Title: TBird AVS Ventilator


1
TBird AVS Ventilator TBird VSO2Scott Richey
RRT
2
The TBird AVS ventilator can be used on adults or
pediatric patients weighing at least 10 kg. The
ventilator can be used for ICU transport.
3
TBird AVS
  • 1. Control Panel
  • 2. Ventilator Set-up
  • 3. Setting Modes
  • 4. Ventilation Parameters
  • 5. Setting Alarms
  • 6. Wave Forms
  • 7. Special Functions

4
Control Panel
5
1. Control Knob 2. Controls Section 3. Monitor
Select button
2
3
1
6
1. Mode Section 2. Alarms Section 3. Special
Controls
Airway Pressure Manometer
Monitor Window
1
2
3
7
Ventilator Set-up
8
1. Connect to power source and turn on. Power
switch on the back of machine.
Expiratory Limb
Inspiratory Limb
9
After turning power ON
  • Power source LED illuminates (if not plugged in
    will run off internal battery)
  • Performs POST
  • Begins operation with last used settings.
  • A Flow Cal Alert is displayed in monitor window
    as reminder to perform manual FC
  • Press MONITOR SELECT to clear display. Vent then
    displays the current Apnea Interval settings.
  • Pressing MS again displays current Bias Flow

10
Vent Performance Check akaUser Verification Tests
  • Do before connecting to patient
  • Press and hold down MONITOR SELECT button as turn
    the ventilator ON. Keep holding down Select until
    UVT ON appears (So if vent on already, turn off
    and back on).
  • When you see UVT ON message and hear audible
    alarm ? Silence alarm
  • Press MONITOR SELECT button again - see message
    REMOVE PT

11
  • Press MONITOR SELECT again to display tests
  • 5 tests appear in order as the CONTROL KNOB is
    rotated alarm, lamp, switch, filter, leak, and
    exit.
  • Alarm test - Press MONITOR SELECT to start test.
    The audible alarm sounds. Press MONITOR SELECT
    again to silence alarm. exit test, and advance to
    next test.
  • Lamp test checks the LEDs

12
  • Switch test check function of control panel.
    Pressing MS button should make SELECT appear in
    window. Press any front panel control and the
    name of the control should appear in the window.
  • Filter test checks the turbine inlet filter.
    Activating this tests accelerates flow to 140 LPM
    and measures the pressure drop across filter.
    The message FILTER 4.5 P indicates the pressure
    and a P that the filter passed.

13
Leak Test
  • Most important test.
  • Can select it first by scrolling through with
    control knob.
  • Before starting test block Pt wye.
  • Press MONITOR SELECT to start test
  • Vent holds a pressure of 60 cmH2O twice.
  • If the difference between the two tests is cmH2O, the test passes.
  • Press MS to exit test and advance to EXIT out of
    special function section.

14
After leak test
  • Set Apnea Interval using Special Functions
  • Set Bias Flow using Special Functions
  • To access Special Functions - Press and Hold
    Control Knob until VENT SETUP appears in the
    monitor window.
  • Accessing Special Functions does not interfere
    with ventilation.

15
Special Functions
  • There are 7 chapters under Special Functions
  • Vent Setup Group, Alarm Setup Group, Transducer
    Data, Transducer Test, Event Codes, User
    Verificiation Tests, Service Verification Tests.
  • Vent Setup includes Bias Flow, and Control Lock.
  • Alarm Setup includes apnea interval

16
Next after checking Bias Flow and Apnea Interval
  • Set Over Pressure Relief valve - located lower
    right section of vent.
  • Set High Pressure Alarm to 120 cmH2O
  • Create high pressure in circuit (block pt wye)
  • Adjust relief valve to the backup setting ( 10
    above high pressure alarm)
  • Set mode
  • Check volume of audible alarm (back panel)

17
Finally
  • Set controls
  • Set alarms

18
MODES
  • Assist-Control in VC, PC and VAPS
  • SIMV in VC, PC and VAPS
  • Pressure Support
  • Volume Assured Pressure Support (VAPS)
  • CPAP

19
To set a mode press desired mode button (LED
light flashes) press again within 5 seconds to
activate mode, otherwise the mode remains
unchanged.
Mode Select Buttons
20
To set pressure control mode 1. Select A/C mode
2. Then push PC pad 3. Dial desired pressure. To
switch back to VC, turn Pressure Control
pressure level to OFF.
21
To add Pressure support 1. Select appropriate
mode (SIMV or CPAP) 2. Push PS pad 3. Dial in PS
setting
1
2
3
22
To set Volume Assured Pressure Support, Select
A/C or SIMV mode, set VT, Peak Flow and Pressure
Control level and push VAP button
23
VAPS for machine breaths only
  • Will give a pressure breath that is flow cycled
    as long as inspiratory flow does not fall to Peak
    Flow setting before target tidal volume
    delivered.
  • If flow falls to Peak Flow setting before target
    volume delivered, ventilator transitions to VC
    mode and delivers flow at the set peak flow until
    target tidal volume delivered.

24
VAPS
Flow falls to set Peak Flow before target volume
delivered, switches to VC
Flow
Press
25
Ventilation Parameters
26
To set ventilation parameters. 1. Press desired
parameter button turn the control knob to
change setting.
27
After setting parameters you can lock the control
panel to prevent setting changes. 1. By pressing
the control knob. 2. A light to the right of the
knob will illuminate. 3. Press again to unlock
2
1
28
ALARMS
29
The main alarms to set on the TBird are
  • Apnea Alarm (10-60 seconds)
  • Alarm Silence (lasts 1 minute)
  • High pressure limit
  • Low peak pressure
  • Low minute volume
  • High respiratory rate

30
To set alarms 1. Push the button for desired
alarm 2. Turn the control knob to change the
setting.
31
WAVEFORMS
32
The TBird has a optional Graphic monitor
interface for viewing wave forms.
33
Special Functions
  • These are accessed by pressing the control knob
    for 2 seconds.
  • Clinical parameters found here include adjustment
    of bias flow apnea period alarm.
  • Bias flow can be adjusted from 10 20 LPM.
  • Press control knob again to Exit.

34
Other Functions
Insp hold press and hold button before start of
inspiration. Set pause time by pressing button
and turning control knob.
35
Battery
  • The battery lasts approximately for 1 Hour.
  • Internal battery charges whenever vent plugged
    into AC source and vent turned off (standby).
  • When power turned off, alarm sounds, press alarm
    silence. Flashing Vent Inop indicator continues
    to flash for several minutes (does not effect
    battery).

36
TBird VSO2
  • What the clinic has
  • Similar to AVS but with less stuff (no NIF, Exp
    hold, VAPS or descending waveform)

37
(No Transcript)
38
Power on
39
Battery from the side
40
TBIRD VSO2
41
(No Transcript)
42
(No Transcript)
43
(No Transcript)
44
(No Transcript)
45
No VAPS
46
No Exp hold or MIP
Pressure relief