Powerpoint template for scientific posters Swarthmore College

1 / 1
About This Presentation
Title:

Powerpoint template for scientific posters Swarthmore College

Description:

IGRT Using A Video-Based 3-D Surface Imaging System in Combination with CBCT ... order to generate a large 3-D surface image which can be detected by the AlignRT ... – PowerPoint PPT presentation

Number of Views:69
Avg rating:3.0/5.0
Slides: 2
Provided by: colin210

less

Transcript and Presenter's Notes

Title: Powerpoint template for scientific posters Swarthmore College


1
IGRT Using A Video-Based 3-D Surface Imaging
System in Combination with CBCT Claus Chunli
Yang Ph.D., Tianxiao Liu Ph.D., Joerg Lehmann
Ph.D., Samir Narayan M.D., Rick Harse R.T.T.,
Srinivasan Vijayakumar M.D., James Purdy
Ph.D. Department of Radiation Oncology,
University of California, Davis, Sacramento, CA
Fig. 1 AlignRT System, VisionRT Ltd. London, UK
Fig4 , Patient registration by patients surface
features. Left Panel Reference image. Reference
image was taken following initial patient setup.
Right Panel Verifying image. Verifying images
were taken after reference image was taken and
before patients were treated and after patients
treatments. The shift determined by the first
verifying image and reference image was compared
with the shift from CBCT. In all cases surface
imaging system verified that the patient shifts
in all three directions are within one millimeter
with the ones determined by CBCT with an
exception that about 20 of the vertical shifts
agree between 1 to 2 mm
Fig. 2 Ball-Baring Device (BBD) and
Ball-bearing Device with a foam phantom added
Fig 3 , Difference between Shifts Detected by
AlignRT and Actual Values
Write a Comment
User Comments (0)