Title: Electron Tomography Tutorial Part I: Instrumentation
1Electron Tomography Tutorial Part I
Instrumentation Advances in Electron Microscope
Instrumentation and Exploration of Biological
Structures
James Bouwer, PhD. and Albert Lawrence,
PhD National Center for Microscopy and Imaging
Research
June 28, 2009
2Core Instrumentation Development Team A
Collaborative Enviroment
PRINCIPAL INVESTIGATORS Mark Ellisman, Nguyen
-Hu Xuong RESOURCE STAFF SCIENTISTS James
Bouwer Albert Lawrence Sebastien
Phan Steven Peltier Tomas
Molina Masako Terada Mason
Mackey Thomas Deerinck Tristan
Shone George Yang GRADUATE STUDENTS Rick
Giuly, Anna Milazzo, Liang Jin
3Overview of the Tomography Process
- What is tomography
- 2D images to 3D volumes
- Inherent aberrations in electron microscopes
- Physics of magnetic lenses
- Aberrations resulting from typical imaging
conditions - Aberrations resulting directly from tilted image
acquisition - Inherent 3D nature of the problem and solutions
- Back-projection
- X-ray transforms
- Non-linear transforms (TxBR)
- New Wide-field imaging systems
- High-resolution 8k x 8k lens-coupled imaging
system for EM - Wide field driving need for image/volume
remapping - New tomography modalities
- Multi-axis tomography, thick section tomography
(MPL,micro-probe STEM)
4What is Electron Tomography
Tomograpy Tilt Series
5Intermediate Voltage Electrons Microscope
Resources
6Toadfish Swimbladder Muscle -gt lead and uranyl
acetate Contrast vs. Voltage
100 keV --gt 100nm thick 200 keV --gt 200 nm
thick 400 keV --gt 500nm thick 1.25 MeV --gt
3microns 3 MeV --gt 10microns
Thickness 80nm
7Limitations of Light Microscopy
8Electron Microscopy for Ultra-wide-field
Multi-Resolution 3D Imaging
9Photo-oxidation of cys4 transfected Hela cells to
prepare for electron tomography
- Cells labeled with ReAsH and imaged live
- Live action frozen with fixatives
- Illuminated to create singlet state oxygen
- Reactive oxygen polymerizes DAB
- Label with Osmium
- embed in epoxy
Live Cell Imaging
photo-oxidized golgi
10The Electron Microscope
11Cartoon of Electron Trajectories in a Solenoidal
Field
from J. Bozzola and L. Russel Electron Microscopy
2nd edition
12Physics of Electron Lensing
Can be solved analytically through separation of
variables in cylindrical coordinates to produce
the equations of motion for the imaging electrons
- Bz z-component of the magnetic field
- ?L Larmor frequency
- e charge of an electron
- me rest mass of an electron
Bell Shaped Field
Image from L. Reimer, TEM 1993
13Effects of Varying the Current to the Objective
Lens
Focus Series of 10nm and 20nm Gold Beads with 1µm
Focus Steps Acquired on 4k x 4k CCD
14Real Electron Trajectories in Rotating Reference
Frame
Electron trajectories for electrons incident
parallel to the optical axis for various B- field
strengths
Image from L. Reimer, TEM 1993
15A Simplified Cartoon of the Electron Tomography
Geometry with a Field Immersion Lens
- Differential magnification
- Differential rotation
- More pronounced for large
- format images
- Rotation and magnification
- are more troublesome for
- tomography
Distortion Correction is Inherently a 3D Problem
(rotation)
(magnification)
16Tilt Geometry Distortions
172-Dimensional Warping is Insufficient
2µm thick Golgi Sample
18Projector Lens Aberrations
19Inherent Electron Microscope Spatial Aberrations
Pincushion
Barrel
Spherical aberration in the projector produces
radially dependent magnification errors
20Largest Spatial Aberrations are S-type
Distortions in the Projector Lens Optics
- EM manufacturer Specs
- lt 1.5 _at_ r 5cm
- For a 4k x 4k detector 6cm diameter
- 40 pixels
- For a 8k x 8k detector 10cm diameter
- ? 80 pixels
-
21S-Distortion Correction Enables Image Mosaicing
Hardware Solution to S-distortion correction for
Montaging
JEM-3200 Drift Tube for S-distortion mitigation
22EM STUDY OF THE AUTOPHAGY PATHWAY AFTER BRAIN
