Title: Principles of Mechanical Ventilation
1Principles of Mechanical Ventilation
- RET 2284
- Module 4.0 Ventilator Management
- - Initial Assessment
2Assessment and Documentation of MV
- This module will review the assessment and
documentation of the patient-ventilator
interaction in the initial stages
3Assessment and Documentation of MV
- Verify the physicians order
- Some orders are specific
- Mode, rate, Vt, FiO2, PEEP
- Some orders are flexible
- Desired range of PaCO2 and PaO2
- Initial settings Mode, rate, Vt, FiO2, PEEP
which can be manipulated to achieve the desired
blood gas results
4Assessment and Documentation of MV
- Verify that the ventilator has passed OVP
- OVP Operational Verification Procedure
- Usually an automated self-test
- Usually includes a system-leak test to check the
integrity of the ventilator circuit, humidifier
and related equipment - Done before placing a patient on the ventilator
for the first time and before reconnecting the
patient to a ventilator if the circuit has been
changed or disassembled for any reason - Must be documented
5Assessment and Documentation of MV
- Place the patient on the ventilator
- Set parameters on ventilator in accordance with
the physicians orders - Connect the patient to the ventilator
- Listen to breath sounds to confirm adequate
volume delivery and proper placement of ET Tube
6Assessment and Documentation of MV
- Check patients vital signs especially heart
rate and blood pressure because these may be
affected by mechanical ventilation - Note Positive pressure ventilation can reduce
venous return to the heart, cardiac output, and
blood pressure
7Assessment and Documentation of MV
- Activate alarms (e.g., apnea, low pressure, low
Vt, and high pressure limits) - Obtain ABG about 15 minutes after ventilation
- Evaluates effectiveness of ventilation and
oxygenation
8Assessment and Documentation of MV
- Chest x-ray Verify ET tube placement
- If the patients clinic presentation indicates a
need, other determinations my be include - CBC
- Glucose, sodium, potassium, chloride, CO2, blood
urea nitrogen, creatinine, phosphate, magnesium - PT, PTT, platelet count
- Blood, sputum, urine culture
9Assessment and Documentation of MV
- Documentation of the PT-Vent System
- AKA Vent Check
- Patient information and ventilator settings
should be documented regularly when a patient is
receiving ventilator support every 1 4 hours
depending on the institutions policy - Documentation is done on a ventilator flow sheet
electronic or paper record
10Assessment and Documentation of MV
- Documentation of the PT-Vent System
- AARC Recommendations
- Data relevant to the patient-ventilator system
check are recorded on the appropriate hospital
form and become part of the patients medical
record - The patient-ventilatory system check includes
patient information and observations of
ventilator settings at the time of the check
11Assessment and Documentation of MV
- Documentation of the PT-Vent System
- AARC Recommendations
- The record should include the physicians order
for mechanical ventilator settings - The patient-ventilatory system check includes a
brief narrative of the clinical observations of
the patients response to mechanical ventilation
at the time of the check
12Assessment and Documentation of MV
- Documentation of the PT-Vent System
- Patient information
13Patient and ventilator parameters
14Patient and ventilator parameters
- Patient and ventilator parameters (cont.)
15Patient and ventilator parameters
- Patient and ventilator parameters (cont.)
16Patient and ventilator parameters
- Patient and ventilator parameters (cont.)
17Patient and ventilator parameters
- Patient and ventilator parameters (cont.)
18Patient and ventilator parameters
- Patient and ventilator parameters (cont.)
