Title: Detection of Ziram by LC/APCI/MS


1
Detection of Ziram by LC/APCI/MS Hiroshi
MORIWAKI Osaka City Institute of Public Health
Environmental Sciences
2
Ziram
use
pesticide antifouling agent
toxicity
endocrine disruptor mutagen carcinogen
((CH3)2NCS.S)2Zn M.W. 305.8 m.p. 240 Log Pow 1.09
3
Analysis of Ziram
Direct detection by GC/MS ?
Conventional method
Decomposition of the ligand to CS2

Chemical speciation of dithiocarbamate pesticides
LC/MS ( Direct detection of ziram )
4
Ion mode
Ionization
Positive ion mode
Negative ion mode
ESI/MS



APCI/MS
Corona currentlt0.6mA

?
Corona currentgt0.6mA
5
Mass spectrum
B)
A)
6
Corona current vs. peak intensity
7
Cone voltage vs. peak intensity
8
Conditions of APCI/MS
9
Column separation
10
Mass chromatogram
11
Calibration curve
12
Summary Direct detection of Ziram by
LC/APCI/MS Corona current gt 0.6mA LC
separation GPC column Instrumental detection
limit 2mg/L
Next step Development of the pretreatment
Use of LC/MS/MS
ref. H. Moriwaki, Y. Yamaguchi and M.
Fukushima Rapid Commun. Mass Spectro., in press
(2001).