ISCHEMIA 2 x 2 mosaic tomogram (8k x 8k pixels)
23Beam Induced Mass Loss (Shrinkage and Sample
Warping)
The Jellyfish Analogy
24Standard X-ray Backprojection The last 40 years
Sample
- Linear electron
- trajectories
- Back project image
- density
- Filtration is necessary
- to remove artifact
25A Tilt Series Shows Aberrations Inherent in
Electron Imaging Aberrations Can Cause Any Number
of Image Errors
Dividing NRK (Rat Kidney) cell photo-oxidized Golg
i ReAsH label
26Fiducial Marker Tracking
Remap the images so that the tracks are
level A. Lawrence - The Tao of TxBR
27Image Remapping Pre-alignment
Remaped Particle Tracks
28Corrected Particle Track Parameters Are Applied
to the tilt images
29Transform Based Backprojection Assumes Curved
Electron Trajectories
Calculation of R is fully parallelizable
30Application of Transform Based Backprojection
Dividing NRK cell photo-oxidized Golgi ReAsH label
31NCMIRs New 8k x 8k Wide-field Ultra-high
Resolution Detector Highlights Tomography
Challenges
32Basic Design of Commercial Slow-Scan CCD
Detectors for TEM
e-
Scintillator
Fiber Optic Relay
TEC(Peltier) Cooling
33A 64 Mega Pixel Digital Detector for TEM
34SupraCam in Cross-Section
Electron beam
JEOL interface flange
Large diameter gate valve
Self-supporting scintillator (260mm diameter)
Scintillator chamber
Pyramidal beam splitter
Leaded glass window
Optics1 Lens
90mm Shutter
Spectral Inst. CCD
Custom 13 axis stages
35(No Transcript)
36(No Transcript)
378k x 8k Alzheimer's Plaque - Human Biopsy Section
38Collaborative Project 4 ENTRY, REPLICATION, AND
ASSEMBLY OF NODAVIRUS IN DROSPHILA CELLS High
Pressure Frozen, Freeze Substituted, 40kx Mag
39Montaging with the Stage and Image Shift Coils
404 Billion Pixel Montage on 8k x 8k Camera
41Montage Tomography
-stage montaging -image shift montaging
422 x 2 Montaged Tomogram (8k x 8k) of a Flock
House Virus Infected Drosophila Cell at 8kx
Magnification
zero - loss energy filtered
43Problem Position
- Thin slices become warped during sample
sectioning, handling and data acquisition (beam
induced mass loss and lens distortions) - Difficulty in stacking volumes in a serial
tomography
44Serial Section Tomography
Two Possible Approaches
- Transform an already reconstructed volume
- Modify the projection maps during the bundle
adjustment procedure
45A shear based warping transformation
Orthogonal warping transformation
work by Sebastien Phan, NCMIR-UCSD
46Neuron Specimen
Shear based transformation
Naoko3A7
47Serial Section Tomography
astrocyte soma at 8k mag
48Serial Section Montage Tomography
-stage montaging -image shift montaging
49 Movie Coming Soon
50The Contrast Transfer Function (CTF)
where
I(k) Image O(k) Object k spatial frequency
and
CTF(defocus, Cs ,?E, source size)
51Wide-field Images of 60 degree tilted sample show
a strong focus gradient (CTF gradient)
?z 10µm
?z -10µm
52Large Field of View Requires CTF Reconstruction
from a Thru Focus Series
2D gt Fourier Transform 3D gt Fourier Integral
Operator
Generated with CTF generator written by Wen
Jiang and Wah Chiu
53Multiple Axis Tomography
54(No Transcript)
55Titan Equipped with 3 Lens Condenser Illumination
System
- Triple Condenser Illumination STEM Mode
- Ability to adjust probe convergence angle
- lt 1mrad convergence angle/parallel probe long
depth of field - Micro-probe STEM Tomography
- Highly parallel probe, 1mrad convergence angle,
2-4 nm resolution - Dynamic focus
- 4k x 4k STEM will increase ability to image wider
fields
Much thicker specimens (gt 6 µm) increase the z
dependent aberrations in micro-probe STEM
56Copper-Lead Stained Spiny Dendrite - MicroProbe
STEM Tomography
57Very High Resolution Tomography tomography in ice
TxBR Reconstruction using a Cubic
Backprojection Shows Actin Sub-units
58Summary
Electrons travel along curved paths
Linear Backprojection Methods are insufficient
Especially true for large format images and for
high resolution tomograms!