19Assessment and Documentation of MV
- Mode
- Institution determines the method for charting
mode - Sensitivity
- Check for autotriggering
- Assess the patients ability to trigger the vent
- Documentation
- Pressure triggered (e.g., - 2 cmH2O)
- Flow triggered (e.g., 2 L/min)
20Assessment and Documentation of MV
- Tidal Volume, Rate, Minute Ventilation
- Usually displayed digitally
- May be verified with respirometer
- Correcting Tubing Compliance
- Volume loss due to tubing compliance
(compressible volume) - Accounts for approximately 1.5 2.5 mL/cm H2O of
lost volume - Newer vents automatically compensate
21Assessment and Documentation of MV
- Determination of Circuit Compressible Volume Loss
- Set VT to 100-200 and PEEP to zero
- Set inspiratory pause at 2 seconds
- Select a minimum flow rate and maximum pressure
limit - Occlude Y-connection and initiate a mechanical
breath - Record the exhaled volume (ml) and peak
inspiratory pressure (cm H2O) - Divide exhaled volume by PIP (V/PIP) circuit
compression factor - Multiply compression factor by the patients PIP
(PIP minus PEEP if PEEP is used) - Example
- Circuit compression factor 150 ml / 50 cm H2O
3 ml/cm H2O - Circuit compression volume 3 ml/cm H2O x (60 cm
H2O PIP 10 cm H2O PEEP) -
- 3 ml/cm H2O x 50 cm H2O 150 ml (circuit
volume Loss)
22Assessment and Documentation of MV
- Alveolar Ventilation
- Because of the use of HMEs and other circuit
adapters that add mechanical dead space to the
ventilator circuit, a knowledge of the effect of
dead space on alveolar volume delivery is
important especially in infants, children, small
adults, and adults with ARDS treated with low
tidal volumes
23Assessment and Documentation of MV
- Alveolar Ventilation
- Anatomical Dead Space
- Normal anatomical dead space (VDanat) is about 1
mL/lb IBW - Note Bypassing the upper airway with an
artificial airway reduces VDanat by about one half
24Assessment and Documentation of MV
- Alveolar Ventilation
- Added Mechanical Deadspace (VDmech)
- Y-connectors, HMEs, and/or flex tubing all add
mechanical deadspace - To measure this volume, fill the device with
water and empty into graduated container) - When determining actual alveolar ventilation, the
volume of these devices must be subtracted from
the tidal volume - Example VT VDanat VDmech Alveolar
Ventilation
25Assessment and Documentation of MV
- Alveolar Ventilation
- Final Alveolar Ventilation
- VA (VT VDanat VDmech) x f
.
26Assessment and Documentation of MV
- Alveolar Ventilation - QUESTION
- A 36-year-old male patient with ARDS is
ventilated with a VT of 400mL. The patients IBW
is 176 lb (80 kg). The HME has a volume of 50
mL. What is the approximate alveolar volume for
one breath for this patient? - 350 mL
- 260 mL
- 400 mL
- 190 mL
27Assessment and Documentation of MV
- Alveolar Ventilation - ANSWER
- VDant is about 180 ml (80 kg IBW x 2.2 lb
176). With an ET tube in place, the VDant is
reduced to about half to 90 mL. The HME adds
about 50 mL of mechanical dead space. VD is
about 140 mL. - VT VD VA
- 400 140 260 mL/breath
- VA 260 mL
28Assessment and Documentation of MV
- Monitoring Airway Pressures
- Ensures that very high pressure limits are not
exceeded - Provide information about the patients conditions
29Assessment and Documentation of MV
- Monitoring Airway Pressures
- Peak Inspiratory Pressure (PIP or PPeak)
30Assessment and Documentation of MV
- Monitoring Airway Pressures
- Peak Inspiratory Pressure (PIP or PPeak)
- The highest pressure observed during inspiration
- Used to calculate dynamic compliance (CD)
- A constant VT with an ? PIP may indicate a ? in
lung compliance (CL) or an ? in Raw - A declining PIP may indicate a leak or may a sign
of improvement in CL or Raw
31Assessment and Documentation of MV
- Monitoring Airway Pressures
- Plateau Pressure (PPlateau)
