VALIDATION OF MOIST HEAT STEILIZATION - PowerPoint PPT Presentation

1 / 22
About This Presentation
Title:

VALIDATION OF MOIST HEAT STEILIZATION

Description:

'The act or process, physical or chemical, that destroys or ... BIER(biological indicator evaluator resistometer) Inoculate the spore into the actual solutions ... – PowerPoint PPT presentation

Number of Views:2486
Avg rating:5.0/5.0
Slides: 23
Provided by: bio73
Category:

less

Transcript and Presenter's Notes

Title: VALIDATION OF MOIST HEAT STEILIZATION


1
VALIDATION OF MOIST HEAT STEILIZATION
  • JM Tech.
  • Do-Young Ahn

2
Definition
  • ? Sterilization
  • The act or process, physical or chemical, that
    destroys or eliminates all viable microbes
    including resistant bacterial spores from a fluid
    or a solid.
  • Examples of sterilization methods are steam
    treatment at 121?, dry heat at 230?, flushing
    with a sterilizing solution such as hydrogen
    peroxide (H2O2) or ozone (O3), irradiation, and
    filtration.
  • ?Sterility
  • The reduction of anticipated levels of
    contamination in a load to the point where the
    probability of survival is less than 10-6.

3
Definition
  • ?D-value
  • The time in minutes required for a one-log or
    90 reduction of a specific microbial population
    under specified lethal conditions. For steam
    sterilization it is determined at a constant
    temperature
  • ?z-value
  • The number of degree of temperature change
    necessary to change the D-value by a factor of
    10.

4
Definition
  • ?F value(lethal rate, instantaneous Fo)
  • The F value is a measurement of sterilization
    effectiveness. F(T,z) is defined as the
    equivalent time at temperature T delivered to a
    container or unit of product for the purpose of
    sterilization, calculated using a specific value
    of z.
  • ?Fo value(accumulated Fo)
  • The term "Fo " is defined as the number of
    equivalent minutes of steam sterilization at
    temperature 121.1C delivered to a container or
    unit of product calculated using a z-value of
    10C.
  • Fo ? 10((121-T)/z)?t

5
Methodology
  • ?Overkill Sterilization
  • ?Provides a minimum 12 log reduction of a
    resistant BI w/ a known D-value of not less than
    1 minute.
  • ?Required minimal information on the bioburden
  • ?Bioburden/Bioindicator Sterilization
  • ?Provides a probability of survival of less than
    1 in 106 for the bioburden as demonstrated using
    a resistant BI w/ a known D-value.
  • ?BI may not be inactivated
  • ?Requires information on the numbers and heat
    resistance of the BI.
  • ?Requires ongoing monitoring or control over
    bioburden.

6
Methodology
  • ?Bioburden Sterilization
  • ?Provides a probability of survival of less than
    1 in 106 for the most resistance bioburden
    expected in the load.
  • ?Requires information on the numbers and heat
    resistance of the BI.
  • ?Requires ongoing monitoring or control over
    bioburden.

7
Sterilizer Cycle
  • ?Gravity Displacement
  • ?Difference of density
  • Density of air at 20? 1.2 g/l
  • Density of steam at 100? 0.6 g/l
  • ?Effectiveness of air elimination depends on the
    rate of steam supply
  • Air pocket too rapidly
  • Diffusion into the steam too slowly, more
    difficult to remove
  • ?Specially designed steam trap permitting the
    passage of large volume of air

8
Sterilizer Cycle
  • ?Prevacuum cycle
  • ?A more effective method
  • ?By means of a mechanical vacuum pump or a steam
    eductor
  • ?Vacuum as low as 1520 mmHg, apply for 810 min.
  • ?Pulsing cycle
  • ?A series of alternating steam pulses followed by
    vacuum excursions
  • ?Air-steam mixture
  • ?Terminal sterilization of large volume
    parenterals
  • ?Air injection required to compensate the great
    expansion of air or nitrogen in the head space
    above the liquid
  • ?Well mixed chamber fan, raining effect by
    external pump w/ cooling

9
Cycle Development
  • ?Consider factors into account
  • ?Nature of the load porous materials, heat
    sensitivity of the products
  • ?Type of the sterilizer
  • ?Employed containers and closures
  • ?Heat stable product overkill approach
  • ?Heat liable product bioburden approach
  • Bioburden studies number of microorganisms
  • D-value studies only highly resistant spore
    formers,
  • BIER(biological indicator evaluator
    resistometer)
  • Inoculate the spore into the actual solutions
  • For solid materials, precut strips

10
Preparing for Validation
  • ?Temperature sensing devices
  • ? T type thermocouples(copper-constantan)
    encased in flexible sheaths
  • ? Premium grades of wire having ?0.1? accuracy
  • ?Temperature standards
  • ? RTD traceable to the National Bureau of
    Standards , IPR, HTR
  • ?Calibration of thermocouples
  • ?At two temperatures 0 ?, 130 ?
  • ?Correction factors
  • ?Stability ?0.03?
  • ?Accuracy ?0.5?

