Title: PTH ASSAY A PROBLEM AND A SOLUTION


1
PTH ASSAY- A PROBLEM AND A SOLUTION
  • DR.SIMON WALKER(Scottish Biochemists Network),
    ANDY ELLIS (on behalf of UKNEQAS),
  • DR.ALISON ALMOND(Scottish Renal Registry)

2
PO4
3
(No Transcript)
4
.....THE PROBLEM
  • 7 DIFFERENT ANALYSERS ACROSS SCOTLAND BUT ALL
    USING SAME METHOD

5
.....THE PROBLEM
  • 7 DIFFERENT ANALYSERS ACROSS SCOTLAND BUT ALL
    USING SAME METHOD
  • WIDE VARIATION IN HOW REFERENCE RANGES
    ESTABLISHED- BUT, REFERENCE RANGES ALL SIMILAR

6
.....THE PROBLEM
  • 7 DIFFERENT ANALYSERS ACROSS SCOTLAND BUT ALL
    USING SAME METHOD
  • WIDE VARIATION IN HOW REFERENCE RANGES
    ESTABLISHED- BUT, REFERENCE RANGES ALL SIMILAR
  • EXTERNAL QUALITY ASSESSMENT DATA SHOW lt3.5
    VARIABILITY ACROSS WORKING RANGE

7
......THE SOLUTION.
  • ADOPT A SINGLE REFERENCE RANGE FOR P04
    ACROSS SCOTLAND.

8
Corrected Calcium
9
(No Transcript)
10
.....THE PROBLEM
11
.....THE PROBLEM
  • 6 different analysers using 2 different
    techniques for total calcium

12
.....THE PROBLEM
  • 6 different analysers using 2 different
    techniques for total calcium
  • 3 different reference ranges for total calcium

13
.....THE PROBLEM
  • 6 different analysers using 2 different
    techniques
  • 3 different reference ranges
  • 2 different techniques for albumin measurements

14
.....THE PROBLEM
  • 6 different analysers using 2 different
    techniques
  • 3 different reference ranges
  • 2 different techniques for albumin measurements
  • Reference ranges for albumin vary

15
.....THE PROBLEM
  • 6 different analysers using 2 different
    techniques
  • 3 different reference ranges
  • 2 different techniques for albumin measurements
  • Reference ranges for albumin vary
  • Lowest albumin reported varies 1g/L-10g/L

16
........THE REAL HEADACHE
17
........THE REAL HEADACHE
  • 4 labs.do not adjust for albumin

18
........THE REAL HEADACHE
  • 4 labs.do not adjust for albumin
  • 4 report only total Ca, 4 report only adjusted
    Ca, 8 report both

19
........THE REAL HEADACHE
  • 4 labs.do not adjust for albumin
  • 4 report only total Ca, 4 report only adjusted
    Ca, 8 report both
  • 7 DIFFERENT ADJUSTMENT FORMULAE!

20
....ROUGHLY TRANSLATED..
  • ALBUMIN CALCIUM
  • RANGE OF CORRECTED Ca
  • 48 2.24
  • 30 2.3
  • 25 2.44
  • 40 2.58
  • 1.99 2.20
  • 2.43 - 2.64
  • 2.65 2.86
  • 2.53 2.7

IN 40 60 OF CASES VARIATION IN cCa CAUSED
BY DIFFERENT FORMULAE WOULD HAVE AFFECTED
CLINICAL DECISIONS

21
.......THE SOLUTION.
  • All labs. will report a corrected calcium, but
    the correction will only apply if albumin is
    gt25g/L.
  • One correction factor will be used universally,
    40/0.02.

22
PARATHYROID HORMONE
23
(No Transcript)
24
ONE PATIENT 5 LABORATORIES
DiaSorin Liaison (GRI)
Beckman Access (BORDERS)
Roche Elecsys (DGRI)
Siemens Centaur (KIRKCALDY)
Siemens Immulite (RIE)
PARATHYROID HORMONE LEVEL (pmol/L
25
ONE PATIENT 5 LABORATORIES
DiaSorin Liaison (GRI)
3.4 FOLD DIFFERENCE
Beckman Access (BORDERS)
Roche Elecsys (DGRI)
Siemens Centaur (KIRKCALDY)
Siemens Immulite (RIE)
PARATHYROID HORMONE LEVEL (pmol/L
26
IN EACH OF THE 17 OTHER PATIENTS.....
1.4 -4.2 FOLD DIFFERENCE
27
PTH ASSAYS USED IN SCOTLAND
  • ASSAY
  • SIEMENS IMMULITE (RIE)
  • SIEMENS CENTAUR (KIRKCALDY)
  • ROCHE ELECSYS (DGRI)
  • DIASORIN LIAISON (GRI)
  • BECKMANS ACCESS (BORDERS)

28
PTH ASSAYS USED IN SCOTLAND
  • ASSAY
  • NORMAL RANGE
  • SIEMENS IMMULITE (RIE)
  • SIEMENS CENTAUR (KIRKCALDY)
  • ROCHE ELECSYS (DGRI)
  • DIASORIN LIAISON (GRI)
  • BECKMANS ACCESS (BORDERS)
  • 1.2 -7.6 pmol/L
  • 1.2 - 7.6 pmol/L
  • 1.6 - 6.9 pmol/L
  • 1.0 6.5 pmol/L
  • 1.2 7.6 pmol/L

29
THE PLAN FOR SRR BONE AUDIT

30
THE PLAN FOR SRR BONE AUDIT
  • THE RENAL UNITS
  • START APRIL 2009
  • BIANNUAL APRIL/NOVEMBER
  • PRE-DIALYSIS BLOODS AFTER SHORT ID GAP
  • TOURNIQUET REMOVED BEFORE WITHDRAWING BLD.
  • CALCIUM/PHOSPHATE/ ALBUMIN in SERUM GEL TUBE
  • EDTA PLASMA FOR PTH

31
THE PLAN FOR SRR BONE AUDIT
  • THE LABORATORIES
  • THE RENAL UNITS
  • CORRECTED CALCIUM REPORTED BY ALL LABS.IF
    ALBUMIN gt25g/L
  • 40/0.02 CORRECTION APPLIED IN ALL LABS.
  • PTH WILL BE REPORTED IN pmol/L BY ALL LABS.
  • NEW ASSAY SPECIFIC PTH REFERENCES WILL BE
    APPLIED (NOVEMBER 2009)
  • START APRIL 2009
  • BIANNUAL APRIL/NOVEMBER
  • PRE-DIALYSIS BLOODS AFTER SHORT ID GAP
  • TOURNIQUET REMOVED BEFORE WITHDRAWING BLD.
  • CALCIUM/PHOSPHATE/ ALBUMIN in SERUM GEL TUBE
  • EDTA PLASMA FOR PTH

32
THANKS TO - SIMON WALKER ANDY ELLIS KAREN