32Assessment and Documentation of MV
- Monitoring Airway Pressures
- Plateau Pressure (PPlateau)
- Obtained by using the ventilators inspiratory
pause of 0.5 1.5 seconds - Static pressure is read when no gas flow is
occurring - Reflects the elastic recoil of the alveolar walls
and thoracic cage against the volume of air in
the lungs - Cannot be measured accurately if the patient
makes active respiratory efforts
33Assessment and Documentation of MV
- Monitoring Airway Pressures
- Set Pressure
- The operator sets a target pressure to be
delivered to the patient during PC-CMV, PC-SIMV
and PSV - Example
22 cm H2O
15 cm H2O
34Assessment and Documentation of MV
- Monitoring Airway Pressures
- Peak Pressure Minus Plateau Pressure
- The difference between PIP and PPlateau is the
transairway pressure (PTA) this is the amount of
pressure required to overcome Raw (Raw
PTA/Flow) -
35Assessment and Documentation of MV
- Monitoring Airway Pressures
- Peak Pressure Minus Plateau Pressure
- A higher than expected difference between PIP and
PPlateau usually indicates increased Raw - Raw increases with secretions, mucosal edema,
bronchospasm, kinked ET tube or patient biting on
ET tube, partly occluded HME (moisture or
secretions)
36Assessment and Documentation of MV
- Monitoring Airway Pressures
- End-Expiratory Pressure (EEP)
- The lowest pressure measured in the expiratory
phase
37Assessment and Documentation of MV
- Monitoring Airway Pressures
- End-Expiratory Pressure (EEP)
- The lowest pressure measured in the expiratory
phase - PEEP Positive End Expiratory Pressure
- CPAP Continuous Positive End Expiratory
Pressure - Auto-PEEP
38Assessment and Documentation of MV
- Monitoring Airway Pressures
- End-Expiratory Pressure (EEP)
- Auto-PEEP air trapping during positive pressure
ventilation AKA - Air trapping
- Breath stacking
- Inadvertent PEEP
- Dynamic hyperinflation
- Occult PEEP
- Intrinsic PEEP
39Assessment and Documentation of MV
- Monitoring Airway Pressures
- Causes of Auto-PEEP
- Increased Raw
- Airway collapse on expiration
- Bronchospasm
- COPD
- Mucosal edema
- Secretions
- Short TE
- Long TI
- Slow peak flow rate
- High rate
- High VE
- High IE ratio
- Increase Raw from ET Tube, expiratory valve, PEEP
valve
40Assessment and Documentation of MV
- Monitoring Airway Pressures
- Auto-PEEP Suspected
- Accessory muscle use
- Decreased breath sounds
- Decreased chest wall movement
- Dyspnea
- Increased resonant percussion
- Increased radiolucency on CXR
- Inspiratory efforts do not trigger ventilator
- Patient still exhaling when vent delivers next
breath - Patients respiratory rate greater than
ventilators response (assuming sensitivity is
set correct)
41Assessment and Documentation of MV
- Monitoring Airway Pressures
- Auto-PEEP Suspected
42Assessment and Documentation of MV
- Monitoring Airway Pressures
- Measuring Auto-PEEP
- Exhalation valve is occluded for 1-2 seconds just
prior to inspiration - Expiratory Pause control on newer ventilators
- Level of auto-PEEP is reflected on the pressure
gauge during the pause - Reading is accurate only if the patient is not
actively breathing - Auto-PEEP Total PEEP set PEEP
43Assessment and Documentation of MV
- Monitoring Airway Pressures
- Methods to reduce Auto-PEEP
- Brochodilators / suction
- Steroids
- Allow more spontaneous breathing (SIMV, PS, CPAP)
- Larger ET Tube
- Apply PEEP (up to 80 of auto-PEEP)
- Increase TE
- Decrease respiratory rate (? TE )
- Decrease VT (? TI)
- Decrease TI
- Increase peak flow rate (? TI)
44Assessment and Documentation of MV
- Monitoring Airway Pressures
- Mean Airway Pressure
- Closely parallels the mean alveolar pressure
- Newer vents calculate and display mean airway
pressure - Affected by
- PIP
- EEP
- TI/Total cycle time
- f
- Inspiratory flow patterns and modes
- Important to tissue oxygenation!!!
