Title: The Accufloc Streaming Current Monitor
1The AccuflocStreaming Current Monitor
2Introduction
- Accufloc Streaming Current Monitor for the
monitoring and control of coagulation - This presentation covers
- Overview of the relevant parts of water treatment
- Background theory of particle charge distribution
- Streaming current theory and usage
- Benefits of SCM usage and applications overview
- Accufloc installation and use
- Accufloc options and ordering
- Maintenance procedures
3Water Treatment
- A conventional water treatment plant doses a
coagulant chemical to cause the particles to
stick together - The water is then gently mixed to cause the floc
size to increase - The water slowly flows through a clarifier, or
settling basin, to cause the floc to settle out.
4- Conventional Plant Overview
5 6- Mixer and flocculator (gentle mixer)
7- Clarifier viewed from top
8- Typical small, above-ground clarifier
9Background Theory
- Most naturally occurring particles in surface
water are clays which have a negative surface
charge - Like charges repel each other.
- The main action of the coagulant is to neutralise
the charge on the particles to allow them to
combine into particles large enough to settle out
of suspension
10- Neutral particles can bind together
11- The double layer model explains the distribution
of ions around each colloidal particle. - The Stern layer ions are tightly bound to the
particle. - A dynamic equilibrium of negative and positive
ions forms outside the Stern layer, known as the
diffuse layer. - The zeta potential is measured outside the Stern
layer where shear occurs. - The zeta potential can be measured in the lab
using a microscope is to observe turbidity
particles in an electrophoresis cell. - Changes in ion concentration, such as pH changes,
affect zeta potential.
12Streaming Current
- The streaming current meter (SCM) was invented
around 1966 by F.W Gerdes. It consists of a
piston driven up and down into a close-ended
chamber - It is based on the effect where the surfaces
which the colloidal particles flow past, quickly
take on their charge characteristics - The water flowing rapidly up and down through the
annulus results in displacement of the
counter-ions. A current flows through the
electrodes to remove the charge separation
13History
- An online version of Zeta-Potential Measurement
- Used (badly) on raw water in the 1970s
- Started to be used successfully in the late 1980s
and 1990s on dosed water - Some countries, such as New Zealand, now require
use of a streaming current meter for the top
grades of water. - Knowledge and correct set-up is important
14Modern SCM Usage
- Feedback control of coagulant dosage
- The SC set-point is determined by jar tests
15- This will compensate for changes in the
concentration of particles, their zeta-potential
and plant flow - However periodic jar tests are required because
- - The sweep floc effect means that zero
zeta-potential does not always result in the best
coagulation, the effect this has will change
depending on raw water pH. - The zero SC reading is typically offset from the
zero zeta-potential because of the intrinsic
surface charge of the piston itself. - Wear and contamination of the piston surface
gradually changes the SC reading relative to the
zeta-potential.
16Relationship between SC and settled water
turbidity
17Jar Testing
- You will still have to do this, although not as
often. - Weekly to 6 monthly, whenever significant
variation in raw water quality occurs. - Set the pump speed to the optimal dosage, then
set the SCM so it reads zero at this point
18Benefits of SCMs
- More consistent clarifier operation
- Reduced coagulant chemical costs
- Longer filter runs
- Automatic dosing adjustments
- Reduced operator call outs
- Better polish to the water
19Suitable Applications almost any coagulant
dosing
- Conventional plants with clarifiers. If the jar
results have a definite minimum then electrical
effects are significant and a SCM can be used. - Direction filtration plants (no clarifier), here
the target SC is typically set to a negative
value to ensure pin-floc forms. - Flotation (DAF) plants.
20Less Suitable Applications
- Plants with low turbidity and high organic
loading. - Plants which use a much higher coagulant dosage
than normal to compensate for poor pH control. - These plants have a high coagulant dose,
significantly above the electrically neutral
point. For example 100 ppm. - Applications where coagulation does not occur
21The Accufloc
Displays
Keypad
Terminal Strip
Mounting plate
Body, contains cam
Motor
Body, contains bearing
Sample Chamber, contains piston
Sensor
Sample Flow
22Installation
- Design of the sampling is important.
- The sample must be completely mixed
- Time for the water to get from the mixer to the
sample point should be less than 30sec. - Sample lines must be resistant to clogging or
fouling and easy to clean or flush - Recommended flow rate is 2-4 L/min, but can go up
to gt10L/min. - A constant head is recommended
- The SCM does not pump the sample through itself.
23Example Installation
24Controller Tuning
- Controller tuning is critical to good results
- PID auto tuners use at your own risk. They are
often not well suited to this type of process. - Controller tuning does not have to be very tight
- Tuning can be done by experience or by formula. A
simple procedure is outlined in the manuals
appendices.