59(No Transcript)
60Procedure for Solving this Set of Eqs.
- Seed computer with the knowns
- -particle positions (xj, yj) in each tilt
- image
- -The approximate (x,y,z) coordinates
- for each particle in the object are
- compute using triangulation
- Use conjugate gradient optimization to compute
- aij? bij? cij? dij? and precise (xj,
yj, zj) - This method of solution is called Bundle
Adjustment
61Transform Based Back-Projection
(TxBR)
A Parallel Algorithm for Volume Computation
In GTS
Albert Lawrence James Bouwer Able Lin Masako
Terada Guy Perkins
62The Story of Transform Based Back Projection
(TxBR).
Errors inherent to tomography with magnetic
lenses result in errors which cause smearing of
the object density.These errors cannot be
corrected using linear back-projection algorithms.
- A. Rick Lawrence
Electrons travel along curved paths
resolution is the goal
63TxBr (Transform Based Backprojection)
- Very High Resolution Tomography
- High Resolution Wide Field Tomography
- Challenging Data Sets
- Non-eucentric goniometers
- Critical Data (sample warping)
64Computation of the backprojection (R) requires
the polynomial approximations to the projection
(R) along electron paths (?)
26 Unknowns!
65Sample calculation of the required number of
fiducal markers
- Quadratic model with 60 tilts and 20 fiducials
- Unknowns (26)(tilts) (particles)(3coords)
- 156060 1620
- Knowns (2coords)(20fiducails)(60tilts) 2400
- Cubic Model (cubic projector lens distortions
- number of unknowns (43)(60)(40)(3) 2700
- number of fiducials 40
- Highly Parallelizable!
- Well Suited for Grid Abstraction
66Very High Resolution Tomography
TxBR Reconstruction of ice embedded actin using
a cubic backprojection shows actin sub-units
67Imod vs. TxBR High-resolution Wide-field
Tetracystine labeled Actin Fibers and Bundles
TxBR with remap
Linear Methods (IMOD)
68(No Transcript)
69Continued Progress in Electron Optical Sectioning
- Collecting data to building accurate point-spread
function models - Working on techniques to deconvolve the sections
- Working with Angus Kirkland to explore techniques
for controlling the stigmator coils to reduce
aberrations
Dev Team J. Bouwer, T. Molina, G. Yang, B.
Smith, Y. Hakozaki
70Challenges in Electron Microscopy New
Approaches to the Tomography Process
- Physics of Magnetic Lenses in EM Tomography
- Inherent aberrations in electron microscopes
- Challenges in tomography
- Contrast Transfer Function is inherently 3D
- New Data Collection Modes
- Large field montaging tomography
- Most Probably Loss Tomography
- Electron Energy Loss Imaging
- Multiple Axis Tomography
- New Instrumentation
- Ultra-wide field Cameras
- HAADF STEM microprobe microscopy
- EELS
71Serial Section Tomography
astrocyte soma at 8k mag