Title: Eichrom Resin Columns: A New breakthrough
1Eichrom Resin Columns A New breakthrough
- Anil Thakkar
- Eichrom Users Group Workshop
- May 3, 2005
- Oak Ridge, TN
2Background details
- Stripping issues of Th on TEVA prepacked columns
reported by numerous customers - Investigated all the components of the prepacaked
column to trouble shoot the problem resin,
column body, resin bead size, frits.
3Previously discussed investigations.
- Column frits identified as the main cause of the
problem - Investigations on frits led to the understanding
of different frit types - Frits with ionic surfactant
- Impact Th elution from TEVA columns
- Impact low level U elution on UTEVA columns
- Frits with no surfactant
- Measured slow flow rates at gravity
- Mechanical pressure required to start the flow
- Measured inconsistent flow rates
4Additional issues reported
- Poor elution of low level U on UTEVA columns
reported - An inconsistent elution of Th on UTEVA columns
has been observed
5Additional research continues
- A non-ionic surfactant frit material found and
put to test
6TEVA Column Performance Test
- Columns packed with non-ionic frit material
- Measured performance of columns over a period of
time
7TEVA Columns Th Yields
8TEVA Columns Pu Yields with 0.5M HCl
INEEL procedure uses 0.5M HCl to elute Pu
9TEVA Columns Am Yields(Am/ Lanthanide
Separation)
10TEVA Columns Tc-99 Yields
11- UTEVA, TRU, Sr and Ln columns tested after
- 2 weeks of packing
12UTEVA Columns U Th Yields
Th eluted with 5M HCl/0.05M oxalic acid
13TRU Columns Am Pu Yields
14Sr Columns Sr Yields Recoveries
15Ln Columns Ac-228 Yields
16Improvements Summary
- Non-ionic surfactant frits improve performance
- Th elution on TEVA
- Th elution on UTEVA
- Pu elution on TEVA
- No impact observed on the performance of the
following products with non-ionic frits - Am Pu on TRU resin
- Tc-99, Am on TEVA resin
- Sr elution on Sr resin
- Ac elution on Ln resin
-
-
17Future resin column inventory
- Preliminary evaluation of prepacked column
products with the new frits (non-ionic
surfactant) has been extremely satisfactory - New frit inventory for future production has been
ordered - Current inventory has old frits (ionic
surfactant) in the prepacked column products - Any users of TEVA resin and UTEVA resin prepacked
columns may consider cartridges in the interim - Contact Eichrom for further assistance