- Affects lung volumes and cardiac output
45Assessment and Documentation of MV
- Monitoring Airway Pressures
- Pressure Limit
- Usually set 10 15 cm H2O above PIP
- Audible and visual alarms are activated if PIP
exceeds a set limit - Activation of high pressure limit alarm ends
inspiration - Often activated by coughing
- High peak pressures may indicate ? Raw or ? CL
46Assessment and Documentation of MV
- Monitoring Airway Pressures
- Low Pressure Alarm
- Visual or audible alarm is activated when
pressure within the vent circuit has fallen
significantly - Leak or disconnect
- Usually set 10 cm H2O below PIP
- Note If the leak is not obvious, the patient
must be manually ventilated until the cause of
the leak is determined
47Assessment and Documentation of MV
- Vital Signs
- Observing and recording the patients blood
pressure (BP), heart rate (HR), temperature (T),
respiratory rate (f), oxygen saturation (SpO2),
and color every few hours help staff members
evaluate possible changes in the patients
overall condition - Moderate changes in vital signs should alert the
practitioner to the possibility of hypoxemia,
impending cardiovascular collapse, or infection
48Assessment and Documentation of MV
- Vital Signs
- Heart Rate
- All patient on ventilatory support must be
continuously monitored with a three-lead ECG - Provides maximum/minimum HR alarms
- ECG electrode disconnection
- Vent disconnection -
- (hypoxemia, hypercapnia)
- Myocardium infarction (MI)
- Anxiety
- Pain
- Stress
49Assessment and Documentation of MV
- Vital Signs
- Temperature
- Hyperthermia
- Infection
- Tissue necrosis
- Metabolic states, e.g., hyperthyroidism
- Atelectasis
- Accidental or surgical trauma
50Assessment and Documentation of MV
- Vital Signs
- Temperature
- Hypothermia
- Metabolic disease
- CNS disorders
- Drugs
- Alcohol, heroin, carbon monoxide
51Assessment and Documentation of MV
- Vital Signs
- Systemic Arterial Blood Pressure
- Monitored intermittently
- Stethoscope/sphygmomanometer
- Automatic BP cuff
- Monitored continuously
- Invasive intravascular arterial catheters
- A-Line
Radial A-Line
52Assessment and Documentation of MV
- Vital Signs
- Central Venous Line (CVP)
- Placed in the superior vena cava or right atrium
- Provide valuable information regarding
- Right heart function
- Fluid status
- Hypervolemia
- Hypovolemia
53Assessment and Documentation of MV
- Vital Signs
- Pulmonary Artery Pressure
- Monitored continuously
- Swan-Ganz catheter
- AKA Balloon-tip, flow-directed, pulmonary artery
catheter (BTFDC) - Used in patients with severe cardiopulmonary
complications
54Assessment and Documentation of MV
- Physical Examination of Chest
- Inspection
- Accessory muscles usage - ? WOB
- Paradoxical breathing - ? WOB
- Abdominal distention
- Gas, air swallowing, bleeding, ascites
- Impairs ventilation
- Palpation
- Percussion
- Auscultation
55Assessment and Documentation of MV
- ET and Tracheostomy Tube Cuffs
- Cuff Pressure Measurement
- Checked once per shift
- Pressures not to exceed
- 27 34 cm H2O (20 25 mm Hg)
- Excessive pressures my cause tracheal damage if
cuff pressures are greater than tracheal
perfusion pressures
56Assessment and Documentation of MV
- ET and Tracheostomy Tube Cuffs
- High Cuff Pressure
- Overinflation of tube cuff
- Necessary to maintain minimum occlusion
- Artificial airway may have moved up into the
larynx of pharynx - Check depth of tube or chest x-ray
- ET tube may be too small
57Assessment and Documentation of MV
- ET and Tracheostomy Tube Cuffs
- No or Low Cuff Pressure
- Cuff not properly inflated
- Cuff leak
- Check inflate cuff and clamp pilot tube
- Resolution change cuff
- Pilot balloon leak
- Check inflate cuff and place stopcock in pilot
balloon in off position - Resolution clamp pilot tube
- Pilot tube leak
- Resolution syringe with blunt-tipped needle
58Assessment and Documentation of MV
- Tube and Mouth Care
- Reposition ET tube every shift (Pilbeam) and
retape if necessary - Most facilities do this every 24 hours
- Prevents pressure injury to gums, mouth, lips
- Usually a two-man procedure
- Mouth care should be performed every shift
- Helps prevents ventilator-acquired pneumonia
(VAP)
59- JADA Continuing Education
- Pneumonia in nonambulatory patients
- The role of oral bacteria and oral hygiene
- Frank A. Scannapieco, DMD, PhD Background.
