Title: Sensitivity: Large Sample Volume Splitless
1Sensitivity Large Sample Volume Splitless
2Speed Sensitivity in a NEW way!
3A breakthrough technique for ULTRA sales
- Large volume techniques OC and PTV
- Overcoming SSL limitations for LV CSR
- When, why, and what of LVSL
- Practical aspects of LVSL injection
- Understanding how LVSL
- Some fundamentals of operation
- Competition
- Conclusions
4Large Sample Volume Injection
- Increased sensitivity
- Simplified sample preparation procedures
- cost reduction
- Better use of the sample material
- Lower amount of both solvent and sample used
5Large Sample Volume Injection
- Large Volume On-Column Injection
- suitable for volatile solute
- few operating parameters to be set
- sensitive to contaminants
- PTV Solvent Split Injection
- robust to contaminants
- difficult for volatile solutes
- several operating parameters to be set
6And with the SSL?
- Sample volume is limited to a few microliters
- The vaporizing chamber must be sufficiently large
to house the whole sample vapour. - The vaporizing chamber cannot be enlarged beyond
a given size.
7SSL Overcoming of Sample Volume Limitation
- Slow injection?
- Introducing liquid at a rate that vapours can be
immediately transferred to the column. - Evaporation inside the needle
- DISCRIMINATION OF HIGH BOILING SOLUTES
- Concurrent Solvent Recondensation (CSR)
- Solvent recondenses into a column at the same
rate as produced by sample evaporation.
8When and why LVSL injection?
- 1uL injection
- Initial Oven T. 60C
9When and why LVSL injection?
- WHEN USE LVSL
- Whenever you use the SSL injector the Worlds
most used injector - When you need more sensitivity than standard
just inject more in the SSL injector!! - WHY USE LVSL
- Simple to operate
- Performs on standard SSL injector
- NO impact on method validation!!!
10LVSL what do you need ?
New Proprietory DPFC with septum purge closure
capability. Forward pressure regulation
with sensor on the purge line and low dead volume
connections/tubes between valves and injector.
Will be the new standard for 250 1000 Kpa DPFC.
Software assist A simple dedicated parameter
selection program (separate from Chromcard)
11LVSL how do you install it?
CSR Large Volume Splitless GC Setup and Injection
Press-fit
Analytical column
Pre-column
12LVSL What else do you have to change?
1330uL injection for 30 times more sensitivity!
- No difference in method setup!!
- No method revalidation due to h/w differences!!
14CSR Large Volume Splitless InjectionSteps of
the Process
- Injection with liquid band formation
- Auto pressure surge
- Concurrent solvent recondensation into the
precolumn - Sample transfer
- Solvent evaporation in the precolumn
15Concurrent Solvent Recondensation (LVSL)
Bottom of the liner
- Restricted Evaporation Rate
- Injection with Liquid Band Formation
- Evaporation from a single droplet
- Increased Transfer Rate
- Solvent recondensation into a precolumn generates
a strong suction effect
Solvent vapour
Liquid sample
Glass wool
High flow
Recondensed solvent
Precolumn
16Concurrent Solvent Recondensation
Transfer occurs concurrently with solvent
evaporation
- The vapour volume no longer limits the injection
volume
17Step 3 visual difference with without CSR
- 35 uL Hexane injection
- BT 69C
- 65C starting temp
- Solvent recondensation
- Strong suction effect -fast pressure
equilibration
- 100C starting temp
- No solvent recondensation,
- No suction effect - long time for pressure
equilibration
18Concurrent Solvent Recondensation
- Injection of 30 ul of a hexane solution of
n-alkanes - Vapour volume 7.2 ml (300C and 50 kPa)
- Liner volume 1.6 ml
19Linearity from 5 to 50 ul
Precolumn 5 m x 0.32 mm id deactivated (Mega)
Column RTX1 15 m x 0.32 mm id, 0.5 um
20Injection Conditions
21Recoveries
- Better than those obtained with conventional SL!
Peak area normalized on C20 ratios calculated on
1 ul OC injection
22Analysis of PAHs with GC-MS
25 ul PAHs mix, 20 ppb in hexane Precolumn 5 m x
0.32 mm id Column RTX5 30 m x 0.25 mm id
23Analysis of PAHs with GC-MS
25 ul, 6 consecutive injections
24Analysis of OPPs with GC-FPD
40 ul OPPs mix, 12 ppb each in hexane Precolumn
5 m x 0.32 mm id Column 15 m x 0.32 mm id, RTX1
0.50 um
25Analysis of OPPs with GC-FPD
40 ul, 6 consecutive injections
26Competition NO COMPETITORS!!!
- Thermo is in dominant position competition is
trying to defuse the bomb. We are in the position
to put pressure
27Competition NO COMPETITORS!!!
28Some Takeaways to keep in mind
- SSL INJECTORS ARE NO LONGER ALL THE SAME!!
- Ours is different in its substance
- TO USE WHENEVER THERE IS SSL INJECTOR
- Basic option and add-on kit (as seen in video!!)
- EASY TO PERFORM AND EXCELLENT RESULTS
- No specific expertise, total recovery and No
discrimination - NO REQUIREMENT FOR METHOD REVALIDATION
- NO COMPETITION
- Only Trace GC Ultra has Large Volume SL
- Only Trace GC Ultra has suitable pneumatics
- PROPRIETARY!!
- WE ARE THE FIRST LETS CAPITALISE!!!
29Some Takeaways to keep in mind
- SSL INJECTORS ARE NO LONGER ALL THE SAME!!
- Ours is different in its substance
- TO USE WHENEVER THERE IS SSL INJECTOR
- Basic option and add-on kit (as seen in video!!)
- EASY TO PERFORM AND EXCELLENT RESULTS
- No specific expertise, total recovery and No
discrimination - NO REQUIREMENT FOR METHOD REVALIDATION
- NO COMPETITION
- Only Trace GC Ultra has Large Volume SL
- Only Trace GC Ultra has suitable pneumatics
- PROPRIETARY!!
- WE ARE THE FIRST LETS CAPITALISE!!!