Title: BACT Reporting: Back to the Basics
1BACT Reporting Back to the Basics
- Common Mistakes, Reporting Errors, and Strategies
for Avoiding Both
Eliza Petzinger March 2007
2Overview
- BACT reporting in detail.
- Common mistakes and reporting errors.
- Strategies for reducing errors.
- The benefits of electronic submission.
3BACT Reporting
- Why focus on BACTs?
- Increasing volume of reporting errors.
- October December 2006, most violations issued
were for reporting errors and omissions.
4BACT Reporting Responsibility
- What if the lab is at fault?
- Ultimately the PWS is responsible for all
elements of BACT reporting. - Letting labs fill out reporting forms without any
oversight is very risky. - Checking laboratory reporting and results is
imperative for avoiding submitting mistakes.
5Error Reduction
- Why is error reduction important?
- Reduced public notification.
- Increased consumer confidence.
- Better assurance of public health.
6The Bacteriological Analysis Report Form
7Unique for each system. This is required for all
sampling reports. This is a mandatory element!!!
Only one PWSID can be reported on a form.
8Month and Year format (ex 032007). Generally
the month the samples were collected. If a
Routine sample is collected at the end of the
month (pos.) and the repeat sample is collected
in the next month, the repeat should still be
reported in the previous months compliance
period (ex positive sample on Feb 28th,
resampled on Mar 1, both should be reported for
February).
9Month, day and year format (MMDDYYYY
03102007). Each form can ONLY contain samples
collected on the same day. Use a new form for
each day.
105 Digit code assigned to each certified lab.
11MMDDYYYY This is an optional element! Identifies
the date the sample was received at the lab.
12MMDDYYYY This is the actual day the sample was
analyzed.
13Coded value for the method analysis (EX Colilert
Presence /Absence 309, Membrane filtration
m-Endo 303) Only one method code can be
reported per page!!! A new page must be used for
each method used on a particular day.
14Coded value for the method analysis (EX Colilert
Presence /Absence 309, Membrane filtration
Coliscan 319) Only one method code can be
reported per page!!! A new page must be used for
each method used on a particular day.
15Routine (RT) These are the same as Distribution
samples. Repeat (RP) These are the same as
check samples. Special (SP) (Ex line breaks,
emergency repairs, line extensions, etc)
Specials do not count toward compliance and can
not be substituted for repeat samples!!!!!
16Coded value indicating the reason for the sample.
A suspected contamination B New plant,
Modification, or line extension C Treatment
Modification D Study/Investigation E Line
break, Emergency repair
17Only two choices yes or leave BLANK. This
section denotes that this sample is a
replacement for a CNFG or TNTC negative sample.
18These are the approved numbers for Routine
samples. SDWIS/State will cross-check location
codes with approved sample sites. If reporting a
Repeat sample (RP) you should enter RPU or RPD in
this column for the upstream and downstream. The
original number should be used for the original
site. If the original sample was RPU or RPD, then
use RPO.
19This is for Repeat samples only! Original site
OR Upstream UP Downstream DN This is a
required element if the sample is a repeat. This
must be completed correctly to avoid a
violation!!!!!!
20Time must consist of 4 digits!!! (24-hour
clock) 710 am 0710 125 pm 1325
21This is the free chlorine residual. Chloramine
systems will not fill this section out. Values
should be in mg/L or ppm. Note inclusion of
decimal point. May be recorded in tenths or
hundredths.
22Chloramine systems must fill this section
out. Chlorine systems may fill this out if they
wish. This data should be in mg/L or ppm. Note
inclusion of decimal point.
23This is a unique number assigned by the lab for
that year only. This can contain up to 8
characters. This can be devised any way a lab
would like. One example is to combine the date
analyzed with the location code (ex 0816031 or
0922F32).
24This is the time the sample was analyzed in the
lab. Do not indicate am or pm. Results must be
four digits (ex 710 am 0710 and 125 pm
1325).
25Enter the total coliform count if the method
being utilized gives a count. If there are no
total coliform present then leave this column
blank!! Do not enter 0. You may also enter TNTC
or CNFG.
