weight Materials
  • Column Breakage

25
Ghost Peaks
  • Contamination
  • Injector
  • Column
  • Septum Bleed
  • Dirty Syringe
  • Sample Too Large
  • Previous Run Terminated Too Soon

BOO!
26
Broad Solvent Front
  • Poorly Installed Column
  • Leak in the Injector
  • Solvent Interacting with Detector
  • Too Low
  • Split Ratio (Flow)
  • Injector Temperature
  • Too Long Purge Activation for Splitless
    Injections
  • Large Injection Volume
  • Low Column Temperature and High Boiling Solvent
  • High Column Temperature and Low Boiling Solvent

27
(No Transcript)
28
Tools
  • Flowmeter
  • Syringe (NEW)
  • Methane or Nonretained Compound
  • Septa, Ferrules, Injection Liners (NEW)
  • Leak Detector
  • Column Test Mixture (Reference)
  • Checkout Column
  • Instrument Manuals

29
Solutions
Quickest Solution ---- Improve Chromatography
--- Do Manual Integration --- Get Results out
Fast !!
NO! WRONG ANSWER!!!
30
Solutions
  • Preventive Maintenance
  • Clean detector
  • Clean injector
  • Clean and repair parts
  • Clean and tighten connections
  • Follow instrument manual suggestions
  • Allow Time for Optimizing System
  • Integration parameters
  • Programs

31
Does anyone have a question?
Question !
Where do you document your trouble shooting?