Title: AACN
1AACN
- Pulmonary Artery Pressure Measurement
2Lecture Content
- Patient Positioning
- Identify Phelbostatic Axis
- Leveling
- Square Wave Test
- Obtain PA Reading
-
AACN PAP Measurement Practice Alert
3PATIENT POSITIONING
AACN PAP Measurement Practice Alert
4LEVELING
- Eliminates effects of hydrostatic forces on the
observed hemodynamic pressures - Ensure air-fluid interface of the transducer is
leveled before zeroing and/or obtaining pressure
readings - Phlebostatic axis
- Level of left atrium
- 4th ICS ½ AP diameter
- Mark the chest with washable felt pen
- AACN PAP Measurement Practice Alert
5IDENTIFY PHLEBOSTATIC AXIS
- Intersection of the 4th ICS and ½ the
anterior-posterior diameter of the chest
McHale DL, Carlson KK. AACN Procedure Manual for
Critical Care 4th ed WB Saunders Philadelphia,
Pa 2001 (479) With permission from Elsevier
AACN PAP Measurement Practice Alert
6Mark location chest wall with washable felt pen
Used with permission of PACEP Collaborative
AACN PAP Measurement Practice Alert
7LEVEL TRANSDUCER SYSTEM
Ensure air-fluid interface of the transducer is
level to phelbostatic axis
Used with permission of PACEP Collaborative
- AACN PAP Measurement Practice Alert
8LEVEL TRANSDUCER SYSTEM
- Relevel the transducer with any
- change in the patients position
- Referencing the system 1 cm
- above the left atrium decreases the
- pressure by 0.73 mm Hg
- Referencing the system 1 cm
- below the left atrium increases the
- pressure by 0.73 mm Hg
- AACN PAP Measurement Practice Alert
9SQUARE WAVE TEST
- Determines the ability of the
- transducer to correctly reflect pressures
- Perform at the beginning of
- each shift
- AACN PAP Measurement Practice Alert
10SQUARE WAVE TEST
- Figure A
- Expected square wave test
- Figure B
- Over damped
- Figure C
- Under damped
Reprinted from Darovic GO. Hemodynamic
Monitoring Invasive and Noninvasive Clinical
Application 2nd ed. Philadelphia,Pa WB Saunders
Co1995161-162 Used with permission from Elsevier
- AACN PAP Measurement Practice Alert
11UNDER DAMPED SYSTEM
- Over responsive, exaggerated, artificially spiked
waveform - SBP erroneously high DBP erroneously low
- Causes small air bubbles, too long of tubing,
defective transducer
Reprinted from Darovic GO. Hemodynamic
Monitoring Invasive and Noninvasive Clinical
Application 2nd ed. Philadelphia,Pa WB Saunders
Co1995161-162 Used with permission from Elsevier
- AACN PAP Measurement Practice Alert
12OVER DAMPED SYSTEM
- Sluggish, artificially rounded blunted
appearance - SBP erroneously low DBP erroneously high
- Causes large air bubbles in system, compliant
tubing, loose/open connections, low fluid level
in flush bag
Reprinted from Darovic GO. Hemodynamic
Monitoring Invasive and Noninvasive Clinical
Application 2nd ed. Philadelphia,Pa WB Saunders
Co1995161-162 Used with permission from Elsevier
- AACN PAP Measurement Practice Alert
13PAP DOCUMENTATION
- Measure at end expiration
- Measure pressures from a graphic tracing
- Measure pulmonary capillary wedge pressure at
end-expiration using the mean of the a wave - a wave indicates atrial contraction and falls
within the P QRS interval of the corresponding
ECG complex
- AACN PAP Measurement Practice Alert
14RESPIRATORY COMPONET
- Changes in intrathoracic pressure during
respiration change PAP readings - Record and trend pressure readings at end
expiration
- AACN PAP Measurement Practice Alert
15Used with permission of PACEP Collaborative
AACN PAP Measurement Practice Alert
16Used with permission of PACEP Collaborative
AACN PAP Measurement Practice Alert
17Used with permission of PACEP Collaborative
AACN PAP Measurement Practice Alert
18RESPIRATORY VARIATION
Mechanical ventilation
Spontaneous ventilation
Aherns Ts, Taylor LK. Hemodynamic Waveform
Analysis. Philadelphia, Pa WB Saunders 1992
27 Used with permission from Elsevier
- AACN PAP Measurement Practice Alert
19Used with permission of PACEP Collaborative
AACN PAP Measurement Practice Alert
20Used with permission of PACEP Collaborative
AACN PAP Measurement Practice Alert
21Used with permission of PACEP Collaborative
AACN PAP Measurement Practice Alert
22Used with permission of PACEP Collaborative
AACN PAP Measurement Practice Alert
23- Thanks
- for your attention