26Indicate Presence (P) or Absence (A) of coliform
in the sample. If a sample is TNTC or CNFG
negative (total coliform is entered as an A),
only a replacement sample must be collected. If a
sample is TNTC or CNFG positive (total coliform
is entered as P), then a set of three Repeat
samples are required.
27If a sample is total coliform positive, the lab
must analyze for E. coli. Enter P for present or
A for Absent. Do not enter anything if total
coliform was negative!
28When a Routine sample (RT) is positive for total
coliform and a set of three Repeat samples have
been collected, each of the repeat samples must
reference the lab sample number of the original
sample. If this is not filled out, this may
result in a violation!!!
29Examples of Correctly Formatted BACT Forms
30(No Transcript)
31(No Transcript)
32(No Transcript)
33(No Transcript)
34(No Transcript)
35(No Transcript)
36(No Transcript)
37A Note About Special Samples
- Do not code special samples with RPU or RPD.
- DWB does not use specials for compliance purposes.
38(No Transcript)
39(No Transcript)
40Examples of Common Errors
- Invalid location codes
- Always use state approved location codes
(sampling sites). - A copy of approved location codes on file can be
obtained through the DOW. - Invalid RP location codes
- Must be original location code, RPU, and RPD.
41(No Transcript)
42(No Transcript)
43Examples of Common Errors
- Miscoded sample type
- SP instead of RP
- RT instead of RP
- RT instead of SP
- Checking for mixing of sample type elements is an
easy way to avoid this.
44(No Transcript)
45(No Transcript)
46Examples of Common Errors
- Invalid or missing RP location codes.
- Must be OR, UP, DN.
- RT samples with repeat location codes.
- Cannot be accepted by state as RP or RT.
47(No Transcript)
48Examples of Common Errors
- Missing sample time, free or total chlorine.
- Missing or miscoded PWSID.
- Compliance period or collection date missing or
do not match. - Sample exceeds 30 hour holding time.
49Examples of Common Errors
- Missing or incorrect lab analysis time.
- Duplicate Lab Sample Numbers.
- Missing or miscoded Total Coliform or E.Coli P/A
indicators. - Missing or incorrect Original Lab Sample Numbers
for RPs.
50Strategies for Reducing Errors
- Proofread BACT forms before they are sent.
- Notify the DWB of mistakes when you catch them.
- Do not allow labs to submit unchecked analysis
forms.
51Electronic Submission
52How Electronic Submission Works
53How Electronic Submission Works
54The Pros
- Catches most mistakes.
- Omissions
- Invalid entries
- Allows for timely correction of mistakes and
resubmission. - Lab notified of mistakes within a few days
- Resubmitted results can still be used for
compliance - Paper copies not required
- Paper copies should still be kept on file.
55What mistakes will be caught?
- Any omission of required elements
- Sample type, location code, analysis time,
chlorine levels, etc. - Duplicate lab sample numbers.
- RPs not properly linked to a positive original
sample. - Replacement samples not properly linked to
original sample. - Invalid location, lab, and method codes.
56Cons
- Mistakes not caught.
- Routines and repeats coded as specials
- Repeats coded as routines
- Harder for PWS and labs to review
- Rejection report allows PWS to review only
samples that need to be corrected instead of
having to review all samples! - Too many to catch mistakes
57 NOTE submitting electronically
- When an error is found by the PWS or corrections
made after NOV -
- Hard copy must be sent in with corrections
- Miscoded samples
- Ex RT instead of RP
-
-
58How can your PWS utilize electronic submission?
- Encourage your lab to use this technology.
- Find a lab that already uses this technology.
59BACT reporting and Compliance
- Not complicated unless you make it complicated
- written procedures
- plans for tackling problems
- examples of correct BACTs
- understand rules and regulations
- Never trust a lab with anything more than testing
your water. - PWS may be held accountable for all mistakes.
60ATTNPWS taking fewer than 5 samples
- Have one (1) or more total coliform-positive
samples - At least five (5) RT samples should be collected
during the next month of service.
61Final Thoughts
- Dont hesitate to call or e-mail the DWB with
questions. - 502-564-3410, ext. 505
- Eliza.Petzinger_at_KY.GOV
- Dont wait until we discover a problem.