11
Preparing for Validation
  • ?Autoclave
  • ?Validation nozzle and adaptor
  • ?Data logger digital output and multi-channel
    device
  • ?BIs or biological challenges
  • ?Loads

12
Validation Protocol
  • Protocol should include
  • ?Objectives of the validation
  • ?Responsibilities of validation personnel and
    operating department personnel
  • ?Identification and description of the sterilizer
    and its process control
  • ?Identification of SOPs equipment
  • ?Calibration of instrument SOPs and/or
    description
  • ?Identification and calibration of the
    temperature monitoring equipment

13
Validation Protocol
  • ?A description of the following studies
  • ?Bioburden determination studies(if applicable)
  • ?Empty chamber heat distribution studies
  • ?Container mapping studies(if applicable)
  • ?Loaded chamber heat penetration studies
  • ?Microbiological challenge studies
  • ?Evaluation of drug product cooling water(if
    applicable)
  • ?Integrity testing of vent filter
  • ?Acceptance criteria
  • ?References
  • ?Review and approval

14
Heat Distribution Studies
  • ?To demonstrate the temperature uniformity and
    stability of the sterilizing medium throughout
    the sterilizer
  • ?Conduct on both the empty and loaded chamber
    with max. and min. load configurations
  • ?Acceptance criteria Less than 1?of the mean
    temperature
  • ?Conduct 3 runs to obtain consistent results
  • ?Distribution of the thermocouples geometrical
    representatives, exhaust drain, adjacent to the
    control sensor
  • ?At least 10 probes, normally 1520 probes

15
Heat Distribution Studies
  • ?At loaded chamber heat distribution test, the
    thermocouples should be positioned in the same
    locations used for empty chamber heat
    distribution
  • ?Avoid contacting solid surfaces
  • ?Do not place within any containers
  • ?Data should be obtained at regular intervals

16
Container Mapping
  • ?To determine the coolest point within the liquid
    filled container
  • ?Temperature mapping should be conducted on all
    the different container types, sizes and fill
    volume to be validated
  • ?The number of the thermocouples used depends on
    the container volume
  • ?Possible to use a single thermocouple at
    different positions,
  • and can be conducted in a smaller autoclave or
    retort
  • ?Penetration thermocouples should be positioned
    at the cold spot having lowest temperature or Fo

17
Heat Penetration Studies
  • ?To determine the coolest point(s) within the
    specified load and configuration, and to assure
    that these points be consistently exposed to
    sufficient heat lethality
  • ?Prior to conduct heat penetration studies,
    determine max. and min. load configurations
  • ?Probed container at the cold spot should be
    distributed uniformly throughout the load
  • ?Penetration thermocouple are positioned at
    points within the process equipment suspected to
    be the most difficult for steam heat penetration

18
Heat Penetration Studies
  • ?Lethal rate can be determined from the
    temperature data
  • by the following formula
  • L log-1(To-Tb)/z 10((To-Tb)/z)
  • ?A summation over time of the lethal rate at a
    series of temperature(accumulated lethality)
  • Fo ? 10((121-T)/z)?t
  • ?Regard to product stability

19
Microbial Challenge Studies
  • ?Biological challenges are employed during heat
    penetration studies in order to demonstrate the
    degree of process lethality provided by the
    sterilization cycle
  • ?Microorganism frequently utilized
  • ?Overkill Bacillus stearothermophilus and
    Clostridium sporogenes
  • ?Bioburden Calibrated BIs from environmental
    and process
  • isolates such as E. coli
  • ?Type of BI
  • Spore strips or spore suspension into the
    suspending medium
  • ?Microbial challenge studies are conducted
    concurrently with the heat penetration studies

20
Validation Report
  • ?Common elements of all reports
  • ?Identification of the task report by number
  • ?Reference to protocol
  • ?A brief summary of the range of operational
    conditions experienced and how they were
    controlled
  • ?A procedure for maintaining control within the
    approved range
  • ?A summary and analysis of the experimental
    results
  • ?A brief description of any deviation
  • ?Conclusion
  • ?Review and approval
  • ?Cycle development reports are not usually a part
    of the validation report

21
Maintenance of Validation
  • ?A routine calibration program for all
    instruments critical to the operation of the
    sterilizer and its support system
  • ?A preventative maintenance program including
    periodic operational rechecks and comparison to
    OQ record
  • ?Routine monitoring of bioburden and periodic BI
    challenges(optionally)
  • ?Operating records and equipment logs
  • ?Process and equipment change control procedures
    including review to establish whether additional
    validations are required
  • ?On-going validation

22
Controversial Issues
  • ?Incubation of the sterility test 7 days vs. 14
    days
  • ?USP provide information concerning critical
    parameters for Parameteric Release
  • ?Reduction extent and frequency of revalidation
  • ?Verification of D-value of BIs
  • ?Use of alternative to B. stearothermophilus as a
    BI
Write a Comment
User Comments (0)
About PowerShow.com