25The Displays
The lower display always shows the unadjusted
streaming current reading. Its span can be
calibrated but it will always read zero when
there is no SC. (This is not necessarily zero
zeta-potential) The upper display shows the
difference between the reading and the target SC.
This is like single display SCMs. When the target
SC is set right, zero on the upper display
corresponds to the desired coagulant dosage.
26Span and Zero
- The span setting affects both displays and
determines the sensitivity of the readings. - There are no absolute units of Streaming Current,
so the span can be calibrated to any value on any
sample. For example 10.0 on raw water. - It is not necessary to change the factory default
calibration. The span only needs to be adjusted
for conveniance. - The easiest way is to open the filter drain and
pour the sample into the top of the weir.
Suggested 5-10 litres of raw water. - A sample dosed with alum will react with air and
the SC reading obtained from the sample will
decrease over time. Not advisable for
calibration. - The zero point needs to be set after the span is
adjusted. This is set on water with the desired
coagulant dosage
27Using the Menus
- A full menu map is in back page of the manual
28Some Useful Menu Items
- Averaging. (Asec) A rolling average is performed
over a time period adjustable from 1 to 60
seconds. - Calibration Value. (CaLr) This is what the
reading will be after a span calibration is
performed. - Should be negative if calibration is performed on
under dosed water, or positive if performed on
overdosed water. - Decimal Point Position. (dP) Can be 0.00 or 00.0
- Set Defaults. (SEt dEF) This returns all settings
which affect the calibration and reading to the
factory defaults, and leaves the rest alone.
29Common Options
- 4-20mA output.
- Corresponds to the zeroed (upper) reading.
- The range of this can be set independently of the
calibration range. - Menu items under 4-20 are reading for 4mA (Rd4)
and reading for 20mA (Rd20). - Built in PID Controller.
- Manually adjustable P and I settings. Auto/Man
button on front. - Remembers its output when the unit is switched
off. - Menu items under CtrL are Proportional Band as a
(Pb) and Integral Reset Rate in minutes (Rt)
30Common Options 2
- Up to 2 alarm relays.
- Each has an independent set-point(SP), hysteresis
(HySt) and mode. - Mode can be configured as high level alarm (HI),
low level alarm(LO) and NC fault alarm (FLt) . - Flushing Mechanism.
- Uses an automatically controlled solenoid valve
to inject high pressure clean water into the
sample chamber. - This reduces the need for manual cleaning, but
does not remove it. - Has adjustable interval(Int), duration(dur) and
post flush hold time (PFht).
31Ordering Information
32Maintenance Cleaning
- Contamination of the sensor and piston surfaces
will result in slower response and drift. - Cleaning should be performed with a brush wetted
with sample water. A toothbrush is ideal. - Do not handle any wetted surfaces, any oily
residue may affect the reading for up to 30
minutes. - Immediately after cleaning the reading may be
disturbed, this will normally stabilize within 5
minutes.
33Weekly cleaning is more than sufficient in most
plants
Does not need cleaning, but shows signs of wear
Really needs cleaning
34Symptoms of Worn Parts
- Zero point drifts away from where it is set
- Large offset, but poor sensitivity to changes
- Vertical scouring on the piston or sensor
- Significant staining of the piston
- Looseness in the piston
- Mechanical slackness
- A worn out SCM can still be useful with frequent
attention, but why live with it?
35Replacement Parts Piston
- Piston may need replacement every 1-2 years.
- This depends on the condition of the surface at
the microscopic level - Extremely dependant on water conditions
Piston (including rod) is replaced by undoing a
bolt at the top
36Replacement Sensor
- Sensor may need replacement every 2-5 years.
- It is subject to the same wear forces as the
piston but is more durable and its condition is
less critical - It is removed and replaced as a unit.
- This takes 5-10 minutes in the field as the front
panel of the electronics must be removed to
connect the wires.
37Replacement Bearings
- Linear bearings may need replacing approx. every
5 years. - They hold the piston straight and need replacing
when it develops horizontal slack - Water quality affects the life of the bottom
bearing - They can be replaced in the field in 30 minutes.
They simply slot into a removable part of the
main body.
38Common Issues with SCMs
- Worn out parts causes drift and inconsistency
- Controller tuned incorrectly. May cycle, esp. at
lower flow rates, or be slow to respond - Poor sample point, incomplete mixing, easily
blocked sample lines - Lack of understanding, such as expecting miracles
39Summary
- A Streaming Current Monitor (SCM) measures the
charge of particles in water - A SCM will be useful for the control of coagulant
dosing - It needs to be set up correctly to be useful
- The Accuflocs operation is relatively user
friendly - It does need, at least, a little maintenance to
continue to be useful - Visit our website www.accufloc.com for more
information