Considerable evidence exists to support a
relationship between poor oral health, the oral
microflora and bacterial pneumonia, especially
ventilator-associated pneumonia in
institutionalized patients. Teeth or dentures
have nonshedding surfaces on which oral biofilms
(that is, dental plaque) form that are
susceptible to colonization by respiratory
pathogens. Subsequent aspiration of respiratory
pathogens shed from oral biofilms into the lower
airway increases the risk of developing a lung
infection. In addition, patients may aspirate
inflammatory products from inflamed periodontal
tissues into the lower airway, contributing to
lung insult.
60Assessment and Documentation of MV
- Mouth Care Kit for MV Patients
61Assessment and Documentation of MV
- Monitoring CL and Raw
- Static Compliance (Cs)
- Normal 70 100 mL/cm H2O (Cs VT/Pplat
PEEP) - Changes in Cs over time is usually considered a
result of change in the patients alveolar
elastic recoil (can be affected by chest wall
compliance) - Causes for ? Cs
- Air trapping
- PE
- Atelectasis
- Consolidation
- Pneumonia
- Pneumothorx
- Hemothorax
- Pleural effusion
62Assessment and Documentation of MV
- Monitoring CL and Raw
- Static Compliance (Cs)
- Chest wall compliance reduced in the following
- Flail chest
- Chest wall muscle tension
- Pneumonmediastinum
- Abdominal distention
63Assessment and Documentation of MV
- Monitoring CL and Raw
- Static Compliance (Cs)
- Pressure Ventilation (PV)
- Reduced CS
- Set pressure remains constant while delivered VT
decreases - Increased CS
- Set pressure remains constant while delivered VT
increases
64Assessment and Documentation of MV
- Monitoring CL and Raw
- Static Compliance (Cs)
- Volume Ventilation (VV)
- Reduction in CS
- Delivered VT remains constant while pressure
increases - Increases in CS
- Delivered VT remains constant while pressure
decreases
65Assessment and Documentation of MV
- Monitoring CL and Raw
- Dynamic Compliance (CD)
- Normal 40 70 mL/cm H2O (CD VT/PIP-PEEP)
- Includes compliance and resistance components
- Lung and chest wall elastic recoil
- Airway resistance
- Decreases when
- CS decreases
- Raw increases
66Assessment and Documentation of MV
- Monitoring CL and Raw
- Differentiating between lung and airway
resistance problems - Volume Ventilation
- ? PIP and ? Pplat (constant PTA) ? CS
- ? PIP and Pplat constant (? PTA) ? Raw
67Assessment and Documentation of MV
- Monitoring CL and Raw
- Airway Resistance (Raw)
- Normal Raw 0.6 2.4 cm H2O/L/sec
- Approximately 6 cm H2O/L/sec or higher in
intubated patients - Can be estimated by PTA/Flow (L/sec)
- Use constant flow (square wave) when measuring
Raw - Many new ventilators calculate Raw
68Assessment and Documentation of MV
- Monitoring CL and Raw
- Airway Resistance (Raw)
- Increased Raw caused by
- Bronchospasm
- Secretions
- Mucosal edema
- Small ET tube
- Kinked or bitten ET tube
- Treat the cause!
69Assessment and Documentation of MV
- Monitoring CL and Raw
- Airway Resistance (Raw)
- Graphs demonstrating ?Raw
VC Ventilation
PC Ventilation
70Assessment and Documentation of MV
- Comment Section of Vent Flow Sheet
- Breath sounds
- Chest movement
- Percussion note
- Skin color
- Level of consciousness
- Changes in vent settings
- Changes in physicians order
- Description of any equipment problems
Some medical centers use SOAP notes