Mouse Bone Nanoindentation Device - PowerPoint PPT Presentation

1 / 17
About This Presentation
Title:

Mouse Bone Nanoindentation Device

Description:

Mouse Bone. Nanoindentation Device. Group Members: Ryan Fischer. Megan Toth ... The goal of this project is to adapt a MTS 858 Bionix servohydraulic testing ... – PowerPoint PPT presentation

Number of Views:245
Avg rating:3.0/5.0
Slides: 18
Provided by: aidedeng
Category:

less

Transcript and Presenter's Notes

Title: Mouse Bone Nanoindentation Device


1
Mouse BoneNanoindentation Device
  • Group Members Ryan Fischer
  • Megan Toth
  • Client Dr. Robert Blank, Ph.D., M.D.
  • Advisor J.A. Gimm, Ph.D.

2
Outline
  • Problem Statement
  • Background Information
  • Product Design Specification
  • Design Alternatives
  • Proposed Solution
  • Future Work

3
Problem Statement
  • The goal of this project is to adapt a MTS 858
    Bionix servohydraulic testing machine to indent
    mouse bone to determine the mechanical strength
    of bones.

4
Background Information
  • Motivation
  • Fragility Fractures
  • Current Research
  • Current Method
  • Three-point bend test
  • Nanoindentation
  • Elastic Modulus

MTS Machine
5
Nanoindentation
Indentation Schematic, 1
Load vs. Displacement, 1
6
Product Design Specification
  • Process to provide homogeneous surface for
    indentation.
  • Stabilizer
  • Attach to 5000 lb. Load cell with eight screws of
    3/8 fine thread, R2.375.
  • Nanoindenter
  • D0.01 mm
  • Provide small indentation depth of 2-3 microns
    under load from 0-1 N.
  • MTS Attachment
  • Attachment for Nanoindenter to Load Cell

7
Design Alternatives
  • Process for Smoothing
  • Stabilizer
  • Piece Model
  • Nanoindentor tip
  • Square Pyramid
  • Nanoindenter Attachment
  • Double Jointed

8
Smoothing Process
  • Fix bone in Petri dish with epoxy.
  • Stabilize Petri dish in epoxy.
  • Saw bone with microscopy saw.
  • Polish bone with serial diamond cloths of
    increasing fineness placed over the saw blade.

9
Piece Model
5,000 Lb. Load Cell
10
Square Pyramid
Bottom View
Side View
11
Double-Jointed
50 Lb. Load Cell
12
Proposed Solution
  • Smoothing Procedure
  • Screw Model for Stabilizer
  • Trigonal Pyramidal Indentor Tip
  • Straight Indenter Attachment

13
Screw Model
14
Trigonal Pyramidal Indenter
Bottom View
Side View
15
Straight Indenter Attachment
50 Lb. Load Cell
16
Future Work
  • Research on
  • Indenter Tip Materials
  • Load Cells
  • Stabilizer Materials
  • Develop Final Design
  • Build Prototype
  • Testing

17
Acknowledgements
  • Dr. Robert Blank
  • Ron McCabe
  • Tony Escarcega
  • 1 http//www.inano.dk/sw262.asp