Analysis of ABG Samples - PowerPoint PPT Presentation

1 / 18
About This Presentation
Title:

Analysis of ABG Samples

Description:

... is done by setting the 'slope' potentiometer to a low pH buffer (6.84) and ... the 'balance' potentiometer with a near-normal buffer (7.384) with an ... – PowerPoint PPT presentation

Number of Views:640
Avg rating:3.0/5.0
Slides: 19
Provided by: jhomer
Category:

less

Transcript and Presenter's Notes

Title: Analysis of ABG Samples


1
Analysis of ABG Samples
2
ABG Analysis, Introduction
  • pH, PaCO2, PaO2 are measured directly by special
    electrodes contained in a device made for that
    purpose
  • Other indirect measurements can be made or
    calculated from the above measurements i.e.,
    HCO3-, O2 Sat.

3
pH
  • pH electrode is constructed of two half cells,
    which develop an electrical potential when
    connected together
  • Reference electrode maintains a constant
    potential and is bathed in KCl
  • Glass electrode (Sanz electrode) develops an
    electrical potential that is proportional to the
    amount of H present

4
PaCO2 Electrode (Severinghaus)
  • Measures carbon dioxide tensions by allowing the
    CO2 gas to undergo a chemical reaction that
    produces hydrogen ions
  • The hydrogen ion concentration produced is
    directly proportional to the PCO2 in contact with
    the membrane of the electrode
  • Operates on the principle of electric potential
    between electrodes
  • CO2 H2O H2CO3 H HCO3-

5
O2 Electrode (Clark)
  • Is a polarographic device that measures oxygen
    tensions by oxidation/reduction reactions, a
    chemical process that generates measurable
    electrical currents
  • Has platinum cathode and silver anode immersed in
    an electrolyte solution
  • Volume of O2 will be directly proportional to the
    number of electrons used in the cathode rxn. and
    by measuring current ? is a measure of O2
    diffused across membrane

6
QA in Blood Gas Analysis
  • ABG lab must be able to assure accurate and
    reliable results
  • The above is accomplished by applying protocols
    in 3 areas - pre-analytic error -
    calibration - quality control

7
Pre-analytic Error
  • All factors that cause variance in lab results
    prior to the sample arriving in the ABG lab.
  • 4 factors assoc. with signif. P. E. are - air
    bubbles in sample
  • - time delay (iced sample with more than 60
    min. or uniced with more than 10 min.)
  • - blood clots in sample
  • - small sample size where excessive
    anticaogulation is suspect

8
Calibration
  • Purpose is assure consistency
  • Def. the systemic standardization of the
    graduation of a quantitative measuring instrument
  • Calibrating standards for blood gas analyzers
    should simulate the physical properties of blood
    and meet manuf. specs.
  • When 2 standards are used ---gt 2-point
    calibration, performed after 50 blood gases or at
    least every 8 hours

9
Calibration (contd)
  • A one-point calibration is an adjustment of the
    electronic response of an electrode to a single
    standard and is performed more freq. than a 2 pt.
    cal., ideally prior to each sample analysis

10
pH Calibration
  • Several types of buffer solutions are acceptable
    for pH electrode calibration
  • Buffers are excellent standards because of their
    long shelf life and stability for days after
    being opened
  • 2-point cal. of pH electrode is done by setting
    the slope potentiometer to a low pH buffer
    (6.84) and setting the balance potentiometer
    with a near-normal buffer (7.384) with an
    accuracy of /- 0.005

11
PaCO2 Calibration
  • PaCO2 electrode may be calibrated by introducing
    a known gas concentration
  • Gases used are 5 and 10 CO2, /- .03 - 1-point
    cal. uses 5 standard to set the balance
    point - 2-point cal. uses 5 for balance
    point and 10 for slope point
  • PCO2 (BP - 47 mmHg) x CO2

12
PO2 Calibration
  • Properly calibrated PO2 electrodes perform within
    the manuf. stated accuracy in PaO2 ranges below
    150 mmHg but may vary 20 at 500 mmHg
  • 0 oxygen is used for the slope point and 12
    or 20 for the balance point
  • PO2 (BP - 47 mmHg) x O2

13
Quality Control
  • Refers to a system that documents the accuracy
    and reliability of the blood gas measurements and
    is essential to assure accuracy in the blood gas
    lab
  • Media available as blood gas controls include
  • - aqueous buffers
  • - glycerin soltn.
  • - human/animal serum and blood - artificial
    blood
  • A QC system must ID problems and specify
    corrective action, document. of accept. oper.

14
QC (contd)
  • Documentation of QC is usu. on Levy-Jennings
    Chart which shows measured results on the y axis
    versus time of measurement on the x axis
  • SD is used to summarize a mass of data the
    difference between a number in a data set and the
    mean of the data set is called a deviation. A
    deviation shows how much a number varies from the
    mean

15
QC (contd)
  • A properly functioning electrode that repeatedly
    analyzes a known value will produce results
    within a rel. small range, e.g., a PaCO2
    electrode that analyzes a 40 mmHg standard 100
    times will produce results where 2/3 of the
    measurements are 39 - 41 mmHg and nearly all
    measurements fall in 38 - 42 range
  • 95 of the control measurements should fall
    within 2 SD

16
QC (contd)
  • Random errors indicates a value outside of 2 SD
    of the mean a single random error has minor
    signif., but if number increased the machine and
    techniques must be evaluated
  • Systematic errors is recurrent measurable
    deviation from the mean
  • Causes of systematic errors
  • - contaminated standard
  • - variations in electrode temp.
  • - inconsistent introduction of standard

17
QC (contd)
  • Causes of systematic error (contd) -
    inconsistent calibration technique - change in
    QC standard storage or prep. - electrode
    problems, e.g., protein contamin., membrane
    malfunction, contamin. electrolyte, or
    electrical problems

18
QC Levels
  • Level 1 simulates a patient hypoventilating
  • Level 2 simulates a patient with normal
    ventilatory status
  • Level 3 simulates a patient hyperventilating
Write a Comment
User Comments (0)
About PowerShow.com