Title: Louisiana State University Radiation Safety Office


1
Louisiana State UniversityRadiation Safety Office
  • Module 12
  • General Rules for Irradiator Use

2
Operating the Irradiators
  • Shepherd Co-60 Irradiator
  • Safety Features
  • The Shepherd Co-60 Irradiator is equipped with an
    Interlock System which prevents the opening of
    the chamber door when the radioactive source is
    in the exposed position. To operate the interlock
    system, a special key is needed. A radiation
    monitoring system (placed on top of the
    irradiator) continuously measures the radiation
    level in the vicinity of the irradiator and will
    set off an alarm whenever the radiation level
    exceeds the limit.

3
Operating the Irradiators
  • Shepherd Co-60 Irradiator
  • Operating Procedures
  • Unlock the irradiator door padlock and place the
    padlock on top of the power panel
  • With the other (interlock system) key, turn the
    panel POWER on
  • Check that the INTERLOCK light is on. Check that
    the LOW PRESSURE light is off.
  • Verify both sources are in SAFE position, i.e.
    the source position indicators are at the bottom
    of the columns.

4
Operating the Irradiators
  • Shepherd Co-60 Irradiator
  • Open the door by pressing the red DOOR SWITCH
    button and at the same time opening the door
    latch. If the door does not open, release the
    DOOR SWITCH and press against the door to make
    sure the door sensor is activated. Then,
    immediately press the DOOR SWITCH and open the
    door latch at the same time.
  • Place the sample in the chamber at a
    pre-determined position. Refer to the chamber
    dose rate graphs to determine the proper sample
    positioning and exposure time. The graphs
    indicate the percent of dose rate in different
    areas of the chamber.

5
Operating the Irradiators
  • Shepherd Co-60 Irradiator
  • Close the door by pushing on the handle. The
    door is very heavy, so make sure your other hand
    is away from the opening.
  • Select the source activity, (i.e. 5 Ci or 3000
    Ci) by setting the source toggle switch
    accordingly.
  • Set the sample exposure time PRESET or MANUAL.
  • (A) For PRESET time operation
  • 1. Set the time toggle switch to PRESET
  • 2. Based on your calculated exposure time, set
    the timer to the nearest 1/100 minute
  • 3. Push the RESET button to clear previous
    setting and accept the new setting.
  • (B) For MANUAL time operation
  • Set the time toggle switch to MANUAL

6
Operating the Irradiators
  • Shepherd Co-60 Irradiator
  • To start the irradiation, press the RED source
    button. Immediatley, the selected source will
    raise to the exposed position, shown by the
    source position indicator position at the top of
    the column. Check that the timer is showing the
    elapsed time.
  • The irradiation can be paused or terminated at
    anytime by pressing the GREEN source button. For
    preset time operation, the timer will stop when
    the source is returned to SAFE position, and will
    continue when you press the RED source button and
    the source raises to the exposed position.

7
Operating the Irradiators
  • Shepherd Co-60 Irradiator
  • For preset time operation, the irradiation is
    automatically terminated at the preset time and
    the source automatically returns to its SAFE
    position.
  • Before opening the chamber door, verify both
    sources are in the SAFE position.
  • Remove the sample from the chamber and close the
    chamber door.
  • Turn the Power off and lock the padlock

8
Operating the Irradiators
  • Shepherd Co-60 Irradiator
  • Safety and Emergency Procedures
  • Do not put explosive or pressurized materials in
    the chamber.
  • Never attempt to defect the chamber door
    interlock. This is a condition of our license to
    operate the irradiator. Always notify Radiation
    Safety Office personnel if the door does not open
    when the sources are in the safe position.
  • The chamber door is very heavy. When closing the
    door, push on the handle and make sure you other
    hand is away from the opening.
  • If the radiation monitor alarms, immediately
    close the chamber door, exit the basement area
    and notify Radiation Safety Office personnel.

9
Operating the Irradiators
  • Pool Irradiator Co-60
  • Safety Features
  • High activity Co-60 radioactive sources are
    placed at the bottom of the pool, and the water
    in the pool is the main radiation shielding to
    keep the radiation level in the room below an
    acceptable limit. A locked metal grid covers the
    pool surface to prevent access into the pool.
    Only a small opening (with a metal grid cover) is
    provided for transferring a special container
    into the pool. A radiation monitoring system is
    placed in the room. The system is placed at one
    corner of the room and monitors the ambient
    radiation level.

10
Operating the Irradiators
  • Pool Irradiator Co-60
  • Operating Procedures
  • Turn on the light in the pool. The switch box is
    located on the wall next to the room radiation
    monitoring system.
  • Unlock the pool padlock, remove the metal bar and
    the grid cover over the small opening.
  • Put the sample(s) in the special container
    located on the preparation bench next to the
    pool.
  • Secure the top of the container (which is already
    attached to the hoist) to the container with the
    three C-clamps provided.

11
Operating the Irradiators
  • Pool Irradiator Co-60
  • Using the hoist system, carefully transfer the
    container through the opening into the pool until
    the container sits in the hole at the bottom of
    the pool. Use your own timer/watch to register
    the start of irradiation.
  • Cover the opening with the metal grill and the
    metal bar, lock the pool padlock, and turn off
    the pool light.
  • At the end of the irradiation period, recover the
    container by unlocking the security grill, lift
    with the hoist and then transfer the container
    back to the preparation bench.
  • Recover the sample(s), leave the container on
    the bench, cover the opening with the metal grill
    and the metal bar, lock the pool padlock, and
    turn off the pool light.

12
Operating the Irradiators
  • Pool Irradiator Co-60
  • Safety and Emergency Procedures
  • Upon entering the room, check if the pool is
    filled with water.
  • The material to be irradiated has to be placed in
    the special container. Do not put explosives or
    dry-ice in the sample container.
  • Make sure you do not drop anything in the pool.
    Report to Radiation Safety Office personnel if
    you accidentally do so.
  • If the radiation monitor alarms, immediately exit
    the basement area and notify the Radiation Safety
    Office.