Title: A UNIVERSAL PROGRAMMABLE TEMPERATURE SPRAY CHAMBER
1A UNIVERSAL PROGRAMMABLE TEMPERATURE SPRAY
CHAMBER
- Jerry Dulude, Glass Expansion, jdulude_at_geicp.com
2Limitations of Room Temperature Spray Chambers
- Intensity drift with temperature
- Excessive plasma loading (volatile solvents)
- Excessive oxide formation
- Insufficient control of analyte transport
3Limitations of Externally Controlled Spray
Chambers
- Requires antifreeze solution
- Requires bulky floor model chiller
- Messy coolant lines from chiller to chamber
- Freezing of condensate on lines
- Requires intricate jacketed chamber
4IsoMist Controlled Temperature Spray Chamber
5IsoMist Encapsulated Spray Chamber
6Features of IsoMist
- Powerful Peltier-effect chiller reaches -5C in 15
minutes - Programmable from -10 to 60C in 1 degree
increments - Maintains temperature to within 0.1 degree
- Custom software runs from any PC
- Compact design (7.5x4x4 inches)
- 100 self-contained (no external lines)
- Incorporates Bluetooth technology for clean
wireless control (USB available)
7PC Screen showing IsoMist Software
8Applications of the IsoMist
- Analyze volatile solvents
- Reduce oxide interferences
- Lower detection limits
- Increase stability
9Determination of Trace Metals in Naphtha
- Interfere with the cracking process
- Poisoning of the catalysts
- Environmental release concerns
- Origin and migration markers
10Analysis of Naphtha by ICP-OESPE Optima 2100 DV
(Axial view)
- Forward Power 1500Watts
- Coolant flow 20L/min
- Auxiliary flow 1.8L/min
- Nebulizer gas flow 0.35L/min
- Injector 1mm capillary bore
- Spray Chamber Twister baffled cyclonic
- Nebulizer SeaSpray glass concentric
- Uptake rate 300ul/min
- IsoMist temperature -10C
11IsoMist on Optima 2100DV
12Spike Recoveries in Naphtha
13Oxide Reduction in ICP-MSIsobaric oxide
interferences
Robert Thomas, Practical Guide to ICP-MS, Marcel
Dekker, NY, 2004.
14Reducing Oxides for ICP-MS
Data courtesy of David Jones, ALS Chemex,
Brisbane, Australia
15Analysis of Limited Sample Volumes
- Biological applications
- Intracellular fluids
- Neonatal samples
- Forensic applications
- Requires micro flow nebulization
- Low DLs desirable
16Effect of Temperature on Normalized Intensity
(20ul/min Uptake) ICP-OES
PE Optima 2100DV
17Effect of Temperature on DL20ul/min Uptake
(ICP-OES)
PE Optima 2100DV
18Stability EnhancementICP-OES
19IsoMist Compatibility
Models
- PE ICP ICP-MS
- Varian ICP ICP-MS
- Agilent ICP-MS
- ThermoFisher ICP ICP-MS
- Horiba J-Y ICP
- Teledyne Leeman ICP
- GBC ICP
- Spectro ICP
Spray Chamber Materials
- Borosilicate glass
- Quartz
- Inert PFA